Evaluation of H-800/FUS-100 automatic urine analyzer performance

被引:0
|
作者
Ercan, Mujgan [1 ]
Oguz, Esra Firat [2 ]
Kaya, Oguzhan [3 ]
Yilmaz, Fatma Meric [4 ]
机构
[1] Bozok Univ, Dept Med Biochem, Fac Med, Yozgat, Turkey
[2] Ankara Pediat Hematol & Oncol Training & Res Hosp, Biochem Lab, Ankara, Turkey
[3] Ankara Numune Training & Res Hosp, Clin Biochem, Ankara, Turkey
[4] Yildirim Beyazit Univ, Dept Biochem, Fac Med, Ankara, Turkey
关键词
Urine microscopy; Interference; FUS-100; H800; Erythrocytes; Leukocytes; URINALYSIS; MICROSCOPY; SEDIMENT; ERROR;
D O I
10.1515/tjb-2016-0248
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Objective: Automated urine analysis is usually preferred for laboratories with intensive workload. The aim of this study was to evaluate the performance of the automated urine analyser H-800/FUS-100 and detect the error sources. Materials and methods: One thousand four hundred fifty nine fresh urine samples were analyzed with H-800/FUS-100 automated systems. The urine sediment of the samples with discrepant strip and microscopy results were confirmed by manual microscopy. Precision and carry over studies were performed. Results: The discrepancy is detected in a ratio of 5.89% between chemical analysis (H-800) and microscopic analysis (FUS-100) of the device. A total of 86 discrepant samples were detected. Fifty six of 86 were erythrocyte discrepancies (65.1%) and 30 of 86 were leukocyte discrepancies (34.9%). The results of carry over analysis for erythrocyte and leukocyte were 21.85% and 13.64%, respectively. Conclusions: Sixteen (1.09%) of 1459 patients' results in FUS-100 were discrepant with manual microscopy. Commonly, yeasts and crystals affected erythrocyte counts and calcium oxalate and amorphous crystals affected the leukocyte counts. Images should be reviewed for every sample when automated systems are used for urine analysis. Especially if discrepancy is detected between chemical and microscobic analysis, the results should also be confirmed with manual microscopy.
引用
收藏
页码:89 / 92
页数:4
相关论文
共 39 条
  • [1] 迪瑞H-800、FUS-100全自动尿液分析仪的两项指标(BACT、BYST)评价
    吴晓燕
    沈荣南
    金红玉
    [J]. 中国医药指南, 2012, (25) : 134 - 135
  • [2] Evaluation of the iChem™ 100 urine chemistry analyzer compared to the Clinitek® 500 urine chemistry analyzer shows equivalent performance
    Roch, T.
    Dorko, M.
    [J]. CLINICAL CHEMISTRY, 2007, 53 (06) : A168 - A168
  • [3] Evaluation of FUS-2000 urine analyzer: analytical properties and particle recognition
    Benovska, Miroslava
    Wiewiorka, Ondrej
    Pinkavova, Jana
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2018, 78 (1-2): : 143 - 148
  • [4] Comparison of Fully Automated Urine Sediment Analyzers H800-FUS100 and Labumat-Urised with Manual Microscopy
    Yuksel, Hatice
    Kilic, Elif
    Ekinci, Aysun
    Evliyaoglu, Osman
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2013, 27 (04) : 312 - 316
  • [5] PERFORMANCE EVALUATION OF BECKMAN AUTOMATIC CHEMICAL ANALYZER
    GALLWAS, GE
    GRAY, JM
    CARNE, HO
    [J]. CLINICAL CHEMISTRY, 1968, 14 (08) : 781 - &
  • [6] The iChem™ 100 urine chemistry analyzer and the AUTION JET™ AJ-4270 urine chemistry analyzer show equivalent performance
    McDermed, J. E.
    [J]. CLINICAL CHEMISTRY, 2007, 53 (06) : A168 - A168
  • [7] Performance evaluation and clinical application of the UA-5600 automatic dry chemistry urine analyzer in renal diseases
    Liu, Yin
    Zheng, Xiaohe
    Wang, Dong
    Fang, Fei
    Chen, Yao
    Hong, Mengzhi
    Wu, Jiali
    Zhang, Chi
    Qiao, Yangyang
    Izevbaye, Iyare
    Zhang, Shihong
    [J]. TRANSLATIONAL ANDROLOGY AND UROLOGY, 2024, 13 (06)
  • [8] EVALUATION OF ENZYMATIC AMMONIA METHOD FOR URINE ON DU PONT AUTOMATIC CLINICAL ANALYZER
    HINDRIKS, FR
    GROEN, A
    [J]. JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1978, 16 (05): : 289 - 291
  • [9] Evaluation of the Atellica(R) UAS 800: a new member of the automated urine sediment analyzer family
    Aper, Stijn J. A.
    Gijzen, Karlijn
    Luimstra, Jolien J.
    van der Valk, Johanna T. M. H.
    Russcher, Anne
    Kocer, Ruya G.
    Liesting, Eline C.
    Jacobs, Leo H. J.
    Lentjes, Eef G. W. M.
    Demir, Ayse Y.
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2021, 81 (07): : 585 - 592
  • [10] FUS-200与H-800联合检测与显微镜检测尿红细胞的对比分析
    沈小琴
    代超
    [J]. 宁夏医科大学学报, 2013, 35 (04) : 454 - 456