Validating a new device for measuring tear evaporation rates

被引:31
|
作者
Rohit, Athira [1 ,2 ]
Ehrmann, Klaus [1 ,2 ]
Naduvilath, Thomas [1 ,2 ]
Willcox, Mark [2 ]
Stapleton, Fiona [1 ,2 ]
机构
[1] Brien Holden Vis Inst, Sydney, NSW, Australia
[2] Univ New S Wales, Sch Optometry & Vis Sci, Sydney, NSW, Australia
关键词
contact lens; lipid layer; tear evaporation; tear film; validation; MACULAR DEGENERATION; HEALTH INFORMATION; INTERNET; QUALITY; OPHTHALMOLOGY; GLAUCOMA; SUPPORT; STATEMENT; MARIJUANA; WEB;
D O I
10.1111/opo.12096
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To calibrate and validate a commercially available dermatology instrument to measure tear evaporation rate of contact lens wearers. Methods: A dermatology instrument was modified by attaching a swim goggle cup such that the cup sealed around the eye socket. Results for the unmodified instrument are dependent on probe area and enclosed volume. Calibration curves were established using a model eye, to account for individual variations in chamber volume and exposed area. Fifteen participants were recruited and the study included a contact lens wear and a no contact lens wear stage. Day and diurnal variation of the measurements were assessed by taking the measurement three times a day over 2 days. The coefficient of repeatability of the measurement was calculated and a linear mixed model assessed the influence of humidity, temperature, contact lens wear, day and diurnal variations on tear evaporation rate. The associations between variables were assessed using Pearson correlation coefficient. Results: Absolute evaporation rates with and without contact lens wear were calculated based on the new calibration. The measurements were most repeatable during the evening with no lens wear (COR = 49 g m(-2) h) and least repeatable during the evening with contact lens wear (COR = 93 g m(-2) h). Humidity (p = 0.007), and contact lens wear (p < 0.01), significantly affected the tear evaporation rate. However, temperature (p = 0.54) diurnal variation (p = 0.85) and different days (p = 0.65) had no significant effect after controlling for humidity. Conclusion: Tear evaporation rates can be measured using a modified dermatology instrument. Measurements were higher and more variable with lens wear consistent with previous literature. Control of environmental conditions is important as a higher humidity results in a reduced evaporation rate.
引用
收藏
页码:53 / 62
页数:10
相关论文
共 50 条
  • [1] Validating a new device for measuring tear evaporation rates (vol 34, pg 53, 2014)
    Rohit, A.
    Ehrmann, K.
    Naduvilath, T.
    Willcox, M.
    Stapleton, F.
    [J]. OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2019, 39 (02) : 127 - 127
  • [2] Repeatability of a New Method for Measuring Tear Evaporation Rates
    Petznick, Andrea
    Tan, Jen Hong
    Boo, Shiao Khee
    Lee, Sze Yee
    Acharya, U. Rajendra
    Tong, Louis
    [J]. OPTOMETRY AND VISION SCIENCE, 2013, 90 (04) : 366 - 371
  • [3] NEW DEVICE FOR MEASURING FLOW RATES
    HASTINGS, CR
    [J]. WATER & WASTES ENGINEERING, 1969, 6 (08): : 54 - &
  • [4] Validation of a new Measuring Device for the Analysis of the Tear Film Osmolarity
    不详
    [J]. OPHTHALMOLOGE, 2019, 116 (06): : 497 - 497
  • [5] OCULAR WATER EVAPORATION AND THE DRY EYE - A NEW MEASURING DEVICE
    MATHERS, WD
    BINARAO, G
    PETROLL, M
    [J]. CORNEA, 1993, 12 (04) : 335 - 340
  • [6] NEW DEVICE FOR MEASURING CRYSTAL-GROWTH RATES
    OFFERMANN, H
    ULRICH, J
    [J]. CHEMIE INGENIEUR TECHNIK, 1979, 51 (11) : 1135 - 1135
  • [7] Measuring and validating social program replacement rates
    Scruggs, Lyle
    [J]. JOURNAL OF EUROPEAN PUBLIC POLICY, 2013, 20 (09) : 1267 - 1284
  • [8] MICROBALANCE FOR MEASURING EVAPORATION RATES IN VACUUM
    FREUNDLICH, MM
    [J]. VACUUM, 1964, 14 (08) : 293 - &
  • [9] CALORIMETER FOR MEASURING LIQUID EVAPORATION RATES
    READING, CA
    REISER, A
    [J]. JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1977, 10 (10): : 1069 - 1071
  • [10] Tear evaporation rates: What does the literature tell us?
    Wong, Stephanie
    Murphy, Paul J.
    Jones, Lyndon
    [J]. CONTACT LENS & ANTERIOR EYE, 2018, 41 (03): : 297 - 306