On the need for tuning the dosimetric leaf gap for stereotactic treatment plans in the Eclipse treatment planning system

被引:25
|
作者
Vieillevigne, Laure [1 ,2 ]
Khamphan, Catherine [3 ]
Saez, Jordi [4 ]
Hernandez, Victor [5 ]
机构
[1] Inst Univ Canc Toulouse, Dept Med Phys, Inst Claudius Regaud, Toulouse, France
[2] Univ Toulouse 3, CNRS Oncopole, Ctr Rech & Cancerol Toulouse, INSERM,ERL5294,UMR1037, Toulouse, France
[3] Inst St Catherine, Med Phys Unit, Avignon, France
[4] Hosp Clin Barcelona, Dept Radiat Oncol, Barcelona, Spain
[5] IISPV, St Joan Reus, Dept Med Phys Hosp, Tarragona, Spain
来源
关键词
dosimetric leaf gap; dynamic conformal arc; stereotactic treatments; tongue-and-groove modeling; volumetric modulated arc therapy; MODULATED ARC THERAPY; BODY RADIOTHERAPY; GUIDELINE; MLC; ACCURACY; QA; IMPLEMENTATION; RADIOSURGERY; PERFORMANCE; LESIONS;
D O I
10.1002/acm2.12656
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The dosimetric leaf gap (DLG) and tongue-and-groove (T&G) effects are critical aspects in the modeling of multileaf collimators (MLC) in the treatment planning system (TPS). In this study, we investigated the dosimetric impact of limitations associated with the T&G modeling in stereotactic plans and its relationship with the need for tuning the DLG in the Eclipse TPS. Measurements were carried out using Varian TrueBeam STx systems from two different institutions. Test fields presenting MLC patterns with several MLC gap sizes (meanGap) and different amounts of T&G effect (TGi) were first evaluated. Secondly, dynamic conformal arc (DCA) and volumetric modulated arc therapy (VMAT) deliveries of stereotactic cases were analyzed in terms of meanGap and TGi. Two DLG values were used in the TPS: the measured DLG (DLG(meas)) and an optimal DLG (DLG(opt)). Measured and calculated doses were compared according to dose differences and gamma passing rates (GPR) with strict local gamma criteria of 2%/2 mm. The discrepancies were analyzed for DLG(meas) and DLG(opt), and their relationships with both TGi and meanGap were investigated. DCA arcs involved significantly lower TGi and larger meanGap than VMAT arcs (p < 0.0001). By using DLG(meas) in the TPS, the dose discrepancies increased as TGi increased and meanGap decreased for both test fields and clinical plans. Dose discrepancies dramatically increased with the ratio TGi/meanGap. Adjusting the DLG value was then required to achieve acceptable calculations and configuring the TPS with DLG(opt) led to an excellent agreement with median GPRs (2%/2 mm) > 99% for both institutions. We also showed that DLG(opt) could be obtained from the results of the test fields. We demonstrated that the need for tuning the DLG is due to the limitations of the T&G modeling in the Eclipse TPS. A set of sweeping gap tests modified to incorporate T&G effects can be used to determine the optimal DLG value.
引用
收藏
页码:68 / 77
页数:10
相关论文
共 50 条
  • [1] A Novel Method to Automate the Optimization of Multileaf Collimator Transmission and Dosimetric Leaf Gap Parameters for Eclipse Treatment Planning System
    DiCostanzo, D.
    Ayan, A.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E571 - E571
  • [2] Effects of the dosimetric leaf gap in treatment planning system on the agreement between predicted and measured doses
    Cui, Congwu
    Wu, Xia
    [J]. MEDICAL PHYSICS, 2017, 44 (08) : 4389 - 4389
  • [3] Dosimetric Comparison of Multiplan and Eclipse Planning Systems for the Stereotactic Body Radiotherapy Treatment of Prostate Cancer
    Li, Fang
    Ozhasoglu, C.
    Clump, D.
    Heron, D.
    Lalonde, R.
    Huq, M.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2961 - 2961
  • [4] Dosimetric Verification of stereotactic Radiation Treatment Planning
    Senn, A. M.
    Garbe, S.
    Muedder, T.
    Schoroth, F.
    Roehner, F.
    Koch, D.
    Schmeel, C.
    Wilhelm-Buchstab, T.
    Leitzen, C.
    Heimann, M.
    Tschirner, S.
    Buecheler, B.
    Baumert, B. G.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2019, 195 : S70 - S70
  • [5] A Systematic Methodology in Dosimetric Verification of Anisotropic Analytical Algorithm in Eclipse Treatment Planning System
    He, R.
    Yang, C.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [6] Comparison of two algorithms for leaf motion calculation in eclipse treatment planning system
    Seoane, A.
    Sanchez-Artunedo, D.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S766 - S766
  • [7] Investigation of the Accuracy of Different Lung SBRT Plans Created in Eclipse Treatment Planning System
    Yildirim, S.
    Cebe, M.
    Mabhouti, H.
    Pacaci, P.
    Serin, E.
    Sanli, E.
    Esitmez, D.
    Kucuk, N.
    Doyuran, M.
    Kucukmorkoc, E.
    Caglar, M.
    Canoglu, D.
    Rzazade, R.
    Acar, H.
    Ozkok, H. Caglar
    [J]. MEDICAL PHYSICS, 2017, 44 (06)
  • [8] Optimizing the effective spot size and the dosimetric leaf gap of the AcurosXB algorithm for VMAT treatment planning
    Passal, V.
    Barreau, M.
    Tiplica, T.
    Dufreneix, S.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (06): : 154 - 161
  • [9] Dosimetric quality control of Eclipse treatment planning system using pelvic digital test object
    Benhdech, Yassine
    Beaumont, Stephane
    Guedon, Jeanpierre
    Crespin, Sylvain
    [J]. MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [10] Dosimetric Uncertainties and Their Impact on Treatment Planning in Stereotactic Proton Radiosurgery
    Geng, C.
    Daartz, J.
    Cheung, K. Lam Tin
    Bussiere, M. R.
    Shih, H. A.
    Paganetti, H.
    Schuemann, J. P.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E618 - E618