Comparison of Push-In versus Pull-Out Tests on Bone-Implant Interfaces of Rabbit Tibia Dental Implant Healing Model

被引:29
|
作者
Seong, Wook-Jin [1 ]
Grami, Shahrzad [2 ]
Jeong, Soo Cheol [1 ]
Conrad, Heather J. [1 ]
Hodges, James S. [3 ]
机构
[1] Univ Minnesota, Div Prosthodont, Dept Restorat Sci, Sch Dent, Minneapolis, MN 55455 USA
[2] Univ Minnesota, Dept Ind & Syst Engn, Coll Sci & Engn, Minneapolis, MN 55455 USA
[3] Univ Minnesota, Sch Publ Hlth, Div Biostat, Minneapolis, MN 55455 USA
关键词
bone-implant interface; pull-out test; push-in test; rabbit tibia healing model; RESONANCE FREQUENCY-ANALYSIS; TITANIUM IMPLANTS; VIVO; STABILITY; OSSEOINTEGRATION; STRENGTH;
D O I
10.1111/j.1708-8208.2011.00357.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: This study aimed to investigate whether push-in and pull-out tests measure mechanical properties of the bone-implant interface differently, and which test is more sensitive to changes over the healing period. Materials and Methods: Two identical self-threading dental implants (3.3x8.5mm) were placed in medial surface of the proximal condyles of left and right tibias of 20 rabbits (40 implants total). Five rabbits each were sacrificed after 1, 4, 8, and 12 weeks of healing. Push-in test was performed on one side's tibia implant and pull-out on the other side's implant, at a rate of 6mm/min. Primary and secondary implant stabilities and tibia weight were measured on all implants. Results: The push-in test generated significantly higher failure load (p=.0001; 530N vs 279N), lower displacement at failure (p=.0003; 0.436mm vs 0.680mm), and higher interface stiffness (p<.0001; 1,641N/mm vs 619N/mm) than pull-out test. Failure load, stiffness, and secondary implant stability were significantly higher for longer compared with shorter healing periods, while displacement, tibia weight, and primary stability were not. Failure load and stiffness differed significantly for four healing times for the push-in but not for the pull-out test. Failure load was significantly correlated with secondary implant stability for both push-in (r=0.66) and pull-out (r=0.48) tests, but stiffness was significantly correlated with secondary stability only for the push-in test (r=0.72; pull-out test r=0.40). Conclusion: The push-in test appeared more sensitive than pull-out to changes in mechanical properties at bone-implant interfaces during healing in rabbit tibia model.
引用
收藏
页码:460 / 469
页数:10
相关论文
共 6 条
  • [1] Nondestructive measurements of implant-bone interface shear modulus and effects of implant geometry in pull-out tests
    Berzins, A
    Shah, B
    Weinans, H
    Sumner, DR
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1997, 34 (03): : 337 - 340
  • [2] Effect of the Acid-Etching on Grit-Blasted Dental Implants to Improve Osseointegration: Histomorphometric Analysis of the Bone-Implant Contact in the Rabbit Tibia Model
    Rios-Carrasco, Blanca
    Lemos, Bernardo Ferreira
    Herrero-Climent, Mariano
    Gil Mur, F. Javier
    Rios-Santos, Jose Vicente
    COATINGS, 2021, 11 (11)
  • [3] Comparison of mechanical behavior between implant-simulated bone tissue and implant-jaw bone tissue interfaces based on Pull Out testing
    Lopez, C.
    Munoz, J. C.
    Pinillos, J. C.
    2ND INTERNATIONAL MEETING FOR RESEARCHERS IN MATERIALS AND PLASMA TECHNOLOGY, 2013, 466
  • [4] Mechanical properties of bone-implant interface: An in vitro model for the comparison of stability parameters affecting various stages during osseointegration for dental implant
    Pan, RLC
    Ying, SH
    PROCEEDINGS OF THE 26TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2004, 26 : 5050 - 5052
  • [5] Comparison Between Bioactive Fluoride Modified and Bioinert Anodically Oxidized Implant Surfaces in Early Bone Response Using Rabbit Tibia Model
    Choi, Jung-Yoo
    Lee, Hyo-Jung
    Jang, Jae-Up
    Yeo, In-Sung
    IMPLANT DENTISTRY, 2012, 21 (02) : 124 - 128
  • [6] Effects of magnesium coating on bone-implant interfaces with and without polyether-ether-ketone particle interference: A rabbit model based on porous Ti6Al4V implants
    Du, Zhe
    Yu, Xiaoming
    Nie, Binen
    Zhu, Zhonglin
    Ibrahim, Muhammad
    Yang, Ke
    Tan, Lili
    Wang, You
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2019, 107 (07) : 2388 - 2396