Standardless, Automated Determination of Chlorine-35 by 35Cl Nuclear Magnetic Resonance

被引:5
|
作者
Watson, Simon A. [1 ]
Edwards, Andy J. [1 ]
Parkinson, John A. [2 ]
机构
[1] GlaxoSmithKline Medicines Res Ctr, Gunnels Wood Rd, Stevenage SG1 2NY, Herts, England
[2] Univ Strathclyde, Dept Pure & Appl Chem, WestCHEM, Glasgow, Lanark, Scotland
关键词
Cl-35; chloride; nuclear magnetic resonance; pharmaceutical analysis; NMR-SPECTROSCOPY; SIGNAL;
D O I
10.1080/00032719.2016.1172636
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A robust, fully automated, walk-up method is reported to quantify chloride in samples using Cl-35 nuclear magnetic resonance. Minimal user input is required, no standards are acquired at the time of analysis; and the submission, acquisition, processing, and production of results are seamlessly integrated within existing software. The method demonstrated good linearity with R-2=0.999 over three orders of magnitude of analyte concentration. The results were highly independent of analyte functionality, and the stability of instrument response was sufficient that analyses of additional standards were not required for a period of several months. At a nominal sample concentration of 10mg/ml in D2O at 400MHz, detection and quantitation limits of 0.1 and 0.5% (w/w) were achieved in a 1-h analysis time. Robust methodology was achieved by applying a rigorous approach to method development and validation to determine and evaluate fully the time- and sample-dependent factors that affect quantitation in these measurements.