A high spatial resolution Stokes polarimeter for motional Stark effect imaging

被引:7
|
作者
Thorman, Alex [1 ]
Michael, Clive [1 ]
Howard, John [1 ]
机构
[1] Australian Natl Univ, Plasma Res Lab, Res Sch Phys & Engn, Canberra, ACT 0200, Australia
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2013年 / 84卷 / 06期
关键词
PLASMA;
D O I
10.1063/1.4811738
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe an enhanced temporally switched interfero-polarimeter that has been successfully deployed for high spatial resolution motional Stark effect imaging on the KSTAR superconducting tokamak. The system utilizes dual switching ferroelectric liquid crystal waveplates to image the full Stokes vector of elliptically polarized and Doppler-shifted Stark-Zeeman Balmer-alpha emission from high energy neutral beams injected into the magnetized plasma. We describe the optical system and compare its performance against a Mueller matrix model that takes account of non-ideal performance of the switching ferro-electric liquid crystal waveplates and other polarizing components. (C) 2013 AIP Publishing LLC.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Spatial heterodyne Stokes vector imaging of the motional Stark-Zeeman multiplet
    Howard, John
    Chung, Jinil
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2012, 83 (10):
  • [2] The motional Stark effect polarimeter in the HL-2A tokamak
    Yu, D. L.
    Wei, Y. L.
    Xia, F.
    Cao, J. Y.
    Chen, C. Y.
    Liu, L.
    Chen, W. J.
    Ji, X. Q.
    Liu, Y.
    Yan, L. W.
    Yang, Q. W.
    Duan, X. R.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2014, 85 (05):
  • [3] Pedestal magnetic field measurements using a motional Stark effect polarimeter
    Lanctot, M. J.
    Holcomb, C. T.
    Allen, S. L.
    Fenstermacher, M. E.
    Luce, T. C.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2012, 83 (10):
  • [4] Imaging motional Stark effect measurements at ASDEX Upgrade
    Ford, O. P.
    Burckhart, A.
    McDermott, R.
    Puetterich, T.
    Wolf, R. C.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (11):
  • [6] Imaging polarimeter with high angle resolution
    Nitzsche, G
    Richter, W
    Fuhrmeister, T
    Rex, S
    DESIGN AND ENGINEERING OF OPTICAL SYSTEMS II, 1999, 3737 : 325 - 331
  • [7] Current profile measurements with motional Stark effect polarimeter in the JT-60U tokamak
    Fujita, T.
    Kuko, H.
    Sugie, T.
    Isei, N.
    Ushigusa, K.
    Fusion Engineering and Design, 1997, 34-35 : 289 - 292
  • [8] Current profile measurements with motional Stark effect polarimeter in the JT-60U tokamak
    Fujita, T
    Kuko, H
    Sugie, T
    Isei, N
    Ushigusa, K
    FUSION ENGINEERING AND DESIGN, 1997, 34-35 : 289 - 292
  • [9] The prototype imaging motional Stark effect diagnostic for ASDEX upgrade
    Ford, O. P.
    Howard, J.
    Wolf, R. C.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2015, 86 (09):
  • [10] The Galway astronomical Stokes polarimeter: an all-Stokes optical polarimeter with ultra-high time resolution
    Patrick Collins
    Gillian Kyne
    David Lara
    Michael Redfern
    Andy Shearer
    Brendan Sheehan
    Experimental Astronomy, 2013, 36 : 479 - 503