Static structural testing of trans-tibial composite sockets

被引:0
|
作者
Current, TA [1 ]
Kogler, GF [1 ]
Barth, DG [1 ]
机构
[1] So Illinois Univ, Sch Med, Orthopaed Bioengn Res Lab, Springfield, IL 62794 USA
关键词
D O I
暂无
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The purpose of this investigation was to quantify the structural strength of various trans-tibial composite sockets. To conduct the study, loading parameters and methods were developed that emulate the International Standards Organisation (ISO) standards for structural testing of lower limb prostheses since specific guidelines for the testing of the trans-tibial socket portion of a prosthesis have not yet been established. The experimental set-up simulated the instant of maximum loading during the late stance phase of gait. Ten trans-tibial sockets were evaluated. Five different reinforcement materials and two resin types were used to construct the sockets. A standard four hole distal attachment plate was used to connect the socket and pylon. Each sample was loaded to failure in a servo-hydraulic materials test machine at 100 N/s. None of the composites in the study met the ISO 10328 standards for level A100, loading condition II (4025 N), as required for other prosthetic componentry. All failures occurred at the site of the pyramid attachment plate. Ultimate strength and failure type were material dependent. Load point deflection was significantly different for the resin variable (p<0.05). Statistical differences according to reinforcement material were noted in composite weight and strength-to-weight ratio (p<0.05). The fibre volume fraction was also estimated and recorded. Reinforcement material type was the primary determinant of performance for the tested samples. Carbon reinforcements performed better than fibreglass reinforcements of similar weave type. The greatest ultimate strength and strength-to-weight ratio was observed with the unidirectional carbon reinforcement.
引用
收藏
页码:113 / 122
页数:10
相关论文
共 50 条
  • [1] Optimised Analysis, Design, and Fabrication of Trans-Tibial Prosthetic Sockets
    Jweeg, Muhsin J.
    Hammoudi, Zaid S.
    Alwan, Bassam A.
    2ND INTERNATIONAL CONFERENCE ON ENGINEERING SCIENCES, 2018, 433
  • [2] Influence of the scale reduction in designing sockets for trans-tibial amputees
    Dakhil, Nawfal
    Tarrade, Tristan
    Behr, Michel
    Mo, Fuhao
    Evin, Morgane
    Thefenne, Laurent
    Liu, Tang
    Llari, Maxime
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2020, 234 (08) : 761 - 768
  • [3] Principal structural testing of trans-tibial prosthetic assemblies: specimen preparation
    Neo, LD
    Lee, PVS
    Goh, JCH
    PROSTHETICS AND ORTHOTICS INTERNATIONAL, 2000, 24 (03) : 241 - 245
  • [4] A new trim line concept for trans-tibial amputation prosthetic sockets
    Söderberg, B
    PROSTHETICS AND ORTHOTICS INTERNATIONAL, 2002, 26 (02) : 159 - 162
  • [5] Quality of polypropylene sockets for trans-tibial prostheses in low-income countries
    Jensen, J. Steen
    Raab, Wilfried
    Fisk, John
    Hartz, Christian
    Saldana, Ariel
    Harte, Carson
    PROSTHETICS AND ORTHOTICS INTERNATIONAL, 2006, 30 (01) : 45 - 59
  • [6] CLINICAL-EVALUATION OF TRANS-TIBIAL PROSTHESIS SOCKETS - A COMPARISON BETWEEN CAD-CAM AND CONVENTIONALLY PRODUCED SOCKETS
    OBERG, T
    LILJA, M
    JOHANSSON, T
    KARSZNIA, A
    PROSTHETICS AND ORTHOTICS INTERNATIONAL, 1993, 17 (03) : 164 - 171
  • [7] Static versus dynamic prosthetic weight bearing in elderly trans-tibial amputees
    Jones, ME
    Steel, JR
    Bashford, GM
    Davidson, IR
    PROSTHETICS AND ORTHOTICS INTERNATIONAL, 1997, 21 (02) : 100 - 106
  • [8] Multi-scale semi-analytical model for fatigue life prediction of trans-tibial prosthetic sockets
    Mankai, Wahbi
    Smida, Brahim Ben
    Chafra, Moez
    Cheikh, Ridha Ben
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2021, 235 (06) : 688 - 700
  • [9] Clinical field-testing of ATLAS prosthetic system for trans-tibial amputees
    Jensen, JS
    Raab, W
    PROSTHETICS AND ORTHOTICS INTERNATIONAL, 2002, 26 (02) : 86 - 92
  • [10] RESULTS OF THE PILOT PHASE OF A CLINICAL-EVALUATION OF COMPUTER-AIDED-DESIGN OF TRANS-TIBIAL PROSTHESIS SOCKETS
    HOLDEN, JM
    FERNIE, GR
    PROSTHETICS AND ORTHOTICS INTERNATIONAL, 1986, 10 (03) : 142 - 148