RADIATION PROTECTION STUDIES FOR MEDICAL PARTICLE ACCELERATORS USING FLUKA MONTE CARLO CODE

被引:9
|
作者
Infantino, Angelo [1 ]
Cicoria, Gianfranco [2 ]
Lucconi, Giulia [2 ]
Pancaldi, Davide [2 ]
Vichi, Sara [2 ]
Zagni, Federico [2 ]
Mostacci, Domiziano [1 ]
Marengo, Mario [2 ]
机构
[1] Univ Bologna, Dept Ind Engn, Lab Montecuccolino, Via Colli 16, I-40136 Bologna, Italy
[2] S Orsola Malpighi Univ Hosp, Dept Med Phys, Via Massarenti 9, I-40138 Bologna, Italy
关键词
NEUTRON-PRODUCTION; CYCLOTRON; ENERGY; SIMULATIONS; ACTIVATION; SPECTRA; SYSTEM;
D O I
10.1093/rpd/ncw302
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Radiation protection (RP) in the use of medical cyclotrons involves many aspects both in the routine use and for the decommissioning of a site. Guidelines for site planning and installation, as well as for RP assessment, are given in international documents; however, the latter typically offer analytic methods of calculation of shielding and materials activation, in approximate or idealised geometry set-ups. The availability of Monte Carlo (MC) codes with accurate up-to-date libraries for transport and interaction of neutrons and charged particles at energies below 250 MeV, together with the continuously increasing power of modern computers, makes the systematic use of simulations with realistic geometries possible, yielding equipment and site-specific evaluation of the source terms, shielding requirements and all quantities relevant to RP at the same time. In this work, the well-known FLUKA MC code was used to simulate different aspects of RP in the use of biomedical accelerators, particularly for the production of medical radioisotopes. In the context of the Young Professionals Award, held at the IRPA 14 conference, only a part of the complete work is presented. In particular, the simulation of the GE PETtrace cyclotron (16.5 MeV) installed at S. Orsola-Malpighi University Hospital evaluated the effective dose distribution around the equipment; the effective number of neutrons produced per incident proton and their spectral distribution; the activation of the structure of the cyclotron and the vault walls; the activation of the ambient air, in particular the production of 41Ar. The simulations were validated, in terms of physical and transport parameters to be used at the energy range of interest, through an extensive measurement campaign of the neutron environmental dose equivalent using a rem-counter and TLD dosemeters. The validated model was then used in the design and the licensing request of a new Positron Emission Tomography facility.
引用
收藏
页码:185 / 191
页数:7
相关论文
共 50 条
  • [1] Radiation protection studies for the INRNE-BAS cyclotron facility using Monte Carlo FLUKA code
    Demerdjiev, A.
    Dimitrov, D. T.
    Tonev, D.
    Goutev, N.
    Dimitrova, G. D.
    Geleva, E.
    Genchev, S. G.
    RADIATION PHYSICS AND CHEMISTRY, 2023, 212
  • [2] FLUKA Monte Carlo Simulation Code As Used for Radiation Studies in the Alice Experiment
    Pastircak, B.
    PHYSICS OF PARTICLES AND NUCLEI LETTERS, 2008, 5 (03) : 301 - 304
  • [3] A Monte Carlo transport code study of the space radiation environment using FLUKA and ROOT
    Wilson, T
    Pinsky, L
    Carminati, F
    Brun, R
    Ferrari, A
    Sala, P
    Empl, A
    MacGibbon, J
    SPACE TECHNOLOGY AND APPLICATIONS INTERNATIONAL FORUM-2001, 2001, 552 : 1234 - 1239
  • [4] FLUKAVAL - A validation framework for the FLUKA radiation transport Monte Carlo code
    Widorski, Markus
    Bozzato, Davide
    Froeschl, Robert
    Kouskoura, Vasiliki
    15TH INTERNATIONAL CONFERENCE ON NUCLEAR DATA FOR SCIENCE AND TECHNOLOGY, ND2022, 2023, 284
  • [5] Validation of the FLUKA Monte Carlo code for predicting induced radioactivity at high-energy accelerators
    Brugger, M.
    Ferrari, A.
    Roesler, S.
    Ulrici, L.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 562 (02): : 814 - 818
  • [6] The FLUKA Monte Carlo code and its applications
    Morone, M. C.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 649 - 650
  • [7] MICRODOSIMETRIC SIMULATIONS OF CARBON IONS USING THE MONTE CARLO CODE FLUKA
    Chiriotti, S.
    Conte, V.
    Colautti, P.
    Selva, A.
    Mairani, A.
    RADIATION PROTECTION DOSIMETRY, 2018, 180 (1-4) : 187 - 191
  • [8] Assessment of the production of medical isotopes using the Monte Carlo code FLUKA: Simulations against experimental measurements
    Infantino, Angelo
    Oehlke, Elisabeth
    Mostacci, Domiziano
    Schaffer, Paul
    Trinczek, Michael
    Hoehr, Cornelia
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2016, 366 : 117 - 123
  • [9] The FLUKA Monte Carlo Code in Ion Beam Therapy
    Rinaldi, I.
    MEDICAL PHYSICS, 2014, 41 (06) : 298 - 298
  • [10] Verification of Monte Carlo transport codes FLUKA, GEANT4 and SHIELD for radiation protection purposes at relativistic heavy ion accelerators
    Beskrovnaia, L.
    Florko, B.
    Paraipan, M.
    Sobolevsky, N.
    Timoshenko, G.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2008, 266 (18): : 4058 - 4060