Nonsurgical placement of a nasolacrimal polyurethane stent: Long-term effectiveness

被引:56
|
作者
Song, HY
Jin, YH
Kim, JH
Suh, SW
Yoon, HK
Kang, SG
Sung, KB
机构
[1] UNIV ULSAN,COLL MED,ASAN MED CTR,DEPT OPHTHALMOL,SONGPA KU,SEOUL 138040,SOUTH KOREA
[2] UNIV ULSAN,COLL MED,ASAN MED CTR,DEPT OTORHINOLARYNGOL,SONGPA KU,SEOUL 138040,SOUTH KOREA
[3] ASAN INST LIFE SCI,MED MAT RES LAB,SEOUL,SOUTH KOREA
关键词
lacrimal gland and duct; interventional procedure; stents and prostheses;
D O I
10.1148/radiology.200.3.8756928
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate the longterm effectiveness of a polyurethane stent in the treatment of obstruction of the lacrimal sac and the nasolacrimal duct. MATERIALS AND METHODS: Fluoroscopically guided placement of a polyurethane nasolacrimal stent was evaluated in 283 obstructed lacrimal systems of 236 patients, with a follow-up period of more than 1 year (range, 52-134 weeks). The obstruction was at the junction between the lacrimal sac and the nasolacrimal duct in 192 systems, at the lacrimal sac in 52, and at the nasolacrimal duct in 39. The causes of obstruction were traumatic in 34 and idiopathic in 249 systems. RESULTS: Stent placement was technically successful in 270 systems (95%). At 7 days after stent placement, 235 (87%) of the 270 systems with successful placement demonstrated complete resolution of epiphora, 27 (10%) had partial resolution, and the remaining eight (3%) had no resolution. There was recurrence in 81 systems due to obstruction of the stent (n = 77) or obstruction of the common canaliculus (n = 4). The recurrence rate was much higher in the systems with obstruction at the lacrimal sac (64%) than in the systems with obstruction at the junction (26%) between the lacrimal sac and nasolacrimal duct or at the nasolacrimal duct (15%). CONCLUSION: Stent placement in the obstructed lacrimal system below the junction between the lacrimal sac and the nasolacrimal duct is valuable as an initial therapy.
引用
收藏
页码:759 / 763
页数:5
相关论文
共 50 条
  • [1] Nonsurgical placement of nasolacrimal polyurethane stent: Long-term effectiveness
    Song, H
    Lee, HK
    Yoon, H
    Goo, DE
    Sung, K
    RADIOLOGY, 1996, 201 : 1518 - 1518
  • [2] NONSURGICAL PLACEMENT OF A NASOLACRIMAL POLYURETHANE STENT
    SONG, HY
    JIN, YH
    KIM, JH
    HUH, SJ
    KIM, YH
    KIM, TH
    SUNG, KB
    RADIOLOGY, 1995, 194 (01) : 233 - 237
  • [3] Nasolacrimal polyurethane stent placement for epiphora:: Technical long-term results
    Pabón, IP
    Díaz, LP
    Grande, C
    López, MAD
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 12 (01) : 67 - 71
  • [4] Treatment of nasolacrimal duct obstruction with polyurethane stent placement:: Long-term results
    Yazici, Z
    Yazici, B
    Parlak, M
    Tuncel, E
    Ertürk, H
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (02) : 491 - 494
  • [5] Nonsurgical placement of a lacrimal polyurethane stent: Results of long-term follow-up
    Song, H
    Kang, S
    Lee, DH
    Ko, GY
    Lee, K
    Sung, K
    RADIOLOGY, 2000, 217 : 551 - 552
  • [6] Nasolacrimal polyurethane stent placement: preliminary results
    Perena, MF
    Castillo, J
    Medrano, J
    De Gregorio, MA
    Loras, E
    Cristobal, JA
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2001, 11 (01) : 25 - 30
  • [7] LONG-TERM PLACEMENT OF SILASTIC NASOLACRIMAL TUBES
    VELOUDIOS, A
    HARVEY, JT
    PHILIPPON, M
    OPHTHALMIC SURGERY AND LASERS, 1991, 22 (04): : 225 - 227
  • [8] The polyurethane nasolacrimal duct stent for lower tear duct obstruction: long-term success rate and complications
    U. Schaudig
    R. Maas
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2000, 238 : 733 - 737
  • [9] The polyurethane nasolacrimal duct stent for lower tear duct obstruction: long-term success rate and complications
    Schaudig, U
    Maas, R
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2000, 238 (09) : 733 - 737
  • [10] FEMOROPOPLITEAL STENT PLACEMENT - LONG-TERM RESULTS
    SAPOVAL, MR
    LONG, AL
    RAYNAUD, AC
    BEYSSEN, BM
    FIESSINGER, JN
    GAUX, JC
    RADIOLOGY, 1992, 184 (03) : 833 - 839