Dosimetric evaluation of the interplay effect in respiratory-gated RapidArc radiation therapy

被引:37
|
作者
Riley, Craig [1 ]
Yang, Yong [2 ]
Li, Tianfang [2 ]
Zhang, Yongqian [2 ]
Heron, Dwight E. [2 ]
Huq, M. Saiful [2 ]
机构
[1] Univ Pittsburgh, Dept Radiat Oncol, Sch Med, Pittsburgh, PA 15232 USA
[2] Univ Pittsburgh, Dept Radiat Oncol, Inst Canc, Pittsburgh, PA 15232 USA
关键词
gating; RapidArc; interplay effect; VMAT; VOLUMETRIC MODULATED ARC; DOSE DELIVERY; ORGAN MOTION; TUMOR MOTION; IMRT; RADIOTHERAPY; PHANTOM;
D O I
10.1118/1.4855956
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Volumetric modulated arc therapy (VMAT) with gating capability has had increasing adoption in many clinics in the United States. In this new technique, dose rate, gantry rotation speed, and the leaf motion speed of multileaf collimators (MLCs) are modulated dynamically during gated beam delivery to achieve highly conformal dose coverage of the target and normal tissue sparing. Compared with the traditional gated intensity-modulated radiation therapy technique, this complicated beam delivery technique may result in larger dose errors due to the intrafraction tumor motion. The purpose of this work is to evaluate the dosimetric influence of the interplay effect for the respiration-gated VMAT technique (RapidArc, Varian Medical Systems, Palo Alto, CA). Our work consisted of two parts: (1) Investigate the interplay effect for different target residual errors during gated RapidArc delivery using a one-dimensional moving phantom capable of producing stable sinusoidal movement; (2) Evaluate the dosimetric influence in ten clinical patients' treatment plans using a moving phantom driven with a patient-specific respiratory curve. Methods: For the first part of this study, four plans were created with a spherical target for varying residual motion of 0.25, 0.5, 0.75, and 1.0 cm. Appropriate gating windows were applied for each. The dosimetric effect was evaluated using EDR2 film by comparing the gated delivery with static delivery. For the second part of the project, ten gated lung stereotactic body radiotherapy cases were selected and reoptimized to be delivered by the gated RapidArc technique. These plans were delivered to a phantom, and again the gated treatments were compared to static deliveries by the same methods. Results: For regular sinusoidal motion, the dose delivered to the target was not substantially affected by the gating windows when evaluated with the gamma statistics, suggesting the interplay effect has a small role in respiratory-gated RapidArc therapy. Varied results were seen when gated therapy was performed on the patient plans that could only be attributed to differences in patient respiratory patterns. Patients whose plans had the largest percentage of pixels failing the gamma statistics exhibited irregular breathing patterns including substantial interpatient variation in depth of respiration. Conclusions: The interplay effect has a limited impact on gated RapidArc therapy when evaluated with a linear phantom. Variations in patient breathing patterns, however, are of much greater clinical significance. Caution must be taken when evaluating patients' respiratory efforts for gated arc therapy. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Dosimetric evaluations of the interplay effect in respiratory-gated intensity-modulated radiation therapy
    Chen, Hungcheng
    Wu, Andrew
    Brandner, Edward D.
    Heron, Dwight E.
    Huq, M. Saiful
    Yue, Ning J.
    Chen, Wen-Cheng
    MEDICAL PHYSICS, 2009, 36 (03) : 893 - 903
  • [2] The Dosimetric and Temporal Effects of Respiratory-Gated Radiation Therapy in Lung Cancer Patients
    Rouabhi, O.
    Gross, B.
    Bayouth, J.
    Xia, J.
    MEDICAL PHYSICS, 2015, 42 (06) : 3304 - 3304
  • [3] Pre-clinical evaluation of respiratory-gated delivery of volumetric modulated arc therapy with RapidArc
    Nicolini, Giorgia
    Vanetti, Eugenio
    Clivio, Alessandro
    Fogliata, Antonella
    Cozzi, Luca
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (12): : N347 - N357
  • [4] Response to the "Comments on 'Dosimetric evaluations of the interplay effect in respiratory-gated intensity-modulated radiation therapy'" [Med. Phys. 36, 2340 (2009)]
    Chen, Hungcheng
    Wu, Andrew
    Brandner, Edward D.
    Heron, Dwight E.
    Huq, M. Saiful
    Yue, Ning J.
    Chen, Wen-Cheng
    MEDICAL PHYSICS, 2009, 36 (06) : 2341 - 2342
  • [5] RESPIRATORY GATED RADIATION THERAPY: A DOSIMETRIC ANALYSIS
    Righetto, R.
    Valenti, M.
    Mantello, G.
    Cardinali, M.
    Maggi, S.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S477 - S477
  • [6] Dosimetric Evaluation of Respiratory-Gated VMAT On a Versa HD Linear Accelerator
    Snyder, J.
    Flynn, R.
    Hyer, D.
    MEDICAL PHYSICS, 2017, 44 (06) : 2896 - 2896
  • [7] Dosimetric Evaluation of Beam-Hold Interruption in Respiratory Gated RapidArc Delivery
    Shen, S.
    Popple, R.
    Duan, J.
    Wu, X.
    Brezovich, I.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [8] Comment on "Dosimetric evaluations of the interplay effect in respiratory-gated intensity-modulated radiation therapy" [Med. Phys. 36, 893-903 (2009)]
    Litzenberg, D. W.
    Moran, J. M.
    Fraass, B. A.
    MEDICAL PHYSICS, 2009, 36 (06) : 2340 - 2340
  • [9] Evaluation of Breathing Patterns for Respiratory-Gated Radiation Therapy Using Respiration Regularity Index
    Cheong, K.
    Lee, M.
    Kang, S.
    Yoon, J.
    Park, S.
    Hwang, T.
    Kim, H.
    Kim, K.
    Han, T.
    Bae, H.
    MEDICAL PHYSICS, 2014, 41 (06) : 170 - 171
  • [10] Evaluation of breathing patterns for respiratory-gated radiation therapy using the respiration regularity index
    Kwang-Ho Cheong
    MeYeon Lee
    Sei-Kwon Kang
    Jai-Woong Yoon
    SoAh Park
    Taejin Hwang
    Haeyoung Kim
    KyoungJu Kim
    Tae Jin Han
    Hoonsik Bae
    Journal of the Korean Physical Society, 2015, 66 : 301 - 313