Evaluation Detector Module of the miniPET-3 Small Animal PET Scanner

被引:0
|
作者
Imrek, J. [1 ]
Hegyesi, Gy. [1 ]
Kalinka, G. [1 ]
Kiraly, B. [1 ]
Molnar, J. [1 ]
Nagy, F. [1 ]
Valastyan, I. [1 ]
Balkay, L. [2 ]
Szabo, Zs. [3 ]
机构
[1] Hungarian Acad Sci, Inst Nucl Res, Debrecen, Hungary
[2] Univ Debrecen, Univ Med Sch Debrecen, PET Ctr, Debrecen, Hungary
[3] Univ Debrecen, Inst Expt Phys, Debrecen, Hungary
关键词
LySO; SiPM; modular readout; Spartan-6; FPGA; miniPET-3;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We report on the development of an evaluation detector module built for our miniPET-3 small animal PET scanner. The module is based on an LySO scintillation matrix exactly matching that used in miniPET-II ( so that the performance of the two PET scanners is easier to compare), followed by a Silicon Photomultiplier (SiPM) matrix array. The frontend electronics is designed to be modular, so that different readout concepts can be evaluated and prototype-tested. The readout electronics is based on a system-on-module with Spartan-6 FPGA and Gigabit Ethernet connectivity. Different SiPM readout concepts were tested using the module, and based on the evaluation results one was selected for implementation and testing.
引用
收藏
页码:3790 / 3793
页数:4
相关论文
共 50 条
  • [1] Experimental scanner setup from miniPET II detector module
    Valastyan, I.
    Imrek, J.
    Hegyesi, Gy.
    Kalinka, G.
    Molnar, J.
    Novak, D.
    Sipos, A.
    Bagamery, I.
    Balkay, L.
    Kerek, A.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3121 - +
  • [2] Performance characteristics of a miniPET scanner dedicated to small animal imaging
    Kis, S. A.
    Valastyan, I.
    Hegyesi, Gy.
    Imrek, J.
    Kalinka, G.
    Molnar, J.
    Novak, D.
    Vegh, J.
    Balkay, L.
    Emri, M.
    Molnar, G.
    Bagamery, I.
    Bukki, T.
    Rozsa, S.
    Szabo, Zs.
    Kerek, A.
    Tron, L.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 1645 - 1648
  • [3] Performance evaluation of the 32-module quadHIDAC small-animal PET scanner
    Schäfers, KP
    Reader, AJ
    Kriens, M
    Knoess, C
    Schober, O
    Schäfers, M
    JOURNAL OF NUCLEAR MEDICINE, 2005, 46 (06) : 996 - 1004
  • [4] Evaluation of scatter effects in phoswich detector modules of a dedicated small animal PET scanner
    Yu, Jianhua
    Pomper, Martin
    Tsui, Benjamin
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [5] Initial evaluation of the Indiana small animal PET scanner
    Rouze, Ned C.
    Soon, Victor C.
    Young, John W.
    Siegel, Stefan
    Hutchins, Gary D.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2394 - 2398
  • [6] A high performance phoswich detector module for small animal PET.
    Lee, IJ
    Vaquero, JJ
    Barbosa, FJ
    Seidel, J
    Green, MV
    JOURNAL OF NUCLEAR MEDICINE, 2000, 41 (05) : 19P - 20P
  • [7] Performance evaluation of the Philips MOSAIC small animal PET scanner
    Huisman, Marc C.
    Reder, Sybille
    Weber, Axel W.
    Ziegler, Sibylle I.
    Schwaiger, Markus
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 (04) : 532 - 540
  • [8] Performance evaluation of the Philips MOSAIC small animal PET scanner
    Marc C. Huisman
    Sybille Reder
    Axel W. Weber
    Sibylle I. Ziegler
    Markus Schwaiger
    European Journal of Nuclear Medicine and Molecular Imaging, 2007, 34 : 532 - 540
  • [9] A Wearable Small Animal PET Scanner
    Yu, Xin
    Liu, Han
    Zhu, Wenjing
    Liang, Da
    Zhao, Huiping
    Zhu, Zhiliang
    Tao, Jinyong
    Zeng, Jiayang
    Zhang, Yibin
    Xu Jianfeng
    Xie, Siwei
    Peng, Qiyu
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [10] Initial performance evaluation of a.modular, large-area detector PET scanner for small animal imaging
    Kao, Chien-Min
    Souris, Jeffrey S.
    Cho, Seungryong
    Penney, Bill C.
    Chen, Chin-Tu
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2081 - 2084