Performance of Vitek 2 for Antimicrobial Susceptibility Testing of Staphylococcus spp. and Enterococcus spp.

被引:59
|
作者
Bobenchik, April M. [1 ]
Hindler, Janet A. [2 ]
Giltner, Carmen L. [1 ]
Saeki, Sandra [2 ]
Humphries, Romney M. [1 ]
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, Dept Pathol & Lab Med, Los Angeles, CA 90095 USA
[2] UCLA Hlth Syst, Los Angeles, CA USA
关键词
INDUCIBLE CLINDAMYCIN RESISTANCE; BD PHOENIX; AUREUS; DAPTOMYCIN; PREVALENCE; FAECIUM; EVOLUTION; COMMUNITY; ACCURACY; FAILURE;
D O I
10.1128/JCM.02432-13
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Vitek 2 (bioMerieux, Inc., Durham, NC) is a widely used commercial antimicrobial susceptibility testing system. We compared MIC results obtained by Vitek 2 to those obtained by the Clinical and Laboratory Standards Institute (CLSI) broth microdilution (BMD) reference method for 134 staphylococcal and 84 enterococcal clinical isolates. Nineteen agents were evaluated, including all those available on Vitek 2 for testing staphylococci and enterococci. The resistance phenotypes tested included methicillin-resistant Staphylococcus aureus (MRSA) (n = 58), S. aureus with inducible clindamycin resistance (ICR) (n = 30), trimethoprim-sulfamethoxazole-resistant MRSA (n = 10), vancomycin-resistant Enterococcus (n = 37), high-level gentamicin-resistant Enterococcus (n = 15), linezolid-resistant Enterococcus (n = 5), and daptomycin-nonsusceptible Enterococcus faecalis (n = 6). For the staphylococci, there was 98.9% categorical agreement (CA). There was one very major error (VME) for gentamicin in a Staphylococcus hominis isolate, six VMEs for inducible clindamycin in S. aureus isolates, and two major errors (ME) for daptomycin in an S. aureus and a Staphylococcus epidermidis isolate. For enterococci, there was 97.3% CA. Two VMEs were observed for daptomycin in isolates of E. faecalis and 2 ME, 1 for high-level gentamicin resistance and 1 for nitrofurantoin, in E. faecium isolates. Overall, there was 98.3% CA and 99% essential agreement for the testing of staphylococci and enterococci by the Vitek 2. With the exception of detecting ICR in S. aureus, Vitek 2 performed reliably for antimicrobial susceptibility testing of staphylococci and enterococci.
引用
收藏
页码:392 / 397
页数:6
相关论文
共 50 条
  • [1] Evaluation of the Vitek 2 system for identification and antimicrobial susceptibility testing of Staphylococcus spp.
    Nonhoff, C
    Rottiers, S
    Struelens, MJ
    CLINICAL MICROBIOLOGY AND INFECTION, 2005, 11 (02) : 150 - 153
  • [2] Two-center collaborative evaluation of the performance of the BD Phoenix Automated Microbiology System for identification and antimicrobial susceptibility testing of Enterococcus spp. and Staphylococcus spp.
    Fahr, AM
    Eigner, U
    Armbrust, M
    Caganic, A
    Dettori, G
    Chezzi, C
    Bertoncini, L
    Benecchi, M
    Menozzi, MG
    JOURNAL OF CLINICAL MICROBIOLOGY, 2003, 41 (03) : 1135 - 1142
  • [3] Isolation and Antimicrobial Testing of Aeromonas spp., Citrobacter spp., Cronobacter spp., Enterobacter spp., Escherichia spp., Klebsiella spp., and Trabulsiella spp. from the Gallbladder of Pigs
    Evangelopoulou, Grammato
    Filioussis, Georgios
    Kritas, Spyridon
    Kantere, Maria
    Burriel, Angeliki R.
    POLISH JOURNAL OF MICROBIOLOGY, 2015, 64 (02) : 185 - 188
  • [4] Antimicrobial susceptibility and β-lactamase production of Prevotella spp. and Porphyromonas spp.
    Tanaka, K
    Kawamura, C
    Fukui, K
    Kato, H
    Kato, N
    Nakamura, T
    Watanabe, K
    Ueno, K
    ANAEROBE, 1999, 5 (3-4) : 461 - +
  • [5] Assessment of the gradient diffusion method for fosfomycin susceptibility testing in Staphylococcus spp. and Enterococcus spp. isolated from the urine of companion dogs in Thailand
    Jariyapamornkoon, Nattha
    Boonkam, Pongthai
    Suanpairintr, Nipattra
    VETERINARY WORLD, 2023, 16 (12) : 2497 - 2503
  • [6] Etiology and antimicrobial resistance of subclinical mastitis pathogens Staphylococcus aureus, Streptococcus spp. and Enterococcus spp. in sheep milk
    Iancu, Ionica
    Igna, Violeta
    Popa, Sebastian Alexandru
    Imre, Kalman
    Pascu, Corina
    Costinar, Luminita
    Degi, Janos
    Gligor, Alexandru
    Iorgoni, Vlad
    Badea, Corina
    Nichita, Ileana
    Herman, Viorel
    VETERINARY RESEARCH COMMUNICATIONS, 2025, 49 (01)
  • [7] Erroneous Reporting of Vancomycin Susceptibility for Staphylococcus spp. by Vitek Software Version 2.01
    Behera, Bijayini
    Mathur, Purva
    JAPANESE JOURNAL OF INFECTIOUS DISEASES, 2009, 62 (04) : 298 - 299
  • [8] Screening of plants with antimicrobial activity against enterobacteria, Pseudomonas spp. and Staphylococcus spp.
    Girondi, Camila Miorelli
    de Oliveira, Analu Barros
    Prado, Jackeline Alves
    Koga-Ito, Cristiane Yumi
    Borges, Aline Chiodi
    Delbem, Alberto Carlos Botazzo
    Pereira, Daniel Freitas Alves
    Salvador, Marcos Jose
    Brighenti, Fernanda Lourencao
    FUTURE MICROBIOLOGY, 2017, 12 (08) : 671 - 681
  • [9] Susceptibility of Pediococcus spp. to antimicrobial agents
    Danielsen, M.
    Simpson, P. J.
    O'Connor, E. B.
    Ross, R. P.
    Stanton, C.
    JOURNAL OF APPLIED MICROBIOLOGY, 2007, 102 (02) : 384 - 389
  • [10] Susceptibility of Pediococcus spp. to antimicrobial agents
    Danielsen, M.
    Simpson, P.J.
    O'Connor, E.B.
    Ross, R.P.
    Stanton, C.
    Journal of Applied Microbiology, 2007, 102 (02): : 384 - 389