Determination of complex microcalorimeter parameters with impedance measurements

被引:12
|
作者
Saab, T [1 ]
Bandler, SR
Chervenak, J
Figueroa-Feliciano, E
Finkbeiner, F
Iyomoto, N
Kelley, RL
Kilbourne, CA
Lindeman, MA
Porter, FS
Sadleir, J
机构
[1] NASA, Goddard Space Flight Ctr, Greenbelt, MD 20771 USA
[2] Univ Wisconsin, Madison, WI 53706 USA
关键词
X-ray spectrometer; microcalorimeter; impedance measurements;
D O I
10.1016/j.nima.2005.12.112
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The proper understanding and modeling of a microcalorimeter's response requires accurate knowledge of a handful of parameters, such as C, G, alpha. While a few of these parameters are directly determined from the TV characteristics, some others, notoriously the heat capacity (C) and alpha, appear in degenerate combinations in most measurable quantities. The consideration of a complex microcalorimeter leads to an added ambiguity in the determination of the parameters. In general, the dependence of the microcalorimeter's complex impedance on these various parameters varies with frequency. This dependence allows us to determine individual parameters by fitting the prediction of the microcalorimeter model to impedance data. In this paper we describe efforts at characterizing the Goddard X-ray microcalorimeters. With the parameters determined by this method, we compare the pulse shape and noise spectra predictions to data taken with the same devices. (c) 2005 Elsevier B.V. All rights reserved.