The feasibility of head motion tracking in helical CT: A step toward motion correction

被引:13
|
作者
Kim, Jung-Ha [1 ]
Nuyts, Johan [2 ,3 ]
Kuncic, Zdenka [4 ]
Fulton, Roger [1 ,4 ,5 ]
机构
[1] Univ Sydney, Sydney, NSW 2141, Australia
[2] Katholieke Univ Leuven, Dept Nucl Med, Louvain, Belgium
[3] Katholieke Univ Leuven, Med Imaging Res Ctr, Louvain, Belgium
[4] Univ Sydney, Sch Phys, Sydney, NSW 2006, Australia
[5] Westmead Hosp, Dept Med Phys, Westmead, NSW 2145, Australia
基金
英国医学研究理事会;
关键词
computed tomography (CT); head motion; motion artifact; motion tracking; PET; IMPLEMENTATION; TOMOGRAPHY;
D O I
10.1118/1.4794481
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To establish a practical and accurate motion tracking method for the development of rigid motion correction methods in helical x-ray computed tomography (CT). Methods: A commercially available optical motion tracking system provided 6 degrees of freedom pose measurements at 60 Hz. A 4 x 4 calibration matrix was determined to convert raw pose data acquired in tracker coordinates to a fixed CT coordinate system with origin at the isocenter of the scanner. Two calibration methods, absolute orientation (AO), and a new method based on image registration (IR), were compared by means of landmark analysis and correlation coefficient in phantom images coregistered using the derived motion transformations. Results: Transformations calculated using the IR-derived calibration matrix were found to be more accurate, with positional errors less than 0 5 mm (mean RMS), and highly correlated image voxel intensities. The AO-derived calibration matrix yielded larger mean RMS positional errors (similar or equal to 1.0 mm), and poorer correlation coefficients. Conclusions: The authors have demonstrated the feasibility of accurate motion tracking for retrospective motion correction in helical CT. Their new IR-based calibration method based on image registration and function minimization was simpler to perform and delivered more accurate calibration matrices. This technique is a useful tool for future work on rigid motion correction in helical CT and potentially also other imaging modalities. (C) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4794481]
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Correction for human head motion in helical x-ray CT
    Kim, J-H
    Sun, T.
    Alcheikh, A. R.
    Kuncic, Z.
    Nuyts, J.
    Fulton, R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (04): : 1416 - 1438
  • [2] Head Motion Correction Based on Filtered Backprojection in Helical CT Scanning
    Jang, Seokhwan
    Kim, Seungeon
    Kim, Mina
    Son, Kihong
    Lee, Kyoung-Yong
    Ra, Jong Beom
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2020, 39 (05) : 1636 - 1645
  • [3] Correction of motion tracking errors for PET head rigid motion correction
    Miranda, Alan
    Kroll, Tina
    Schweda, Vanessa
    Staelens, Steven
    Verhaeghe, Jeroen
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (17):
  • [4] A Method of Motion Tracking during CT for Motion Correction
    Kim, Jung-Ha
    Nuyts, Johan
    Kuncic, Zdenka
    Fulton, Roger
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3816 - 3819
  • [5] Breathing-Motion Correction for Helical CT
    van Stevendaal, Udo
    Klinder, Tobias
    Lorenz, Cristian
    Koehler, Thomas
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 4707 - 4710
  • [6] Cardiac-Motion Correction for Helical CT
    Isola, Alfonso A.
    van Stevendaal, Udo
    Schaefer, Dirk
    Niessen, Wiro J.
    Grass, Michael
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3247 - +
  • [7] Correction of Breathing Motion in the Thorax for Helical CT
    Thomas Khler
    Tobias Klinder
    Udo van Stevendaal
    Cristian Lorenz
    Peter Forthmann
    TsinghuaScienceandTechnology, 2010, 15 (01) : 87 - 95
  • [8] A head motion estimation algorithm for motion artifact correction in dental CT imaging
    Hernandez, Daniel
    Eldib, Mohamed Elsayed
    Hegazy, Mohamed A. A.
    Cho, Myung Hye
    Cho, Min Hyoung
    Lee, Soo Yeol
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (06):
  • [9] Quantification of Local Reconstruction Accuracy for Helical CT with Motion Correction
    Sun, Tao
    Clackdoyle, Rolf
    Fulton, Roger
    Nuyts, Johan
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [10] Cardiac Motion Correction for Helical CT Scan With an Ordinary Pitch
    Kim, Seungeon
    Chang, Yongjin
    Ra, Jong Beom
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (07) : 1587 - 1596