Development and Characterization of a New Cylindrical Ionization Chamber for Dosimetry of 60Co Beams

被引:2
|
作者
Neves, Lucio P. [1 ]
Perini, Ana P. [1 ]
Caldas, Linda V. E. [1 ]
机构
[1] Comissao Nacl Energia Nucl IPEN CNEN SP, Inst Pesquisas Energet & Nucl, BR-05508000 Sao Paulo, Brazil
关键词
Gamma beam dosimetry; ionization chamber; radiotherapy;
D O I
10.1109/TNS.2012.2219319
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A cavity ionization chamber, made of PVC coated with graphite, was developed and tested to verify its applicability as a reference dosimeter for Co-60 beams. In order to characterize this new ionization chamber, several tests were made, following international recommendations, as saturation, ion collection efficiency, polarity effect, short- and medium-term stabilities, linearity of response, angular dependence and leakage current. All results obtained were within those recommended internationally, showing that this prototype may be used as a dosimetric standard system at the Calibration Laboratory of IPEN.
引用
收藏
页码:712 / 715
页数:4
相关论文
共 50 条
  • [1] A new standard cylindrical graphite-walled ionization chamber for dosimetry in 60Co beams at calibration laboratories
    Neves, Lucio P.
    Perini, Ana P.
    Caldas, Linda V. E.
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2014, 104 : 235 - 239
  • [2] Development and characterization of a graphite-walled ionization chamber as a reference dosimeter for 60Co beams
    Perini, Ana P.
    Neves, Lucio P.
    Caldas, Linda V. E.
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2014, 104 : 248 - 251
  • [3] Determination of the Kwall correction factor for a cylindrical ionization chamber to measure air-kerma in 60Co gamma beams
    Laitano, RF
    Toni, MP
    Pimpinella, M
    Bovi, M
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (14): : 2411 - 2431
  • [4] Ionization chamber dosimetry based on 60Co absorbed dose to water calibration for diagnostic kilovoltage x-ray beams
    Araki, Fujio
    Ohno, Takeshi
    Umeno, Suzuna
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (18):
  • [5] A prototype of an ionization chamber for gamma radiation beams of 60Co: Experimental and Monte Carlo preliminary results
    Vahabi, Seyed Milad
    Zafarghandi, Mojtaba Shamsaie
    Ghasemi, Mohammadreza
    Alipoor, Anita
    Shahvar, Arjang
    [J]. RADIATION MEASUREMENTS, 2013, 59 : 284 - 287
  • [6] Experimental and Monte Carlo evaluation of an ionization chamber in a 60Co beam
    Perini, A. P.
    Neves, L. P.
    Santos, W. S.
    Caldas, L. V. E.
    [J]. 8TH BRAZILIAN CONGRESS ON METROLOGY (METROLOGIA 2015), 2016, 733
  • [7] Experimental investigation of the effect of air cavity size in cylindrical ionization chambers on the measurements in 60Co radiotherapy beams
    Swanpalmer, John
    Johansson, Karl-Axel
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (22): : 7093 - 7107
  • [8] Optically stimulated luminescence in vivo dosimetry for radiotherapy: physical characterization and clinical measurements in 60Co beams
    Mrcela, I.
    Bokulie, T.
    Izewska, J.
    Budanec, M.
    Froebe, A.
    Kusic, Z.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (18): : 6065 - 6082
  • [9] FERROUS SULFATE DOSIMETER FOR CONTROL OF IONIZATION-CHAMBER DOSIMETRY OF ELECTRON AND CO-60 GAMMA-BEAMS
    MATTSSON, LO
    JOHANSSON, KA
    SVENSSON, H
    [J]. ACTA RADIOLOGICA ONCOLOGY, 1982, 21 (02): : 139 - 144
  • [10] A new parallel-plate graphite ionization chamber as a 60Co gamma radiation reference instrument
    Perini, Ana P.
    Neves, Lucio P.
    Fernandez-Varea, Jose M.
    Cassola, Vagner F.
    Kramer, Richard
    Khoury, Helen J.
    Caldas, Linda V. E.
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2014, 95 : 106 - 108