Top-down and bottom-up approaches for the estimation of measurement uncertainty in coagulation assays

被引:7
|
作者
Lim, Yong Kwan [1 ]
Kweon, Oh Joo [1 ]
Lee, Mi-Kyung [1 ]
Kim, Bohyun [2 ]
Kim, Hye Ryoun [1 ]
机构
[1] Chung Ang Univ, Coll Med, Dept Lab Med, 102 Heukseok Ro, Seoul 06973, South Korea
[2] Soonchunhyang Univ, Cheonan Hosp, Dept Lab Med, Cheonan, South Korea
基金
新加坡国家研究基金会;
关键词
bias; blood coagulation; clinical laboratory tests; laboratory proficiency testing; measurement uncertainty; INTERNATIONAL SENSITIVITY INDEX; PROTHROMBIN TIME; QUALITY-CONTROL; STANDARDIZATION; LABORATORIES; ANALYTES;
D O I
10.1515/cclm-2020-0038
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The assessment of measurement uncertainty (MU) in clinical laboratories is essential to the reliable interpretation of results in clinical laboratories. However, despite the introduction of various methods for the expression of uncertainty in measurement, the MUs of coagulation tests have not been extensively studied. The aim of this study was to quantify the MU of various coagulation assays according to international guidelines and to report an expected confidence in the quality of coagulation assays. Methods: We selected activated partial thromboplastin time, international normalized ratio (INR), protein C/S, antithrombin, fibrinogen, and Factor V/VIII/X to quantify the MUs of two coagulation testing systems: ACL TOP 750 CTS (Instrumentation Laboratory, Bedford, MA, USA) and STA Compact (Diagnostica Stago, Asnieres-sur-Seine, France). We used international standards and interlaboratory comparison results in accordance with international guidelines in a top-down approach to the assessment of MU. For INR, MU was estimated in a bottom-up approach using reference thromboplastin and certified plasmas. Results: Top-down approaches resulted in MUs between 3.3% and 21.3% for each measurand. In the bottom-up approach, MUs of INR values ranged from 10.9% to 26.4% and showed an upward trend as INR increased. Conclusions: In this study, we were successful in quantifying MU of coagulation assays using practical methods. Our results demonstrated that top-down and bottom-up approaches were adequate for coagulation assays. However, some assays showed significant biases against international standards; therefore, standardization would be necessary to ensure more reliable patient results.
引用
收藏
页码:1525 / 1533
页数:9
相关论文
共 50 条
  • [1] Comparison between bottom-up and top-down approaches in the estimation of measurement uncertainty
    Lee, Jun Hyung
    Choi, Jee-Hye
    Youn, Jae Saeng
    Cha, Young Joo
    Song, Woonheung
    Park, Ae Ja
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (07) : 1025 - 1032
  • [2] Sustainability and uncertainty: Bottom-up and top-down approaches
    Jensen, K. Klint
    [J]. ITALIAN JOURNAL OF ANIMAL SCIENCE, 2007, 6 : 853 - 855
  • [3] Comparison between top-down and bottom-up approaches in the estimation of measurement uncertainty in Bisphenol A analysis by HPLC-FLD
    Kilincer, Mehmet
    Ozyurek, Mustafa
    [J]. JOURNAL OF CHEMICAL METROLOGY, 2023, 17 (02): : 225 - 237
  • [4] Bottom-up and top-down uncertainty quantification for measurements
    Burr, T.
    Croft, S.
    Favalli, A.
    Krieger, T.
    Weaver, B.
    [J]. CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2021, 211
  • [5] Top-down versus bottom-up approaches in proteomics
    Wehr, Tim
    [J]. LC GC NORTH AMERICA, 2006, 24 (09) : 1004 - +
  • [6] Top-Down and Bottom-Up Approaches to Company Competitiveness
    Kotik, Vitezslav
    Voracek, Jan
    [J]. 15TH INTERNATIONAL FORUM ON KNOWLEDGE ASSET DYNAMICS (IFKAD 2020): KNOWLEDGE IN DIGITAL AGE, 2020, : 100 - 119
  • [7] General modelling approaches: Top-down or bottom-up?
    Worsley, TE
    Harris, RCE
    [J]. NATIONAL TRANSPORT MODELS: RECENT DEVELOPMENTS AND PROSPECTS, 2002, : 57 - 60
  • [8] TOP-DOWN AND BOTTOM-UP APPROACHES TO PROJECT SELECTION
    不详
    [J]. RESEARCH MANAGEMENT, 1978, 21 (02): : 22 - 24
  • [9] Bottom-up and top-down measurement uncertainty evaluation for multivariate spectrophotometric procedures
    Couto, Aldo Renato
    Lourenco, Felipe Rebello
    [J]. MICROCHEMICAL JOURNAL, 2023, 193
  • [10] Bottom-up or top-down in dream neuroscience? A top-down critique of two bottom-up studies
    Foulkes, David
    Domhoff, G. William
    [J]. CONSCIOUSNESS AND COGNITION, 2014, 27 : 168 - 171