Pulsed Electron Source for Atmospheric Pressure Chemical Ionization in Ion Mobility Spectrometry

被引:0
|
作者
Bunert, Erik [1 ]
Heptner, Andre [1 ]
Kirk, Ansgar T. [1 ]
Kaebein, Oliver [1 ]
Zimmermann, Stefan [1 ]
Kusch, Alexander [2 ]
Wurz, Marc Christopher [2 ]
机构
[1] Leibniz Univ Hannover, Inst Elect Engn & Measurement Technol, Hannover, Germany
[2] Leibniz Univ Hannover, Inst Micro Prod Technol, Hannover, Germany
关键词
non-radioacitve electron source; ion mobility spectrometry; APCI; pulsed electron emission; thermionic emission; field emission; emission current control; ENERGY;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Ion mobility spectrometers (IMS) are measurement devices for fast and ultra-sensitive trace gas analysis. Most IMS employ radioactive electron sources, such as H-3 or Ni-63, to provide free electrons with high kinetic energy at atmospheric pressure for initiating a chemical gas phase ionization of the analytes. The disadvantage of these radioactive materials are legal restrictions and the electron emission cannot be adjusted or turned off. Therefore, we developed a non-radioactive electron source and replaced the H-3-source of our existing IMS, leading to comparable spectra. An advantage of our non-radioactive electron source is that it can operate in a fast pulsed mode. By optimizing the geometric parameters and developing fast control electronics, we can achieve short electron emission pulses with high intensities and adjustable pulse width down to a few nanoseconds. This allows to control the ionization process, which can enhance the analytical performance of the IMS. Furthermore, a miniaturized non-radioactive electron source is desirable, e.g. for hand-held IMS devices. Therefore, we developed an emission current control for field emitter cathodes and investigated their suitability for this application.
引用
收藏
页码:102 / 103
页数:2
相关论文
共 50 条
  • [1] Atmospheric pressure chemical ionization of fluorinated phenols in atmospheric pressure chemical ionization mass spectrometry, tandem mass spectrometry, and ion mobility spectrometry
    Eiceman, GA
    Bergloff, JF
    Rodriguez, JE
    Munro, W
    Karpas, Z
    [J]. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1999, 10 (11) : 1157 - 1165
  • [2] Using dopants in the atmospheric pressure chemical ionization ion source to determine the site of protonation by ion mobility spectrometry
    Valadbeigi, Younes
    Mirzahosseini, Fatemeh
    Ilbeigi, Vahideh
    Matejcik, Stefan
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2024, 38 (18)
  • [3] Formation of anionic clusters of triols in the atmospheric pressure chemical ionization source: An ion mobility spectrometry and theoretical study
    Valadbeigi, Younes
    Ilbeigi, Vahideh
    Khodabandeh, Zahra
    [J]. CHEMICAL PHYSICS LETTERS, 2021, 777 (777)
  • [4] Study of atmospheric pressure chemical ionization of PETN by ion mobility spectrometry/tandem mass spectrometry
    A. A. Filipenko
    E. K. Malkin
    [J]. Journal of Analytical Chemistry, 2011, 66 : 1464 - 1469
  • [5] Study of atmospheric pressure chemical ionization of phthalates in air by ion mobility spectrometry/mass spectrometry
    Moravsky, Ladislav
    Michalczuk, Bartosz
    Fateh Borkhari, Arian
    Papp, Peter
    Sysoev, Alexey A.
    Matejcik, Stefan
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2021, 35 (17)
  • [6] Study of atmospheric pressure chemical ionization of PETN by ion mobility spectrometry/tandem mass spectrometry
    Filipenko, A. A.
    Malkin, E. K.
    [J]. JOURNAL OF ANALYTICAL CHEMISTRY, 2011, 66 (14) : 1464 - 1469
  • [7] Characterization of Triacetone Triperoxide by Ion Mobility Spectrometry and Mass Spectrometry Following Atmospheric Pressure Chemical Ionization
    Ewing, Robert G.
    Waltman, Melanie J.
    Atkinson, David A.
    [J]. ANALYTICAL CHEMISTRY, 2011, 83 (12) : 4838 - 4844
  • [8] Dopant-enhanced atmospheric pressure glow discharge ionization source for ion mobility spectrometry
    Dong, Can
    Wang, Lina
    Wang, Weiguo
    Hou, Keyou
    Li, Haiyang
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (10):
  • [9] Atmospheric pressure ionization of chlorinated ethanes in ion mobility spectrometry and mass spectrometry
    Ewing, Robert G.
    Atkinson, David A.
    Benson, Michael T.
    [J]. INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2015, 18 (1-2) : 51 - 58
  • [10] Application of a Nonradioactive Pulsed Electron Source for Ion Mobility Spectrometry
    Gunzer, Frank
    Zimmermann, Stefan
    Baether, Wolfgang
    [J]. ANALYTICAL CHEMISTRY, 2010, 82 (09) : 3756 - 3763