Characterization and reduction of artifacts in limited angle tomography

被引:130
|
作者
Frikel, Juergen [1 ,2 ]
Quinto, Eric Todd [3 ]
机构
[1] Tech Univ Munich, Zentrum Math, D-80290 Munich, Germany
[2] Helmholtz Zentrum Munchen, Inst Computat Biol, Munich, Germany
[3] Tufts Univ, Dept Math, Medford, MA 02155 USA
基金
美国国家科学基金会;
关键词
X-RAY TRANSFORM; LOCAL TOMOGRAPHY; IMAGE-RECONSTRUCTION; TOMOSYNTHESIS; INVERSION; SPACE;
D O I
10.1088/0266-5611/29/12/125007
中图分类号
O29 [应用数学];
学科分类号
070104 ;
摘要
We consider the reconstruction problem for limited angle tomography using filtered backprojection (FBP) and lambda tomography. We use microlocal analysis to explain why the well-known streak artifacts are present at the end of the limited angular range. We explain how to mitigate the streaks and prove that our modified FBP and lambda operators are standard pseudodifferential operators, and so they do not add artifacts. We provide reconstructions to illustrate our mathematical results.
引用
收藏
页数:21
相关论文
共 50 条
  • [1] How strong are streak artifacts in limited angle computed tomography?
    Nguyen, Linh V.
    INVERSE PROBLEMS, 2015, 31 (05)
  • [2] Mathematical Reduction of Artifacts in Limited Projection and Limited Angle Cone Beam CT
    Wei, Y.
    Gersh, J.
    Mckee, M.
    Baydush, A.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [3] Limited-Angle Computer Tomography with Truncated Projection Artifacts Removal
    Shen, Enxiang
    Wang, Yuxin
    Yuan, Jie
    Carson, Paul L.
    APPLIED SCIENCES-BASEL, 2022, 12 (22):
  • [4] DualRes-UNet: Limited Angle Artifact Reduction for Computed Tomography
    Zhang, Tao
    Gao, Hewei
    Xing, Yuxiang
    Chen, Zhiqiang
    Zhang, Li
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [5] Data Consistent Artifact Reduction for Limited Angle Tomography with Deep Learning Prior
    Huang, Yixing
    Preuhs, Alexander
    Lauritsch, Gunter
    Manhart, Michael
    Huang, Xiaolin
    Maier, Andreas
    MACHINE LEARNING FOR MEDICAL IMAGE RECONSTRUCTION, MLMIR 2019, 2019, 11905 : 101 - 112
  • [6] Reduction of Ring Artifacts in Computer Tomography
    Laskov V.V.
    Simonov E.N.
    Biomedical Engineering, 2016, 49 (5) : 274 - 277
  • [7] Iodine quantification in limited angle tomography
    Michielsen, Koen
    Rodriguez-Ruiz, Alejandro
    Reiser, Ingrid
    Nagy, James G.
    Sechopoulos, Ioannis
    MEDICAL PHYSICS, 2020, 47 (10) : 4906 - 4916
  • [8] Sparse regularization in limited angle tomography
    Frikel, Juergen
    APPLIED AND COMPUTATIONAL HARMONIC ANALYSIS, 2013, 34 (01) : 117 - 141
  • [9] Linear extrapolator for limited angle tomography
    Yau, Sze Fong
    Wong, Shun Him
    Biomedical Engineering - Applications, Basis and Communications, 1996, 8 (03): : 249 - 260
  • [10] Limited angle tomography and ridge functions
    Kazantsev, IG
    Matej, S
    Lewitt, RA
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1706 - 1710