Novel Method of Temperature Stabilization for SiPM-Based Detectors

被引:4
|
作者
Stolin, Alexander V. [1 ]
Majewski, Stan [1 ]
Raylman, Raymond R. [1 ]
机构
[1] W Virginia Univ, Dept Radiol, Ctr Adv Imaging, Morgantown, WV 26506 USA
关键词
Nuclear medicine imaging; PET Instrumentation; silicon photomultipliers; temperature stabilization;
D O I
10.1109/TNS.2013.2273398
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
One of the drawbacks of silicon-based photomultipliers is variation of performance with even small temperature fluctuations. The purpose of this work is to investigate a novel method of temperature stabilization of SiPM-based imaging device. Our prototype cooling system utilizes optically transparent cooling fluid as means of removing heat from the light detector and spreading scintillation light among several SiPM units. Cooled liquid passes through a 2 mm opening between two Mylar sheets. The Mylar sheets are held together with an aluminum frame. An array of SiPMs and an array of LYSO scintillation crystals are attached to opposite sides of the cooling light guide. Temperature uniformity of the optical surface was evaluated with an infrared camera. A temperature gradient of 0.09 degrees C/cm was measured along the direction of the coolant flow. The system was tested with a 5 x 5 array of Hamamatsu S10943 SiPMs coupled to a 16 x 16 matrix of 1.5 x 1.5 x 10 mm LYSO crystals. Testing procedures included evaluation of temperature dependence of detector gain, integrated noise, ability to resolve individual crystals and energy resolution. Detector gain varied by 9.3% over the range of 5 degrees C. Energy resolution change of 5% was observed over the same temperature range. Integrated noise contribution showed moderate increase of 2.2% over the range of 7 degrees C. The separation of individual crystals was found adequate with only a weak dependence of peak-to-valley contrast ratios on temperature. An assessment of performance stability at constant temperature revealed 0.2% variation in gain and 0.7% variation in energy resolution during an hour-long acquisition. It is concluded that the proposed method is a viable option for cooling and temperature stabilizing of SiPM-based systems.
引用
收藏
页码:3181 / 3187
页数:7
相关论文
共 50 条
  • [1] SiPM-based Gamma Detectors of GECAM
    Feng, Pei-Yi
    Sun, Xi-Lei
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2024, 1069
  • [2] Research on the temperature characteristics of SiPM-based LaBr3(Ce) detectors
    Tao Yu
    Siming Guo
    Jinjie Wu
    Zhenghua An
    Xiaoyu Qie
    Kaiyue Guo
    Journal of the Korean Physical Society, 2023, 82 : 12 - 18
  • [3] Research on the temperature characteristics of SiPM-based LaBr3(Ce) detectors
    Yu, Tao
    Guo, Siming
    Wu, Jinjie
    An, Zhenghua
    Qie, Xiaoyu
    Guo, Kaiyue
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2023, 82 (01) : 12 - 18
  • [4] Applications of Innovative SiPM-based PVT Scintillator Detectors
    Meshkian, M.
    Allwork, C.
    Gendotti, U.
    Ellis, M.
    Schotanus, P.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [5] Gain stabilization and consistency correction approach for multiple SiPM-based gamma-ray detectors on GECAM
    Zhang, Dali
    Li, Xinqiao
    Wen, Xiangyang
    Xiong, Shaolin
    An, Zhenghua
    Xu, Yanbing
    Sun, Xilei
    Qiao, Rui
    Li, Zhengwei
    Gong, Ke
    Guo, Dongya
    Hou, Dongjie
    Li, Yanguo
    Liang, Xiaohua
    Liu, Xiaojing
    Liu, Yaqing
    Peng, Wenxi
    Yang, Sheng
    Zhang, Fan
    Zhao, Xiaoyun
    Zheng, Chao
    Li, Chaoyang
    Yi, Qibin
    Liu, Jiacong
    Xiao, Shuo
    Cai, Ce
    Wang, Chenwei
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2022, 1027
  • [6] Hybrid charge division multiplexing method for SiPM-based high-resolution PET detectors
    Park, Haewook
    Ko, Guen Bae
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57
  • [7] The data acquisition algorithm designed for the SiPM-based detectors of GECAM satellite
    Y. Q. Liu
    K. Gong
    X. Q. Li
    X. Y. Wen
    Z. H. An
    C. Cai
    Z. Chang
    G. Chen
    C. Chen
    Y. Y. Du
    M. Gao
    R. Gao
    D. Y. Guo
    J. J. He
    D. J. Hou
    Y. G. Li
    C. Y. Li
    G. Li
    L. Li
    X. F. Li
    M. S. Li
    X. H. Liang
    X. J. Liu
    F. J. Lu
    H. Lu
    B. Meng
    W. X. Peng
    F. Shi
    X. L. Sun
    H. Wang
    J. Z. Wang
    Y. S. Wang
    H. Z. Wang
    X. Wen
    S. Xiao
    S. L. Xiong
    Y. B. Xu
    Y. P. Xu
    S. Yang
    J. W. Yang
    Q. B. Yi
    Fan Zhang
    D. L. Zhang
    S. N. Zhang
    C. Y. Zhang
    C. M. Zhang
    Fei Zhang
    X. Y. Zhao
    Y. Zhao
    X. Zhou
    Radiation Detection Technology and Methods, 2022, 6 : 70 - 77
  • [8] Packaging strategies for large SiPM-based cryogenic photo-detectors
    Kochanek I.
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2020, 980
  • [9] Development of a SiPM-based CsI(Tl) spectrometer with gain stabilization designs for rapid temperature variations
    Gong, Pin
    Han, Zhenyang
    Gao, Le
    Wang, Peng
    Zhang, Jinzhao
    Tang, Xiaobin
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2019, 927 : 301 - 305
  • [10] Ultra Precise Timing with SiPM-Based TOF PET Scintillation Detectors
    Seifert, Stefan
    Vinke, Ruud
    van Dam, Herman T.
    Loehner, Herbert
    Dendooven, Peter
    Beekman, Freek J.
    Schaart, Dennis R.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 2329 - +