Occlusion trajectory and a concept of a device for testing operating life of dentures

被引:3
|
作者
Jalbrzykowski, Marek [1 ]
Derlatka, Marcin [2 ]
Urban, Dariusz [3 ]
机构
[1] Bialystok Tech Univ, Fac Mech Engn, Dept Mat & Biomed Engn, St Wiejska 45C, PL-15351 Bialystok, Poland
[2] Bialystok Tech Univ, Fac Mech Engn, Dept Mech & Construct Machinery, Bialystok, Poland
[3] Med Univ Bialystok, Dept Prosthodont, Bialystok, Poland
来源
ADVANCES IN MEDICAL SCIENCES | 2016年 / 61卷 / 02期
关键词
Occlusion trajectory; SMART system; Testing machine for dentures; MOTION; WEAR;
D O I
10.1016/j.advms.2015.12.005
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Purpose: This paper presents an original method for the assessment of occlusion trajectory. On this basis, a special device for the assessment of operating life of dentures was designed. Material/methods: For this purpose, the SMART system by BTS for a comprehensive movement analysis was used. In order to analyze occlusion trajectory, characteristic points on patients' heads were appointed in which markers were placed, in accordance with the rules of measurement in dentistry. Markers' movement was recorded by means of 6 cameras, and then composition of coordinates was performed in a 3D system. Results: In this way, curves representing movements of the characteristic points were plotted which, after the composition, with a considerable approximation, can be regarded as occlusion trajectory. Conclusions: On the basis of the obtained results, a thesis was put forward to the effect that traditional tribological testing machines based on systems of the pin-on-disc, ball-on-disc, etc. types, due to the simplicity of their working movements, are not adequately precise for the purpose of operational assessment of elements of prosthodontics. On this basis, a tribological node for a specialist testing machine for the assessment of operating life of dentures was designed. (C) 2016 Medical University of Bialystok. Published by Elsevier Sp. z o.o. All rights reserved.
引用
收藏
页码:180 / 186
页数:7
相关论文
共 50 条
  • [1] Development of a Device to Study Fatigue Life of Fixed Partial Dentures
    Gutierrez, S. C.
    Meseguer, M. D.
    Estal, R.
    Folguera, F.
    Vidal, V.
    [J]. 4TH MANUFACTURING ENGINEERING SOCIETY INTERNATIONAL CONFERENCE (MESIC 2011), 2012, 1431 : 877 - 884
  • [2] Features of one operating regime of a device with a rosette scanning trajectory
    Omelaev, AI
    Kazakov, VN
    [J]. JOURNAL OF OPTICAL TECHNOLOGY, 1997, 64 (11) : 1039 - 1040
  • [3] Influence of guidance concept in complete dentures on oral health related quality of life - Canine guidance vs. bilateral balanced occlusion
    Schierz, Oliver
    Reissmann, Daniel
    [J]. JOURNAL OF PROSTHODONTIC RESEARCH, 2016, 60 (04) : 315 - 320
  • [4] CONCEPT OF OPTIMIZATION AND DESIGN OF A TESTING DEVICE FOR TESTING SEPARATING GRILLES
    Daneshjo, Naqib
    Majernik, Milan
    Pajerska, Erika Dudas
    Danishjoo, Mirwajs
    [J]. MM SCIENCE JOURNAL, 2018, 2018 : 2585 - 2592
  • [5] OPTIMIZATION OF OPERATING CONDITIONS IN A LABORATORY SOFC TESTING DEVICE
    Skalar, Tina
    Lubej, Martin
    Marinsek, Marjan
    [J]. MATERIALI IN TEHNOLOGIJE, 2015, 49 (05): : 731 - 738
  • [6] Research and Design of Testing Device of Hand Operating Capacity
    Yang Lifang
    Liu Jia
    [J]. 2009 IEEE 10TH INTERNATIONAL CONFERENCE ON COMPUTER-AIDED INDUSTRIAL DESIGN & CONCEPTUAL DESIGN, VOLS 1-3: E-BUSINESS, CREATIVE DESIGN, MANUFACTURING - CAID&CD'2009, 2009, : 1416 - 1419
  • [7] Device for testing and control of IBM PC operating mode
    Kochan, R
    Kopylchak, A
    Korkishko, T
    [J]. MODERN PROBLEMS OF RADIO ENGINEERING, TELECOMMUNICATIONS AND COMPUTER SCIENCE, PROCEEDINGS, 2002, : 179 - 180
  • [8] Calibration of the perthometer concept device for testing contour deviation
    Seferovic, E.
    Halilagic, M.
    Begic, D.
    [J]. Annals of DAAAM for 2003 & Proceedings of the 14th International DAAAM Symposium: INTELLIGENT MANUFACTURING & AUTOMATION: FOCUS ON RECONSTRUCTION AND DEVELOPMENT, 2003, : 407 - 408
  • [9] A combined surgical-endovascular device concept for giant aneurysm neck occlusion
    Wellman, BJ
    Loftus, CM
    Noh, D
    Barnhart, WH
    Howard, MA
    [J]. NEUROSURGERY, 1998, 42 (06) : 1364 - 1368
  • [10] Clinical Testing of the Apnea Prevention Device: Proof of Concept Data
    Zornow, Mark H.