Validation of the Technicon H*1 for the automated cell count and cell differentiation of feline blood samples

被引:0
|
作者
Liedl, R
Hirschberger, J
机构
来源
KLEINTIERPRAXIS | 1997年 / 42卷 / 09期
关键词
D O I
暂无
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
The Technicon H*1 gives a reliable CBC in the cat except the thrombocyte count. The differentiation of neutrophils and lymphocytes is accurate enough for clinical purposes. The differentiation of eosinophiles is not accurate. The monocyte estimations of the Technicon H*1 differ from those of the reference method, a 400 cell leucocyte differential cell count. Storage of the blood at room temperature and at 4 degrees C is possible for three days, but storage at room temperature for seven days results in a significant decrease of neutrophils.
引用
收藏
页码:703 / &
页数:5
相关论文
共 50 条
  • [1] EVALUATION OF THE TECHNICON H.1, AN AUTOMATED COMPLETE BLOOD-COUNT AND DIFFERENTIAL ANALYZER
    CARLSON, D
    VELAZQUEZ, F
    DAIGNEAULT, R
    STATLAND, BE
    CLINICAL CHEMISTRY, 1986, 32 (06) : 1173 - 1173
  • [2] A COMPARISON OF THE DIFFERENTIAL WHITE BLOOD-CELL COUNT USING THE TECHNICON-H1, THE TECHNICON-H6000 AND THE COULTER-VCS
    STOLSNES, B
    SKARE, EV
    TAVERNER, P
    OMMUNDSEN, T
    AGA, IB
    SANDBERG, S
    EUROPEAN JOURNAL OF HAEMATOLOGY, 1990, 45 : 37 - 39
  • [3] COMPARISON OF AUTOMATED DIFFERENTIAL BLOOD-CELL COUNTING WITH THE TECHNICON-H1 AND THE COULTER-VCS
    KATZ, N
    MARTI, B
    LENZ, T
    JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1988, 26 (11): : 753 - 753
  • [4] TECHNICON H-1 AUTOMATED WHITE CELL PARAMETERS IN THE DIAGNOSIS OF MEGALOBLASTIC ERYTHROPOIESIS
    TAYLOR, C
    BAIN, BJ
    EUROPEAN JOURNAL OF HAEMATOLOGY, 1991, 46 (04) : 248 - 249
  • [5] ESTIMATION OF CD4 (T4) CELL COUNT USING THE TECHNICON H-1 AUTOMATED HEMATOLOGY ANALYZER
    IRANI, MS
    BANEZ, EI
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1988, 90 (04) : 511 - 511
  • [6] THE TECHNICON H-STAR-1 - AN AUTOMATED HEMATOLOGY ANALYZER FOR TODAY AND TOMORROW - COMPLETE BLOOD-COUNT PARAMETERS
    BOLLINGER, PB
    DREWINKO, B
    BRAILAS, CD
    SMEETON, NA
    TRUJILLO, JM
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1987, 87 (01) : 71 - 78
  • [7] Validation of the Celltac alpha automated hematology analyzer for canine and feline blood samples
    McDaniel, Barbara Jane
    Hirschberger, Johannes
    Weber, Karin
    VETERINARY CLINICAL PATHOLOGY, 2013, 42 (01) : 11 - 18
  • [8] Pseudobasophilia on the technicon automated cell counters
    Bizzaro, N
    CLINICAL AND LABORATORY HAEMATOLOGY, 1996, 18 (04): : 298 - 299
  • [9] LEUKEMIC BLASTS DETECTED BY THE TECHNICON H-1 BLOOD-CELL COUNTER
    KAWARABAYASHI, K
    TSUDA, I
    TATSUMI, N
    OKUDA, K
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1987, 88 (05) : 624 - 627
  • [10] COMPARATIVE-STUDY OF THE DIFFERENTIAL WHITE BLOOD-CELL COUNT USING 3 AUTOMATED ANALYZERS - COULTER STKS, SYSMEX NE-8000 AND TECHNICON H-1
    LESESVE, JF
    LACOMBE, F
    BOISSEAU, MR
    BERNARD, P
    ANNALES DE BIOLOGIE CLINIQUE, 1994, 52 (04) : 295 - 297