Quantitative assessment of colon distention for polyp detection in CT virtual colonoscopy

被引:0
|
作者
Van Uitert, Robert [1 ]
Bitter, Ingmar [1 ]
Summers, Ronald M. [1 ]
Choi, J. Richard [2 ,3 ]
Pickhardt, Perry J. [4 ]
机构
[1] NIH, Dept Diagnost Radiol, Bethesda, MD 20892 USA
[2] Uniformed Serv Univ Hlth Sci, Bethesda, MD 20814 USA
[3] Walter Reed Army Med Ctr, Washington, DC 20307 USA
[4] Univ Wisconsin, Dept Radiol, Madison, WI 53792 USA
关键词
CT; colon; 3D reconstruction; colon cancer; image processing; computer-aided detection;
D O I
10.1117/12.653205
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Virtual colonoscopy is becoming a more prevalent way to diagnose colon cancer. One of the critical elements in detecting cancerous polyps using virtual colonoscopy, especially in conjunction with computer-aided detection of polyps, is that the colon be sufficiently distended. We have developed an automatic method to determine from a CT scan what percentage of the colon is distended by 1 cm or larger and compared our method with a radiologist's assessment of quality of the scan with respect to successful colon polyp detection. A radiologist grouped 41 CT virtual colonoscopy scans into three groups according to the degree of colonic distention, "well", "medium", and "poor". We also employed a subvoxel accurate centerline algorithm and a subvoxel accurate distance transform to each dataset to measure the colon distention along the centerline. To summarize the colonic distention with a single value relevant for polyp detection, the distention score, we recorded the percentage of centerline positions in which the colon distention was 1 cm or larger. We then compared the radiologist's assessment and the computed results. The sorting of all datasets according to the distention score agreed with the radiologist's assessment. The "poor" cases had a mean and standard deviation score of 78.4% +/- 5.2%, the "medium" cases measured 88.7% +/- 1.9%, and the "well" cases 98.8% +/- 1.5%. All categories were shown to be significantly different from each other using impaired two sample t-tests. The presented colonic distention score is an accurate method for assessing the quality of colonic distention for CT colonography.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Impact of multislice CT on polyp detection in virtual colonoscopy
    Rogalla, P
    Hein, E
    Vogel, F
    Kopka, L
    Hamm, BK
    [J]. RADIOLOGY, 2000, 217 : 369 - 369
  • [2] Improved Colon Navigation for Efficient Polyp Detection in Virtual Colonoscopy
    Ismail, Marwa
    Farag, Aly
    Hassouna, M. Sabry
    Dryden, Gerald
    Falk, Robert
    [J]. ABDOMINAL IMAGING: COMPUTATION AND CLINICAL APPLICATIONS, 2013, 8198 : 117 - 125
  • [3] Colonic wall thickness using level sets for CT virtual Colonoscopy visual assessment and polyp detection
    Van Uitert, Robert L.
    Summers, Ronald M.
    [J]. MEDICAL IMAGING 2007: PHYSIOLOGY, FUNCTION, AND STRUCTURE FROM MEDICAL IMAGES, 2007, 6511
  • [4] Colorectal polyp detection with CT colography (virtual colonoscopy): A blinded, prospective study
    Hara, AK
    Johnson, CD
    Reed, JE
    MacCarty, RL
    Ahlquist, DA
    Ehman, RL
    [J]. RADIOLOGY, 1996, 201 : 669 - 669
  • [5] Colorectal polyp detection with CT colography (Virtual colonoscopy): Technical and clinical applications
    Hara, AK
    Johnson, CD
    Reed, JE
    Ahlquist, DA
    Nelson, H
    Ehman, RL
    [J]. RADIOLOGY, 1996, 201 : 535 - 535
  • [6] Lines of curvature for polyp detection in virtual colonoscopy
    Zhao, Lingxiao
    Botha, Charl P.
    Bescos, Javier O.
    Truyen, Roel
    Vos, Frans M.
    Post, Frits H.
    [J]. IEEE TRANSACTIONS ON VISUALIZATION AND COMPUTER GRAPHICS, 2006, 12 (05) : 885 - 892
  • [7] Impact of different window settings on colon polyp measurements with CT virtual colonoscopy: a phantom study
    Wu, Xing-Wang
    Wang, Wan-Qin
    Xu, Jian-Ming
    Liu, Bin
    [J]. CLINICAL IMAGING, 2011, 35 (04) : 274 - 278
  • [8] Virtual colonoscopy with computer-assisted polyp detection
    Vining, DJ
    Ge, Y
    Ahn, DK
    Stelts, DR
    [J]. COMPUTER-AIDED DIAGNOSIS IN MEDICAL IMAGING, 1999, 1182 : 445 - 452
  • [9] Reduction of false positives by internal features for polyp detection in CT-based virtual colonoscopy
    Wang, ZA
    Liang, ZR
    Li, LH
    Li, X
    Li, B
    Anderson, J
    Harrington, D
    [J]. MEDICAL PHYSICS, 2005, 32 (12) : 3602 - 3616
  • [10] Comparison of Optical Colonoscopy and CT Colonography for Polyp Detection
    Dachman, Abraham H.