Shock outcome prediction before and after CPR: A comparative study of manual and automated active compression-decompression CPR

被引:11
|
作者
Box, M. S. [1 ]
Watson, J. N. [2 ]
Addison, P. S. [2 ]
Clegg, G. R. [3 ]
Robertson, C. E. [3 ]
机构
[1] S Western Ambulance Serv NHS Trust, Bournemouth BH8 8LE, Dorset, England
[2] Elvingston Sci Ctr, CardioDigital Ltd, Edinburgh, Midlothian, Scotland
[3] Royal Infirm Edinburgh NHS Trust, Dept Accid & Emergency Med, Edinburgh, Midlothian, Scotland
基金
英国惠康基金;
关键词
electrocardiography; cardiopulmonary resuscitation (CPR); active compression/decompression-CPR; cardiac arrest; shock outcome prediction;
D O I
10.1016/j.resuscitation.2008.03.225
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
We report on a study designed to compare the relative efficacy of manual CPR (M-CPR) and automated mechanical CPR (ACD-CPR) provided by an active compression-decompression (ACD) device. The ECG signals of out-of-hospital cardiac arrest patients of cardiac aetiology were analysed just prior to, and immediately after, cardiopulmonary resuscitation (CPR) to assess the likelihood of successful defibrillation at these time points. The cardioversion outcome prediction (COP) measure previously developed by our group was used to quantify the probability of return of spontaneous circulation (ROSC) after countershock and was used as a measure of the efficacy of CPR. An initial validation study using COP to predict shock outcome from the patient data set resulted in a performance of 60% specificity achieved at 100% sensitivity on a blind test of the data. This is comparable with previous studies and provided confidence in the robustness of the technique across hardware platforms. Significantly, the COP marker also displayed an ability to stratify according to outcomes: asystole, ventricular fibrillation (VF), pulseless electrical activity (PEA), normal sinus rhythm (NSR). We then used the validated COP marker to analyse the ECG data record just prior to and immediately after the chest compression segments. This was initially performed for 87 CPR segments where VF was both the pre- and post-CPR waveform. An increase in the mean COP values was found for both CPR types. A signed rank sum test found the increase due to manual CPR not to be significant (p > 0.05) whereas the automated CPR was found to be significant (p < 0.05). This increase was larger for the automated CPR (1.26, p = 0.024) than for the manual CPR (0.99, p = 0.124). These results indicate that the application of CPR does indeed provide beneficial. preparation of the heart prior to defibrillation therapy whether manual or automated CPR is applied. The COP marker shows promise as a definitive, quantitative determinant of the immediate positive effect of both types of CPR regardless of the details of use. In work of a more exploratory nature we then used the validated COP marker to analyse the ECG pre- and post-CPR for all rhythm types (212 traces). We show a significant increase in the COP measure (p < 0.001 in both cases) as indicated by a shift in the median COP marker distribution values. This increase was more pronounced for automated ACD-CPR than for manual CPR. However, a detailed statistical analysis carried out between the groups adjusted for pre-CPR value showed no significant difference between the two methods of CPR (p = 0.20). Similarly, adjusting for length of CPR showed no significant difference between the groups. Secondary, subgroup analysis of the ECG according to the length of time for which CPR was performed showed that both types of CPR led to an increase in the likelihood Of Successful defibrillation after increasing durations of CPR, however results were less reliable after longer periods of continuous CPR. (C) 2008 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:265 / 274
页数:10
相关论文
共 50 条
  • [1] ACTIVE COMPRESSION-DECOMPRESSION CPR
    MAZZIOTTI, A
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1994, 272 (17): : 1325 - 1326
  • [2] ACTIVE COMPRESSION-DECOMPRESSION CPR - REPLY
    LURIE, KG
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1994, 272 (17): : 1326 - 1326
  • [3] Hemodynamic effects of active compression-decompression after prolonged CPR
    Malzer, R
    Zeiner, A
    Binder, M
    Domanovits, H
    Knappitsch, G
    Sterz, F
    Laggner, AN
    RESUSCITATION, 1996, 31 (03) : 243 - 253
  • [4] ACTIVE COMPRESSION-DECOMPRESSION CPR - A PROGRESS REPORT
    LURIE, KG
    RESUSCITATION, 1994, 28 (02) : 115 - 122
  • [5] Airway management during active compression-decompression CPR
    Rottenberg, EM
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (06): : 448 - 449
  • [6] Comparative effectiveness of standard CPR vs active compression-decompression CPR with CardioPump for treatment of cardiac arrest
    Gunaydin, Yahya Kemal
    Cekmen, Bora
    Akilli, Nazire Belgin
    Koylu, Ramazan
    Sert, Ekrem Taha
    Cander, Basar
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2016, 34 (03): : 542 - 547
  • [7] Airway management during active compression-decompression CPR - Reply
    Stiell, IG
    Hebert, PC
    Wells, GA
    Laupacis, A
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (06): : 449 - 449
  • [8] AUTOMATIC MECHANICAL DEVICE TO STANDARDIZE ACTIVE COMPRESSION-DECOMPRESSION CPR
    WENZEL, V
    FUERST, RS
    IDRIS, AH
    BANNER, MJ
    RUSH, WJ
    ORBAN, DJ
    ANNALS OF EMERGENCY MEDICINE, 1995, 25 (03) : 386 - 389
  • [10] PLUNGERS AND POLEMICS - ACTIVE COMPRESSION-DECOMPRESSION CPR AND FEDERAL-POLICY
    OLSON, CM
    RENNIE, D
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1995, 273 (16): : 1299 - 1300