Adaptive optics enhanced simultaneous en-face optical coherence tomography and scanning laser ophthalmoscopy

被引:100
|
作者
Merino, D [1 ]
Dainty, C
Bradu, A
Podoleanu, AG
机构
[1] Natl Univ Ireland Univ Coll Galway, Appl Opt Grp, Dept Expt Phys, Galway, Ireland
[2] Univ Kent, Sch Phys Sci, Appl Opt Grp, Canterbury CT2 7NR, Kent, England
基金
英国工程与自然科学研究理事会;
关键词
D O I
10.1364/OE.14.003345
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A novel combination of adaptive optics with a simultaneous en-face OCT/SLO system for high resolution imaging of the in-vivo human retina is presented. Pairs of retinal images are shown and performance of the system is evaluated with and without dynamic wavefront correction. The adaptive optics closed loop system operates at a frame rate of 9 Hz and incorporates a Shack-Hartmann wavefront sensor based on a highly sensitive Andor camera and a 37 actuator OKO membrane deformable mirror to correct for ocular aberrations. The system produces C-scan pairs of images at a frame rate of 2 Hz. The correction of aberrations produced by the adaptive optics closed-loop system increased the signal-to-noise ratio in images obtained from volunteer eyes by up to 6 dB in the OCT channel and up to 9 dB in the SLO channel. A slight improvement in the lateral resolution was also obtained, from 6.5 mu m before to 5 mu m after closing the adaptive optics loop. (c) 2006 Optical Society of America.
引用
收藏
页码:3345 / 3353
页数:9
相关论文
共 50 条
  • [1] Adaptive Optics Loop for en-face Optical Coherence Tomography and Laser Scanning Confocal Microscopy
    Tuohy, Simon
    Bradu, Adrian
    Harms, Fabrice
    Chateau, Nicolas
    Podoleanu, Adrian Gh.
    1ST CANTERBURY WORKSHOP ON OPTICAL COHERENCE TOMOGRAPHY AND ADAPTIVE OPTICS, 2008, 7139
  • [2] En-face optical coherence tomography
    Podoleanu, AG
    Rogers, JA
    Dunne, S
    Jackson, DA
    SARATOV FALL MEETING 2000: OPTICAL TECHNOLOGIES IN BIOPHYSICS AND MEDICINE II, 2001, 4241 : 174 - 181
  • [3] En-face optical coherence tomography
    Podoleanu, AG
    Rosen, RB
    Rogers, JA
    Dobre, GM
    Cucu, RG
    Jackson, DA
    Dunne, S
    Amaechi, BT
    SARATOV FALL MEETING 2002: OPTICAL TECHNOLOGIES IN BIOPHYSICS AND MEDICINE IV, 2002, 5068 : 248 - 254
  • [4] Retinal Imaging with a Combined Adaptive Optics - Optical Coherence Tomography and Adaptive Optics - Scanning Laser Ophthalmoscopy System
    Zawadzki, Robert J.
    Jones, Steven M.
    Pilli, Suman
    Kim, Dae Yu
    Olivier, Scot S.
    Werner, John S.
    OPHTHALMIC TECHNOLOGIES XX, 2010, 7550
  • [5] Combined Adaptive Optics - Optical Coherence Tomography and Adaptive Optics - Scanning Laser Ophthalmoscopy system for retinal imaging
    Zawadzki, Robert J.
    Jones, Steven M.
    Chen, Diana
    Choi, Stacey S.
    Evans, Julia W.
    Olivier, Scot S.
    Werner, John S.
    OPHTHALMIC TECHNOLOGIES XIX, 2009, 7163
  • [6] Simultaneous imaging of human cone mosaic with adaptive optics enhanced scanning laser ophthalmoscopy and high-speed transversal scanning optical coherence tomography
    Pircher, M.
    Zawadzki, R. J.
    Evans, J. W.
    Werner, J. S.
    Hitzenberger, C. K.
    OPTICS LETTERS, 2008, 33 (01) : 22 - 24
  • [7] En-face coherence gating of the retina with adaptive optics
    Jonnal, RS
    Qu, J
    Thorn, K
    Miller, DT
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 : U275 - U275
  • [8] En-face optical coherence tomography revival
    Bradu, Adrian
    Kapinchev, Konstantin
    Barnes, Frederick
    Podoleanu, Adrian Gh.
    HIGH-END MEDICINE BASED ON LASER AND BIOTECHNOLOGIES, 2016, 9670
  • [9] Handheld simultaneous scanning laser ophthalmoscopy and optical coherence tomography system
    LaRocca, Francesco
    Nankivil, Derek
    Farsiu, Sina
    Izatt, Joseph A.
    BIOMEDICAL OPTICS EXPRESS, 2013, 4 (11): : 2307 - 2321
  • [10] Comparison of retinal vessel measurements using adaptive optics scanning laser ophthalmoscopy and optical coherence tomography
    Arichika, Shigeta
    Uji, Akihito
    Ooto, Sotaro
    Muraoka, Yuki
    Yoshimura, Nagahisa
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 2016, 60 (03) : 166 - 171