Dosimetric verification of lung phantom calculated by collapsed cone convolution: A Monte Carlo and experimental evaluation

被引:8
|
作者
Najafzadeh, Milad [1 ,2 ]
Nickfarjam, Abolfzal [1 ,3 ]
Jabbari, Keyvan [4 ]
Markel, Daniel [5 ,6 ]
Chow, James C. L. [5 ,6 ]
Takabi, Fatemeh Shirani [1 ]
机构
[1] Shahid Sadoughi Univ Med Sci, Fac Med, Dept Med Phys, Prof Hesabi Blvd,Alem Sq, Yazd, Iran
[2] Hormozgan Univ Med Sci, Fac Paramed, Dept Radiol, Bandare Abbas, Iran
[3] Shahid Sadoughi Univ Med Sci, Radiotherapy Res Ctr, Yazd, Iran
[4] Isfahan Univ Med Sci, Fac Med, Dept Med Phys, Esfahan, Iran
[5] Univ Toronto, Dept Radiat Oncol, Toronto, ON, Canada
[6] Princess Margaret Canc Ctr, Radiat Med Program, Toronto, ON, Canada
关键词
Monte Carlo; treatment planning; collapsed cone convolution; 3D conformal radiotherapy; DOSE CALCULATION ALGORITHMS; ANALYTICAL ANISOTROPIC ALGORITHM; TUMOR-CONTROL PROBABILITY; IMRT TREATMENT; PHOTON; RADIOTHERAPY; DISTRIBUTIONS; SIMULATIONS; MODEL; THERAPY;
D O I
10.3233/XST-180425
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
OBJECTIVE: To evaluate the dose calculation accuracy in the Prowess Panther treatment planning system (TPS) using the collapsed cone convolution (CCC) algorithm. METHODS: The BEAMnrc Monte Carlo (MC) package was used to predict the dose distribution of photon beams produced by the Oncor (R) linear accelerator (linac). The MC model of an 18MV photon beam was verified by measurement using a p-type diode dosimeter. Percent depth dose (PDD) and dose profiles were used for comparison based on three field sizes: 5x5, 10x10, and 20 x 20 cm(2). The accuracy of the CCC dosimetry was also evaluated using a plan composed of a simple parallel-opposed field (11 x 16 cm(2)) in a lung phantom comprised of four tissue simulating media namely, lung, soft tissue, bone and spinal cord. The CCC dose calculation accuracy was evaluated by MC simulation and measurements according to the dose difference and 3D gamma analysis. Gamma analysis was carried out through comparison of the Monte Carlo simulation and the TPS calculated dose. RESULTS: Compared to the dosimetric results measured by the Farmer chamber, the CCC algorithm underestimated dose in the planning target volume (PTV), right lung and lung-tissue interface regions by about -0.11%, -1.6 %, and -2.9%, respectively. Moreover, the CCC algorithm underestimated the dose at the PTV, right lung and lung-tissue interface regions in the order of -0.34%, -0.4% and -3.5%, respectively, when compared to the MC simulation. Gamma analysis results showed that the passing rates within the PTV and heterogeneous region were above 59% and 76%. For the right lung and spinal cord, the passing rates were above 80% for all gamma criteria. CONCLUSIONS: This study demonstrates that the CCC algorithm has potential to calculate dose with sufficient accuracy for 3D conformal radiotherapy within the thorax where a significant amount of tissue heterogeneity exists.
引用
收藏
页码:161 / 175
页数:15
相关论文
共 50 条
  • [1] Verification of lung dose in an anthropomorphic phantom calculated by the collapsed cone convolution method
    Butson, MJ
    Elferink, R
    Cheung, T
    Yu, PKN
    Stokes, M
    Quach, KY
    Metcalfe, P
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11): : N143 - N149
  • [2] Dosimetric and Monte Carlo verification of jaws-only IMRT plans calculated by the Collapsed Cone Convolution algorithm for head and neck cancers
    Duong Thanh Tai
    Luong Thi Oanh
    Nguyen Dong Son
    Truong Thi Hong Loan
    Chow, James C. L.
    [J]. REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2019, 24 (01) : 105 - 114
  • [3] STUDY OF THE DOSIMETRIC ACCURACY OF COLLAPSED CONE CONVOLUTION SUPERPOSITION ALGORITHM VERSUS MONTE CARLO SIMULATIONS FOR SMALL LUNG VOLUMES
    Calvo, O.
    Gutierrez, A.
    Stathakis, S.
    Rassiah-Szegedi, P.
    Papanikolaou, N.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S412 - S412
  • [4] Dosimetric Verification of collapsed cone convolution algorithm in heterogeneous media
    Guan, Yu-min
    Zhou, Ling-hong
    Zhen, Xin
    Lu, Wen-ting
    Zhang, Shu-xu
    [J]. 2010 4TH INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOMEDICAL ENGINEERING (ICBBE 2010), 2010,
  • [5] Comparison between collapsed cone convolution and Monte Carlo dose distributions for small lung lesions
    Liu, H.
    Rassiah-Szegedi, P.
    Stathakis, S.
    Shi, C.
    Papanikolaou, N.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2557 - 2557
  • [6] DOSIMETRIC PERFORMANCE OF IPLAN MONTE CARLO AND INITIAL COMPARISON TO THE COLLAPSED CONE ALGORITHM
    Kuenzler, T.
    Fotina, I.
    Stock, M.
    Georg, D.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S485 - S485
  • [7] Comparison of Collapsed Cone Convolution Superposition and Monte Carlo Dose Calculations for Spine SBRT Dose
    Parenica, H.
    Rembish, J.
    Saenz, D.
    Rasmussen, K.
    Kirby, N.
    Shi, Z.
    Li, Y.
    Fakhreddine, M.
    Crownover, R.
    Ha, C.
    Papanikolaou, N.
    Stathakis, S.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E163 - E164
  • [8] Comparative study of the collapsed cone convolution and Monte Carlo algorithms for radiation therapy planning of canine sinonasal tumors reveals significant dosimetric differences
    Lee, Ber-In
    Boss, Mary-Keara
    LaRue, Susan M.
    Martin, Tiffany
    Leary, Del
    [J]. VETERINARY RADIOLOGY & ULTRASOUND, 2022, 63 (01) : 91 - 101
  • [9] Dosimetric Evaluation of Collapsed Cone Convolution (CCC) Algorithm in Predicting Dose Beyond Air Gap
    Oyewale, S.
    Rana, S.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [10] Dosimetric test of collapsed cone convolution superposition algorithm in a heterogeneous media
    Oliveira e Silva, L. F.
    Mangueira, T. F.
    Dias, J. R.
    Avelino, S. R.
    [J]. EUROPEAN JOURNAL OF CANCER, 2013, 49 : S231 - S232