Long-term results with Matrix coils vs. GDC: an angiographic and histopathological comparison

被引:11
|
作者
Roth, C. [1 ,2 ]
Struffert, T. [3 ]
Grunwald, I. Q. [2 ]
Romeike, B. F. M. [4 ]
Krick, C. [2 ]
Papanagiotou, P. [2 ]
Krampe, P. [2 ]
Reith, W. [2 ]
机构
[1] Univ Clin Saarland, Dept Diagnost & Intervent Neuroradiol, D-66421 Homburg, Germany
[2] Univ Saarland, Dept Diagnost & Intervent Neuroradiol, D-6650 Homburg, Germany
[3] Univ Hosp Erlangen, Dept Neuroradiol, Erlangen, Germany
[4] Univ Jena, Dept Neuropathol, Jena, Germany
关键词
experimental aneurysm model; GDC; matrix; coil; rabbit;
D O I
10.1007/s00234-008-0392-8
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Introduction The aim of the study was to compare standard platinum Guglielmi detachable coils (GDC) with coated platinum coils (Matrix; both Boston Scientific, Fremont, CA) regarding handling, complications, occlusion and recanalization rate after 3 and 6 (A) over bar months. Methods Aneurysms in the right common carotid artery were created in 25 rabbits. The animals were divided into five groups of five animals each. The animals of group 1 (the control group) received no treatment of the induced aneurysms, the animals of groups 2 and 3 (killed at 3 and 6 (A) over bar months) were treated with standard GDC, and the animals of groups 4 and 5 (killed at 3 and 6 (A) over bar months) were treated with Matrix coils. Results Histopathological evaluation showed organized thrombus formation and connective tissue with neovascularization around the implanted coils in all the treated groups. The achieved occlusion rates in groups 2 and 3 were identical to those in groups 4 and 5. Thus the long-term results of aneurysm treatment with GDC and Matrix coils show no differences regarding occlusion and recanalization rates. The only noticeable difference was the difference in handling. More force was required to pushing the Matrix coils forward through the microcatheter and there was more friction in coil interaction in the aneurysm. Conclusion The bioactive coating of the Matrix coil produces no significant benefit in achieving higher occlusion and lower recanalization rates, and the coil is more difficult to handle. Future bioactive coils must be shown to produce significantly better long-term results than GDC and their ease of handling has to be improved.
引用
收藏
页码:693 / 699
页数:7
相关论文
共 50 条
  • [1] Long-term results with Matrix coils vs. GDC: an angiographic and histopathological comparison
    C. Roth
    T. Struffert
    I. Q. Grunwald
    B. F. M. Romeike
    C. Krick
    P. Papanagiotou
    P. Krampe
    W. Reith
    Neuroradiology, 2008, 50 : 693 - 699
  • [2] Angiographic analysis of immediate and long-term results of PTCR vs. PTCA in complex lesions (COBRA study)
    Dietz, U
    Rupprecht, HJ
    Ekinci, O
    Dill, T
    Erbel, R
    Kuck, KH
    Abdollahnia, R
    Rippin, G
    Meyer, J
    Hamm, C
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2001, 53 (03) : 359 - 367
  • [3] Angiographic and histological comparison of canine bifurcation aneurysms treated with first generation matrix and standard GDC coils
    Turk, Aquilla S.
    Luty, Christopher M.
    Carr-Brendel, Victoria
    Polyakov, Igor
    Consigny, Dan
    Grinde, Julie
    Mukherjee, Rajat
    Strother, Charles M.
    NEURORADIOLOGY, 2008, 50 (01) : 57 - 65
  • [4] Angiographic and histological comparison of canine bifurcation aneurysms treated with first generation matrix and standard GDC coils
    Aquilla S. Turk
    Christopher M. Luty
    Victoria Carr-Brendel
    Igor Polyakov
    Dan Consigny
    Julie Grinde
    Rajat Mukherjee
    Charles M. Strother
    Neuroradiology, 2008, 50 : 57 - 65
  • [5] Long-term angiographic results of coronary endarterectomy
    Ferraris, VA
    Harrah, JD
    Moritz, DM
    Striz, M
    Striz, D
    Ferraris, SP
    ANNALS OF THORACIC SURGERY, 2000, 69 (06): : 1737 - 1743
  • [6] Endovascular treatment of ruptured intracranial aneurysms with detachable coils: Long-term clinical and serial angiographic results
    Sluzewski, M
    van Rooij, WJ
    Rinkel, GJE
    Wijnaida, D
    RADIOLOGY, 2003, 227 (03) : 720 - 724
  • [7] Iv IgG in CIDP - initial long-term therapy vs. recurrent long-term therapy - comparison of effectiveness
    Zielinska, M.
    EUROPEAN JOURNAL OF NEUROLOGY, 2010, 17 : 301 - 301
  • [8] Long-term angiographic results of coronary endarterectomy - Discussion
    Baumgartner, WA
    Sundt, TM
    Snow, N
    Urschel, HC
    ANNALS OF THORACIC SURGERY, 2000, 69 (06): : 1743 - 1743
  • [9] Long-Term Results of Aortic Valve Replacement - Mechanical Prostheses vs. Bioprostheses
    Ikeno, Yuki
    Okita, Yutaka
    CIRCULATION JOURNAL, 2014, 78 (11) : 2627 - 2630
  • [10] A comparison of long-term results after Baerveldt 250 implantation in advanced uveitic vs. other forms of glaucoma
    Cazana, Ioana Maria
    Boehringer, Daniel
    Reinhard, Thomas
    Anton, Alexandra
    Ness, Thomas
    Lubke, Jan
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2022, 260 (09) : 2991 - 3000