A Biomechanical Evaluation of All-Inside 2-Stitch Meniscal Repair Devices With Matched Inside-Out Suture Repair

被引:17
|
作者
Ramappa, Arun J. [1 ]
Chen, Alvin [1 ]
Hertz, Benjamin [1 ]
Wexler, Michael [1 ]
Bournissaint, Leandro Grimaldi [1 ]
DeAngelis, Joseph P. [1 ]
Nazarian, Ara [1 ]
机构
[1] Harvard Univ, Beth Israel Deaconess Med Ctr, Ctr Adv Orthopaed Studies, Dept Orthopaed Surg,Sch Med, Boston, MA 02215 USA
来源
AMERICAN JOURNAL OF SPORTS MEDICINE | 2014年 / 42卷 / 01期
关键词
meniscal repair; all inside; inside out; initial displacement; biomechanics; 2-stitch repair; WEIGHT POLYETHYLENE SUTURE; HORIZONTAL SUTURE; IN-VITRO; STRENGTH; SYSTEM; ARROW;
D O I
10.1177/0363546513505190
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Many all-inside suture-based devices are currently available, including the Meniscal Cinch, FasT-Fix, Ultra FasT-Fix, RapidLoc, MaxFire, and CrossFix System. These different devices have been compared in various configurations, but to our knowledge, the Sequent meniscal repair device, which applies running sutures, has not been compared with the Ultra FasT-Fix, nor has it been compared with its suture, No. 0 Hi-Fi, using an inside-out repair technique. Purpose: To assess the quality of the meniscal repair, all new devices should be compared with the gold standard: the inside-out repair. To that end, this study aims to compare the biomechanical characteristics of running sutures delivered by the Sequent meniscal repair device with 2 vertical mattress sutures applied using the Ultra FasT-Fix device and with 2 vertical mattress sutures using an inside-out repair technique with No. 0 Hi-Fi suture. Study Design: Controlled laboratory study. Methods: Paired (medial and lateral), fresh-frozen porcine menisci were randomly assigned to 1 of 3 groups: Sequent (n = 17), Ultra FasT-Fix (n = 19), and No. 0 Hi-Fi inside-out repair (n = 20). Bucket-handle tears were created in all menisci and were subjected to repair according to their grouping. Once repaired, the specimens were subjected to cyclic loading (100, 300, and 500 cycles), followed by loading to failure. Results: The Sequent and Ultra FasT-Fix device repairs and the suture repair exhibited low initial displacements. The Sequent meniscal repair device demonstrated the lowest displacement in response to cyclic loading. No. 0 Hi-Fi suture yielded the highest load to failure. Conclusion: With the development of the next generation of all-inside meniscal repair devices, surgeons may use these findings to select the method best suited for their patients. Clinical Relevance: The Sequent meniscal repair device displays the least amount of displacement during cyclic loading but has a similar failure load to other devices.
引用
收藏
页码:194 / 199
页数:6
相关论文
共 50 条
  • [1] Comparison of All-Inside Meniscal Repair Devices With Matched Inside-Out Suture Repair
    Rosso, Claudio
    Kovtun, Konstantin
    Dow, William
    McKenzie, Brett
    Nazarian, Ara
    DeAngelis, Joseph P.
    Ramappa, Arun J.
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2011, 39 (12): : 2634 - 2639
  • [2] Comparison of All-Inside Meniscal Repair Devices with Matched Inside-Out Suture Repair
    Rosso, Claudio
    Kovtun, Konstantin
    Nazarian, Ara
    DeAngelis, Joseph P.
    Ramappa, Arun
    [J]. SWISS MEDICAL WEEKLY, 2011, 141 : 30S - 30S
  • [3] Biomechanical Comparison of All-inside Meniscal Repair Devices with their Matched Inside-Out Suture Repair: Introducing 10.000 and 100.000 Loading Cycles
    Rosso, Claudio
    Mueller, Sebastian
    Buckland, Daniel
    Schwenk, Tanja
    Zimmermann, Simon
    de Wild, Michael
    Valderrabano, Victor
    [J]. SWISS MEDICAL WEEKLY, 2014, 144 : 26S - 26S
  • [4] Comparison of Arthroscopic Medial Meniscal Suture Repair Techniques Inside-Out Versus All-Inside Repair
    Choi, Nam-Hong
    Kim, Tae-Hyung
    Victoroff, Brian N.
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2009, 37 (11): : 2144 - 2150
  • [5] A Comparison of All-inside and Inside-out Meniscal Repair in Elite Athletes
    Borque, Kyle A.
    Laughlin, Mitzi S.
    Webster, Emily
    Jones, Mary
    Pinheiro, Vitor Hugo
    Williams, Andy
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2023, 51 (03): : 579 - 584
  • [6] All-inside meniscal repair devices
    King, DJ
    Matava, MJ
    [J]. OPERATIVE TECHNIQUES IN SPORTS MEDICINE, 2004, 12 (03) : 161 - 169
  • [7] The Biomechanical Performance of the Latest All-Inside Meniscal Repair Devices
    Barber, F. Alan
    Howard, Michael S.
    Ashraf, Wasik
    Spenciner, David B.
    [J]. ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2020, 36 (12): : 3001 - 3007
  • [8] Biomechanical Characteristics of All-Suture Meniscal Repair Devices Compared With PEEK-Anchored Devices and Inside-Out Suture for Meniscal Repair: A Porcine Study
    Barber, F. Alan
    Davidson, Philip A.
    Henninger, Heath B.
    [J]. ORTHOPAEDIC JOURNAL OF SPORTS MEDICINE, 2024, 12 (05)
  • [9] All-Inside Meniscal Repair Devices Compared With Their Matched Inside-Out Vertical Mattress Suture Repair Introducing 10,000 and 100,000 Loading Cycles
    Rosso, Claudio
    Mueller, Sebastian
    Buckland, Daniel M.
    Schwenk, Tanja
    Zimmermann, Simon
    de Wild, Michael
    Valderrabano, Victor
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2014, 42 (09): : 2226 - 2233
  • [10] All-inside meniscal repair devices and techniques
    Tuman, Jeff
    Haro, Marc S.
    Foley, Sarah
    Diduch, David
    [J]. EXPERT REVIEW OF MEDICAL DEVICES, 2012, 9 (02) : 147 - 157