Laboratory vacuum spectrometer for the soft x-ray region

被引:3
|
作者
Yuryev, Yu. N. [1 ]
Lee, Hwack-Joo [1 ]
Kim, Ju-Hwang [1 ]
Cho, Yang-Koo [1 ]
Lee, Min-Kyu [1 ]
Pogrebitsky, K. Ju. [2 ]
机构
[1] Korea Res Inst Standards & Sci, Taejon 305600, South Korea
[2] AF Ioffe Phys Tech Inst, St Petersburg 194021, Russia
关键词
D O I
10.1002/xrs.1022
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A laboratory surface x-ray absorption near edge structure ((S)XANES) spectrometer used for measurements in the soft x-ray region is described. The x-ray beam path of the spectrometer is enclosed in a vacuum chamber directly connected to the x-ray generator output port. With this setup, the absorption of the x-rays in the air is avoided. The developed spectrometer uses a monochromator equipped with cylindrically bent PET(002), KAP(001), and Ge(111) crystals of different Rowland radii working in the Johann focusing geometry. Nine stepping motors are used to control the positions of the monochromator, receiving slit, sample, and detectors. An x-ray photon energy available in experiment ranges from about 1 keV to 5 keV. A registration of the photoelectrons emitted from the sample with the channeltron is used to carry out surface-sensitive measurements. The performance of the developed spectrometer is illustrated by spectra measured at the absorption K-edges of several elements from the Na-Ti set. Copyright (c) 2008 John Wiley & Sons, Ltd.
引用
收藏
页码:476 / 481
页数:6
相关论文
共 50 条
  • [1] A VACUUM X-RAY SPECTROMETER
    STEPHENSON, ST
    MASON, FD
    PHYSICAL REVIEW, 1947, 72 (08): : 744 - 744
  • [2] A VACUUM X-RAY SPECTROMETER
    TRAPEZNI.VA
    NEMNONOV, SA
    PHYSICS OF METALS AND METALLOGRAPHY-USSR, 1968, 25 (01): : 49 - &
  • [3] PHOTOMULTIPLIER SOFT X-RAY SPECTROMETER
    JACOB, L
    NOBLE, R
    YEE, H
    JOURNAL OF SCIENTIFIC INSTRUMENTS, 1960, 37 (12): : 460 - 463
  • [4] A high-vacuum x-ray spectrometer
    Siegbahn, M
    Thoraeus, R
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA AND REVIEW OF SCIENTIFIC INSTRUMENTS, 1926, 13 (02): : 235 - 242
  • [5] DIFFRACTION VACUUM X-RAY PHOTOELECTRON SPECTROMETER
    KOVALCHUK, MV
    SHILIN, YN
    DENISOV, AG
    GRAVSHIN, YM
    ZELTSER, IA
    LYAPIN, VM
    SENICHKINA, RS
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1987, 30 (03) : 715 - 719
  • [6] PHOTON-COUNTING SPECTROMETER FOR ATTENUATION MEASUREMENTS IN SOFT X-RAY REGION
    BEDO, DE
    TOMBOULIAN, DH
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1961, 32 (02): : 184 - &
  • [7] A GEIGER COUNTER VACUUM SPECTROMETER AND ITS USE FOR THE STUDY OF SOFT X-RAY LINES
    ROGERS, JL
    CHALKLIN, FC
    PROCEEDINGS OF THE PHYSICAL SOCIETY OF LONDON SECTION B, 1954, 67 (412): : 348 - 356
  • [8] A SOFT X-RAY SPECTROMETER WITH IMPROVED DRIVE
    CUTHILL, JR
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1970, 41 (03): : 422 - &
  • [9] SOLAR OBSERVATIONS WITH A SOFT X-RAY SPECTROMETER
    NEUPERT, WM
    BEHRING, WE
    JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 1962, 2 (04): : 527 - 532
  • [10] The soft X-ray spectrometer polarimeter SolpeX
    J. Sylwester
    M. Stȩślicki
    J. Bąkała
    S. Płocieniak
    Ż. Szaforz
    M. Kowaliński
    D. Ścisłowski
    P. Podgórski
    T. Mrozek
    J. Barylak
    A. Makowski
    M. Siarkowski
    Z. Kordylewski
    B. Sylwester
    S. Kuzin
    A. Kirichenko
    A. Pertsov
    S. Bogachev
    Experimental Astronomy, 2019, 47 : 199 - 223