Anterior Segment Optical Coherence Tomography Versus Ultrasound Pachymetry to Measure Corneal Thickness in Endothelial Keratoplasty Donor Corneas
被引:9
|作者:
Fante, Ryan J.
[1
]
Shtein, Roni M.
[1
]
Titus, Michael S.
[2
]
Woodward, Maria A.
[1
]
机构:
[1] Univ Michigan, Dept Ophthalmol & Visual Sci, Ann Arbor, MI 48105 USA
[2] Midwest Eye Banks, Ann Arbor, MI USA
来源:
关键词:
optical coherence tomography;
ultrasound pachymetry;
Descemet stripping automated endothelial keratoplasty;
corneal thickness;
GRAFT THICKNESS;
KERATOCONUS;
D O I:
10.1097/ICO.0b013e31827c0b5b
中图分类号:
R77 [眼科学];
学科分类号:
100212 ;
摘要:
Purpose: To analyze corneal thickness (CT) by anterior segment optical coherence tomography and ultrasound pachymetry in precut endothelial keratoplasty (EK) donor corneas at the eye bank. Methods: Thirty-six corneas were analyzed. All corneas were dissected to create lamellar grafts for EK using the standard eye bank protocol. We measured central and peripheral (3 mm from center) CT by ultrasound after lamellar tissue processing and by optical coherence tomography after the tissue was transferred to a viewing chamber. We performed paired t tests and Pearson correlation analyses to compare ultrasound and optical coherence tomography measurements. Results: Central CT measurements by optical coherence tomography versus ultrasound in donor corneas were not statistically different (mean: 169 vs. 177 mu m, respectively, P = 0.09). There was a strong correlation between measurements of central CT (r = 0.73, P, < .0001). There was a statistically significant difference in mean peripheral CT measurements by optical coherence tomography and ultrasound (mean: 190 vs. 236 mu m, respectively, P < 0.0001); at peripheral locations, measurements correlated moderately (r = 0.52, P = 0.002). Conclusions: Optical coherence tomography measurements of EK-prepared tissue averaged 8 mm thinner than the ultrasound measurements, similar to clinical measurements of central CT. Given the growing interest in the effect of EK graft thickness on patient outcomes, it is important to understand the differences in measurements obtained by different devices used to assess donor corneas.
引用
收藏
页码:E79 / E82
页数:4