Outcomes of the Recalled Articular Surface Replacement Metal-on-Metal Hip Implant System: A Systematic Review

被引:29
|
作者
Laaksonen, Inari [1 ]
Donahue, Gabrielle S. [2 ]
Madanat, Rami [2 ,3 ]
Makela, Keijo T. [1 ]
Malchau, Henrik [2 ,3 ]
机构
[1] Turku Univ Hosp, Dept Orthoped & Traumatol, Luolavuorentie 2, Turku 20700, Finland
[2] Massachusetts Gen Hosp, Harris Orthopaed Lab, Boston, MA 02114 USA
[3] Harvard Med Sch, Dept Orthopaed Surg, Boston, MA USA
来源
JOURNAL OF ARTHROPLASTY | 2017年 / 32卷 / 01期
关键词
articular surface replacement; total hip arthroplasty; hip resurfacing arthroplasty; systematic review; metal-on-metal; FINNISH ARTHROPLASTY REGISTER; DIAMETER TOTAL HIP; ION LEVELS; RESURFACING ARTHROPLASTY; FOLLOW-UP; ACETABULAR INCLINATION; ADVERSE-REACTION; ASR; COMPONENT; REVISION;
D O I
10.1016/j.arth.2016.06.060
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The aim of this review was to synthesize the main findings of clinical studies that have evaluated outcomes of the articular surface replacement (ASR) Hip System. Methods: We performed a systematic literature search to identify all articles published between January 2008 and June 2015 that included ASR hip resurfacing arthroplasty (ASR HRA) or ASR total hip arthroplasty (ASR XL THA) outcomes according to the PRISMA statement. Results: A total of 56 studies were assessed. The prevalence of adverse local tissue reactions (ALTRs) and revision rates were found to be high. ALTR prevalence varied from 12.5% to 69% (mean, 33.5%). Mean revision rate for any reason at 4-year to 7-year follow-up was 13.8% (range, 5.6%-31%) for ASR HRA and 14.5% (range, 0%-37%) for ASR XL THA. Femoral head size <53 mm was found to correlate with higher blood metal ion levels. Femoral head size >44 mm was not associated with higher ALTR prevalence or revision rates in ASR XL THA. High blood metal ion levels (>7 mu g/L Co, >7 mu g/L Cr) were associated with higher failure rates and bearing-related complications. The role of cup positioning was found to be controversial. Conclusion: ALTR prevalence and failure rates were high. High blood metal ion levels were a risk factor for ALTR and failure. Surprisingly, the role of cup positioning and large femoral head size in ASR XL THA were controversial. These findings should be considered in the clinical follow-up and risk stratification of patients with the ASR Hip System. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:341 / 346
页数:6
相关论文
共 50 条
  • [1] Outcomes of a metal-on-metal total hip replacement system
    Matharu, G. S.
    Theivendran, K.
    Pynsent, P. B.
    Jeys, L.
    Pearson, A. M.
    Dunlop, D. J.
    [J]. ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2014, 96 (07) : 530 - 535
  • [2] Relationship of Blood Metal Ion Levels and Leukocyte Profiles in Patients With Articular Surface Replacement Metal-on-Metal Hip Replacement
    Markel, David C.
    Bergum, christopher
    Flynn, Jeffrey
    Jackson, Nancy
    Bou-Akl, Therese
    Ren, Weiping
    [J]. ORTHOPEDICS, 2018, 41 (03) : E424 - E431
  • [3] Revision rates for metal-on-metal hip resurfacing and metal-on-metal total hip arthroplasty - a systematic review
    Sorensen, Sofie-Amalie L. Ras
    Jorgensen, Henrik L.
    Sporing, Sune L.
    Lauritzen, Jes B.
    [J]. HIP INTERNATIONAL, 2016, 26 (06) : 515 - 521
  • [4] Metal-on-metal hip surface replacement THE ROUTINE USE IS NOT JUSTIFIED
    Dunbar, M. J.
    Prasad, V.
    Weerts, B.
    Richardson, G.
    [J]. BONE & JOINT JOURNAL, 2014, 96B : 17 - 21
  • [5] Metal-on-metal total hip replacement
    Silva, M
    Heisel, C
    Schmalzried, TP
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (430) : 53 - 61
  • [6] Revision of metal-on-metal hip replacements and resurfacings for adverse reaction to metal debris: a systematic review of outcomes
    Matharu, Gulraj S.
    Pynsent, Paul B.
    Dunlop, David J.
    [J]. HIP INTERNATIONAL, 2014, 24 (04) : 311 - 320
  • [7] Influence of implant design on blood metal ion concentrations in metal-on-metal total hip replacement patients
    Matharu, Gulraj S.
    Berryman, Fiona
    Brash, Lesley
    Pynsent, Paul B.
    Treacy, Ronan B.
    Dunlop, David J.
    [J]. INTERNATIONAL ORTHOPAEDICS, 2015, 39 (09) : 1803 - 1811
  • [8] Influence of implant design on blood metal ion concentrations in metal-on-metal total hip replacement patients
    Gulraj S. Matharu
    Fiona Berryman
    Lesley Brash
    Paul B. Pynsent
    Ronan B. Treacy
    David J. Dunlop
    [J]. International Orthopaedics, 2015, 39 : 1803 - 1811
  • [9] The 510(k) Ancestry of a Metal-on-Metal Hip Implant
    Ardaugh, Brent M.
    Graves, Stephen E.
    Redberg, Rita F.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2013, 368 (02): : 97 - 100