Intravascular optical coherence tomography imaging at 3200 frames per second

被引:83
|
作者
Wang, Tianshi [1 ]
Wieser, Wolfgang [2 ]
Springeling, Geert [1 ]
Beurskens, Robert [1 ]
Lancee, Charles T. [1 ]
Pfeiffer, Tom [2 ]
van der Steen, Antonius F. W. [1 ,3 ]
Huber, Robert [2 ]
van Soest, Gijs [1 ]
机构
[1] Erasmus Univ, Med Ctr, Ctr Thorax, NL-3000 CA Rotterdam, Netherlands
[2] Univ Munich, Fak Phys, Lehrstuhl BioMol Opt, D-80538 Munich, Germany
[3] Interuniv Cardiol Inst Netherlands, NL-3051 DG Utrecht, Netherlands
基金
欧洲研究理事会;
关键词
IN-VIVO; ULTRASOUND; ENDOSCOPE; CATHETER;
D O I
10.1364/OL.38.001715
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We demonstrate intravascular optical coherence tomography (OCT) imaging with frame rate up to 3.2 kHz (192,000 rpm scanning). This was achieved by using a custom-built catheter in which the circumferential scanning was actuated by a 1.0 mm diameter synchronous motor. The OCT system, with an imaging depth of 3.7 mm (in air), is based on a Fourier domain mode locked laser operating at an A-line rate of 1.6 MHz. The diameter of the catheter is 1.1 mm at the tip. Ex vivo images of human coronary artery (78.4 mm length) were acquired at a pullback speed of 100 mm/s. True 3D volumetric imaging of the entire artery, with dense and isotropic sampling in all dimensions, was performed in < 1 second acquisition time. (C) 2013 Optical Society of America
引用
收藏
页码:1715 / 1717
页数:3
相关论文
共 50 条
  • [1] Ultrahigh-speed intravascular optical coherence tomography imaging at 3200 frames per second
    Wang, Tianshi
    Wieser, Wolfgang
    Springeling, Geert
    Beurskens, Robert
    Lancee, Charles T.
    Pfeiffer, Tom
    van der Steen, Antonius F. W.
    Huber, Robert
    van Soest, Gijs
    OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE TECHNIQUES VI, 2013, 8802
  • [2] Intravascular optical coherence tomography: Cellular imaging
    Briain D. MacNeill
    Brett E. Bouma
    Hiroshi Yabushita
    Ik-Kyung Jang
    Guillermo J. Tearney
    Journal of Nuclear Cardiology, 2005, 12 : 460 - 465
  • [3] Intravascular optical coherence tomography: Cellular imaging
    MacNeill, BD
    Bouma, BE
    Yabushita, H
    Jang, LK
    Tearney, GJ
    JOURNAL OF NUCLEAR CARDIOLOGY, 2005, 12 (04) : 460 - 465
  • [4] In vivo intravascular photoacoustic imaging at a high speed of 100 frames per second
    Xie, Zhihua
    Shu, Chengyou
    Yang, Daya
    Chen, Hao
    Chen, Chaobin
    Dai, Gang
    Lam, Kwok Ho
    Zhang, Jinke
    Wang, Xiatian
    Sheng, Zonghai
    Gao, Duyang
    Liu, Chengbo
    Song, Liang
    Gong, Xiaojing
    BIOMEDICAL OPTICS EXPRESS, 2020, 11 (11): : 6721 - 6731
  • [5] High-speed time-domain optical coherence tomography with an imaging speed of ten frames per second with 2000 A-scan
    Kim, Young Kwan
    Kim, Yong Pyung
    OPTICAL ENGINEERING, 2010, 49 (05)
  • [6] Intravascular imaging in acute coronary syndrome Intravascular ultrasound and optical coherence tomography
    Mrevlje, B.
    Legutko, J.
    Tillwich, F.
    Jakala, J.
    Dudek, D.
    Noc, M.
    Rzeszutko, L.
    Kleczynski, P.
    Birkemeyer, R.
    Aboukoura, M.
    Nienaber, C. A.
    KARDIOLOGE, 2015, 9 (01): : 47 - 53
  • [7] Intravascular optical coherence tomography
    Tearney, Guillermo J.
    EUROPEAN HEART JOURNAL, 2018, 39 (41) : 3685 - 3686
  • [8] Intravascular optical coherence tomography
    Bouma, Brett E.
    Villiger, Martin
    Otsuka, Kenichiro
    Oh, Wang-Yuhl
    BIOMEDICAL OPTICS EXPRESS, 2017, 8 (05): : 2660 - 2686
  • [9] Intravascular ultrasound and optical coherence tomography imaging of coronary atherosclerosis
    Costopoulos, Charis
    Brown, Adam J.
    Teng, Zhongzhao
    Hoole, Stephen P.
    West, Nick E. J.
    Samady, Habib
    Bennett, Martin R.
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2016, 32 (01): : 189 - 200
  • [10] Method for parametric imaging of attenuation by intravascular optical coherence tomography
    ZHENG, S. U. N.
    FEI, Y. A. N. G.
    JIAN, S. U. N.
    BIOMEDICAL OPTICS EXPRESS, 2021, 12 (04) : 1882 - 1904