Imaging Characteristics of 124I Between 3D and 2D on Siemens ECAT HR plus PET Scanner

被引:5
|
作者
Lee, Young Sub [1 ,2 ]
Kim, Jin Su [1 ]
Kim, Hee-Joung [2 ]
Woo, Sang-Keun [1 ]
Kim, Jong Guk [1 ]
Park, Ji Ae [1 ]
Choi, Chang Woon [1 ]
Lim, Sang Moo [1 ]
Kim, Kyeong Min [1 ]
机构
[1] Korea Inst Radiol & Sci, Seoul 139706, South Korea
[2] Yonsei Univ, Dept Radiol Sci, Wonju 220710, South Korea
关键词
Image quality; reconstruction algorithms; spatial resolution; whole-body PET; PERFORMANCE EVALUATION; POSITRON EMITTERS;
D O I
10.1109/TNS.2013.2240012
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
I-124 has a complex decay scheme with high gamma energy and low positron abundance. In this study, comparative performance measurements of I-124 were performed in terms of spatial resolution, sensitivity, and image quality. All measurements were performed using both 2D and 3D PET in both brain mode and whole body mode. The transverse and axial spatial resolutions at 1 cm were 5.56 and 6.07 mm for I-124, and were 4.58 and 4.77 mm and for F-18, respectively. Sensitivities were 0.5 kcps/MBq (2D) and 3.4 kcps/MBq (3D) for I-124, and 1.8 kcps/MBq (2D) and 9.8 kcps/MBq (3D) for F-18. The %contrast of 3D was higher than that of 2D in I-124. For I-124 PET imaging, 3D acquisition with brain mode was highest achievable imaging acquisition mode with finer spatial resolution and higher contrast. This result will be useful for I-124 PET imaging
引用
收藏
页码:797 / 801
页数:5
相关论文
共 50 条
  • [1] Comparative measurement of I-124 and F-18 on Siemens ECAT HR plus PET Scanner
    Lee, Y.
    Kim, J.
    Kim, K.
    Yu, A.
    Kim, J.
    Park, J.
    Lee, W.
    Woo, S.
    Kim, H.
    An, K.
    Cheon, G.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 : S334 - S334
  • [2] Accuracy of CPU versus vector processor based reconstructions of 2D and 3D PET scans with the ECAT HR plus PET scanner
    Balen, SCM
    Boellaard, R
    Lammertsma, AA
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2001, 28 (08): : 1136 - 1136
  • [3] Clinical oncology imaging with the ECAT EXACT HR plus operated in 3D
    Beyer, T
    Nicoletti, R
    Bruckbauer, T
    Rocak, D
    Aigner, R
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2001, 28 (08): : 1081 - 1081
  • [4] Comparison of Imaging Characteristics of 124I PET for Determination of Optimal Energy Window on the Siemens Inveon PET
    Yu, A. Ram
    Kim, Hee-Joung
    Lim, Sang Moo
    Kim, Jin Su
    [J]. BIOMED RESEARCH INTERNATIONAL, 2016, 2016
  • [5] Quantitative and qualitative assessment of I124 using 2D/3D PET
    Peters, Brooke
    Farhoud, Mohammed
    Weichert, Jamey
    Jaskowiak, Christine
    Hall, Lance
    Perlman, Scott
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2013, 54
  • [6] Quantitative imaging of yttrium-86 PET with the ECAT EXACT HR+ in 2D mode
    Kull, T
    Ruckgaber, J
    Weller, R
    Reske, S
    Glatting, G
    [J]. CANCER BIOTHERAPY AND RADIOPHARMACEUTICALS, 2004, 19 (04) : 482 - 490
  • [7] 3D PET reconstruction with the ECAT EXACT HR using Fourier rebinning
    Defrise, M
    Sibomana, M
    Michel, C
    Newport, D
    [J]. 1995 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD, VOLS 1-3, 1996, : 1316 - 1320
  • [8] Evaluation of the ECAT EXACT HR+ 3D PET scanner in O-15-water brain activation studies
    MorenoCantu, JJ
    Thompson, CJ
    Zatorre, RJ
    [J]. 1996 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1997, : 1280 - 1284
  • [9] Comparison of 2D and 3D myocardial PET imaging.
    Knesaurek, K
    Machac, J
    Krynyckyi, BR
    Almeida, OD
    Parisi, JC
    Kim, CK
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2001, 42 (05) : 46P - 46P
  • [10] Performance characteristics of a newly developed PET/CT scanner using NEMA standards in 2D and 3D modes
    Mawlawi, O
    Podoloff, DA
    Kohlmyer, S
    Williams, JJ
    Stearns, CW
    Culp, RF
    Macapinlac, H
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2004, 45 (10) : 1734 - 1742