Fatigue Fracture in Dual Modular Revision Total Hip Arthroplasty Stems Failure Analysis and Computed Tomography Diagnostics in Two Cases

被引:46
|
作者
Norman, Pontus [1 ]
Iyengar, Srinivasan [2 ]
Svensson, Ingrid [3 ]
Flivik, Gunnar [1 ]
机构
[1] Lund Univ, Dept Orthopaed, Skane Univ Hosp, Lund, Sweden
[2] Lund Univ, Div Mat Engn, Lund, Sweden
[3] Lund Univ, Div Solid Mech, Lund, Sweden
来源
JOURNAL OF ARTHROPLASTY | 2014年 / 29卷 / 04期
关键词
modular; hip prosthesis; fatigue failure; CT; FEMORAL STEMS; COMPONENT; CORROSION; IMPLANT; REPLACEMENT; JUNCTION;
D O I
10.1016/j.arth.2013.09.008
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report on two patients with fracture of a modular, tapered and distally fixed, uncemented titanium revision hip stem, not previously described. A failure analysis revealed that the cause of the fractures was the development of fatigue cracks in the mid-stem cobalt-chromium modular junction ending in corrosion-fatigue failure. No material defects or stress risers were found in any of the implants. The diameter of the mid-stem modular junction might be undersized for use in heavy and active patients. We also report a new way of detecting an undisplaced fracture at the modular junction, using the scout image from a computed tomography (CT) scan; a technique that can be used when plain radiographs are inconclusive. (C) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:850 / 855
页数:6
相关论文
共 50 条
  • [1] Modular revision stems in total hip arthroplasty
    Jones, RE
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (420) : 142 - 147
  • [2] Modular Femoral Stems for Revision Total Hip Arthroplasty
    Restrepo, Camilo
    Mashadi, Magdalena
    Parvizi, Javad
    Austin, Matthew S.
    Hozack, William J.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (02) : 476 - 482
  • [3] Revision Total Hip Arthroplasty With Modular Femoral Stems
    Stimac, Jeffrey D.
    Boles, John
    Parkes, Natalie
    Della Valle, Alejandro Gonzalez
    Boettner, Friedrich
    Westrich, Geoffrey H.
    JOURNAL OF ARTHROPLASTY, 2014, 29 (11): : 2167 - 2170
  • [4] Failure mechanisms in CoCrMo modular femoral stems for revision total hip arthroplasty
    Wang, Qiong
    Parry, Michael
    Masri, Bassam A.
    Duncan, Clive
    Wang, Rizhi
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2017, 105 (06) : 1525 - 1535
  • [5] Nonmodular Stems Are a Viable Alternative to Modular Stems in Revision Total Hip Arthroplasty
    Clair, Andrew J.
    Cizmic, Zlatan
    Vigdorchik, Jonathan M.
    Poultsides, Lazaros A.
    Schwarzkopf, Ran
    Rathod, Parthiv A.
    Deshmukh, Ajit J.
    JOURNAL OF ARTHROPLASTY, 2019, 34 (07): : S292 - S296
  • [6] The use of proximal fixed modular stems in revision of total hip arthroplasty
    Higuera, Carlos A.
    Hanna, George
    Florjancik, Karen
    Allan, D. Gordon
    Robinson, Raymond
    Barsoum, Wael K.
    JOURNAL OF ARTHROPLASTY, 2006, 21 (04): : 112 - 116
  • [7] Modular Fluted Tapered Stems in Aseptic Revision Total Hip Arthroplasty
    Abdel, Matthew P.
    Cottino, Umberto
    Larson, Dirk R.
    Hanssen, Arlen D.
    Lewallen, David G.
    Berry, Daniel J.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2017, 99 (10): : 873 - 881
  • [8] Outcomes of dual modular cementless femoral stems in revision hip arthroplasty
    Ghoz, Ali
    Broadhead, Matthew L.
    Morley, John
    Tavares, Shawn
    McDonald, David
    ORTHOPEDIC REVIEWS, 2014, 6 (01) : 43 - 47
  • [9] Complex revision total hip arthroplasty with modular stems at a mean of 14 years
    McCarthy, Joseph C.
    Lee, Jo-ann
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (465) : 166 - 169
  • [10] Influence of Technique With Distally Fixed Modular Stems in Revision Total Hip Arthroplasty
    Patel, Preetesh D.
    Klika, Alison K.
    Murray, Trevor G.
    Elsharkawy, Karim A.
    Krebs, Viktor E.
    Barsoum, Wael K.
    JOURNAL OF ARTHROPLASTY, 2010, 25 (06): : 926 - 931