Monitoring of the EEG using the bispectrum analysis on the anesthetic depth

被引:0
|
作者
Ye, Soo Young [1 ]
Jeong, Do Un [1 ]
Shon, Jung Man [1 ]
Park, Jong Duk [1 ]
Choi, Byeong Cheol
Jeon, Gye Rok
机构
[1] Pusan Natl Univ, Dept Interdisciplinary Biomed Engr, Pusan, South Korea
关键词
D O I
暂无
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
In this study, a distribute characteristics on the bispectral coupling 01 EEG were observed according to depth of anesthesia using bispectum analysis method in a type of nonlinear signal processing. 25 patients of the general anesthesia were studied. In the preprocess, base line correction and linear detrend was used. In awakening state of patient in pre-operation and post operation, appearance rate of bispectral coupling was observed that a strong appearance rate was represented in major frequency zone such as (15, 15) Hz but weak appearance rate investigated in the other minor frequency zone. During the anesthesia of patient, a strong appearance rate revealed the low frequency range below (10, 10) Hz and a weak appearance rate viewed in the other frequency range. Also the result was compared with the delta ratio and spectrum edge frequency(SEF). Delta ratio shows significant lower values at indudion (3.60 center dot center dot (0.98) and maintenance(5.0 center dot 1.74) of anesthesia than preoperation (10.30 center dot 2.44) and postoperation (13.03 center dot 2.29)(P < 0.05). These result was consistent with a trend of delta ratio that the index of evaluation for depth of anesthesia known as a reflex of the conscious level of patient's during anesthesia. While these results were not consistent with the analysis result of SEF that the index of evaluation for suppression level of patient's during recovery stage of anesthesia. SEF shows lower values at maintenance (17.00 center dot center dot 1.22) and postoperation (12.20 center dot center dot 1.53) than preoperation (22.50 center dot center dot 1.75) and induction (22.40 center dot center dot 1.18) of anesthesia(p < 0.05). The analysis method of bispecirum was useful information about the index of evaluation according to the level of unconsciousness and suppression among other evaluation factor on depth of anesthesia.
引用
收藏
页码:3211 / 3215
页数:5
相关论文
共 50 条
  • [1] EEG MONITORING FOR ANESTHETIC DEPTH ASSESSMENT
    HAMEROFF, SR
    WATT, RC
    JOLLY, T
    [J]. PROCEEDINGS OF THE ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, PTS 1-4, 1988, : 1851 - 1852
  • [2] Computerized EEG monitoring of anesthetic depth: Quo Vadis?
    Doyle, J
    [J]. CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2000, 47 (10): : 1044 - 1045
  • [3] Computerized EEG monitoring of anesthetic depth: Quo Vadis?
    D. John Doyle
    [J]. Canadian Journal of Anesthesia, 2000, 47 : 1044 - 1045
  • [4] Monitoring depth of anaesthesia using auditory evoked potential and bispectrum
    Wang, Zicong
    Li, Yingchun
    Wen, Peng
    Meng, Max Q. -H.
    [J]. 2007 IEEE INTERNATIONAL CONFERENCE ON INTEGRATION TECHNOLOGY, PROCEEDINGS, 2007, : 316 - +
  • [5] Monitoring of anesthetic depth and EEG band power using phase lag entropy during propofol anesthesia
    Hye Won Shin
    Hyun Jung Kim
    Yoo Kyung Jang
    Hae Sun You
    Hyub Huh
    Yoon Ji Choi
    Seung Uk Choi
    Ji Su Hong
    [J]. BMC Anesthesiology, 20
  • [6] Monitoring of anesthetic depth and EEG band power using phase lag entropy during propofol anesthesia
    Shin, Hye Won
    Kim, Hyun Jung
    Jang, Yoo Kyung
    You, Hae Sun
    Huh, Hyub
    Choi, Yoon Ji
    Choi, Seung Uk
    Hong, Ji Su
    [J]. BMC ANESTHESIOLOGY, 2020, 20 (01)
  • [7] Study of epileptic rat's EEG using bispectrum analysis
    Wu, HY
    Xia, Y
    Lai, YX
    Liu, YS
    Yao, DZ
    [J]. 2005 First International Conference on Neural Interface and Control Proceedings, 2005, : 119 - 122
  • [8] Predicting anesthetic depth using EEG and models of drug interaction.
    Hu, C
    Greenwald, S
    Chamoun, N
    Shafer, SL
    [J]. CLINICAL PHARMACOLOGY & THERAPEUTICS, 1997, 61 (02) : PII45 - PII45
  • [9] Monitoring Depth of Anesthesia Using Detrended Fluctuation Analysis Based on EEG Signals
    Li, Xiaoou
    Wang, Feng
    Wu, Guilong
    [J]. JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2017, 37 (02) : 171 - 180
  • [10] Monitoring Depth of Anesthesia Using Detrended Fluctuation Analysis Based on EEG Signals
    Xiaoou Li
    Feng Wang
    Guilong Wu
    [J]. Journal of Medical and Biological Engineering, 2017, 37 : 171 - 180