Time-based multisyringe flow injection system for the spectrofluorimetric determination of aluminium

被引:39
|
作者
de Armas, G
Miró, M
Cladera, A
Estela, JM
Cerdà, V [1 ]
机构
[1] Univ Balearic Isl, Dept Chem, E-07071 Palma de Mallorca, Spain
[2] Univ Havana, Dept Analyt Chem, Mat & Reagents Inst, Havana, Cuba
关键词
multisyringe flow injection analysis (MSFIA); time-based injection; multicommutation; aluminium; 8-hydroxyquinoline-5-sulphonic acid; drinking water; spectrofluorimetry;
D O I
10.1016/S0003-2670(01)01588-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A time-based multisyringe flow injection procedure with spectrofluorimetric detection is proposed in this paper for the determination of aluminium in drinking water. The flow methodology is based on the simultaneous or sequential injection of sample and chelating reagent (viz. 8-hydroxyquinoline-5-sulphonic acid) plugs using a multicommutation approach so that three successive injections may be performed with a sole displacement of the piston driver bar of the burette. Thus, an injection throughput as high as 154 h(-1) is achieved by sampling a 182 mul sample zone. In order to enhance the luminescence, the reaction is carried out in micellar medium using hexadecyltrimethylammonium chloride as surfactant. The influence of geometric and hydrodynamic variables as well as several parameters such as multicommutation timing, ligand and surfactant concentration and reagent pH was assessed. Under the selected working conditions, a linear dynamic range from 10 to 500 mugl(-1) Al(III). a 3sigma detection limit of 0.5 mug l(-1) and a coefficient of variation of 0.6% at the 30 mug l(-1) level were obtained. The analytical features were compared with those reported in previous flow injection and sequential injection methods. The multisyringe technique was successfully applied to the determination of aluminium in drinking water at low mineralisation levels, validating the results by inductively coupled plasma atomic emission spectrometry. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:149 / 157
页数:9
相关论文
共 50 条
  • [1] Spectrophotometric determination of chloride in waters using a multisyringe flow injection system
    Maya, Fernando
    Manuel Estela, Jose
    Cerda, Victor
    [J]. TALANTA, 2008, 74 (05) : 1534 - 1538
  • [2] Multisyringe flow injection spectrofluorimetric determination of warfarin at trace levels with on-line solid-phase preconcentration
    de Armas, G
    Miró, M
    Estela, JM
    Cerdà, V
    [J]. ANALYTICA CHIMICA ACTA, 2002, 467 (1-2) : 13 - 23
  • [3] Conductometric determination of ammonium by a multisyringe flow injection system applying gas diffusion
    Henriquez, Camelia
    Horstkotte, Burkhard
    Cerda, Victor
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2013, 93 (12) : 1236 - 1252
  • [4] A multisyringe flow injection system coupled with a gas diffusion cell for ammonium determination
    Klimundová, J
    Forteza, R
    Cerdá, V
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2003, 83 (03) : 233 - 246
  • [5] AUTOMATIC MULTISYRINGE FLOW INJECTION SYSTEM FOR THE SPECTROPHOTOMETRIC DETERMINATION OF TINIDAZOLE IN PHARMACEUTICAL PREPARATIONS
    Guzman-Mar, J. L.
    Hinojoza-Reyes, L.
    Hernandez-Ramirez, A.
    Peralta-Hernandez, J. M.
    Lopez-Chuken, V. J.
    Lopez-De-Alba, P. L.
    Cerda-Martin, V.
    [J]. JOURNAL OF THE CHILEAN CHEMICAL SOCIETY, 2010, 55 (02): : 215 - 218
  • [6] Development of Flow Injection Spectrofluorimetric Detection System for the Determination of Homocysteine
    Nouroozi, Siavash
    Biglary, Hassan
    Haghighi, Behzad
    [J]. JOURNAL OF FLUORESCENCE, 2012, 22 (01) : 365 - 371
  • [7] Development of Flow Injection Spectrofluorimetric Detection System for the Determination of Homocysteine
    Siavash Nouroozi
    Hassan Biglary
    Behzad Haghighi
    [J]. Journal of Fluorescence, 2012, 22 : 365 - 371
  • [8] TIME-BASED FLOW-INJECTION ANALYSIS
    STEWART, KK
    [J]. ANALYTICA CHIMICA ACTA, 1986, 179 : 59 - 68
  • [9] Flow injection extraction spectrofluorimetric determination of aluminium as the tetraphenylphosphonium aluminium(III) 8-hydroxyquinoline-5-sulphonate
    Burns, DT
    Pornsinlapatip, P
    [J]. ANALYTICAL LETTERS, 2002, 35 (06) : 1085 - 1093
  • [10] Multisyringe flow injection spectrophotometric determination of uranium in water samples
    Guzman Mar, Jorge L.
    Lopez Martinez, Leticia
    Lopez de Alba, Pedro L.
    Ornelas Soto, Nancy
    Cerda Martin, Victor
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2009, 281 (03) : 433 - 439