A Simple Dewar/Cryostat for Thermally Equilibrating Samples at Known Temperatures for Accurate Cryogenic Luminescence Measurements

被引:0
|
作者
Weaver, Phoebe G. [1 ]
Jagow, Devin M. [1 ]
Portune, Cameron M. [1 ]
Kenney, John W., III [1 ]
机构
[1] Concordia Univ, Chem Phys Lab, Montreal, PQ, Canada
来源
关键词
Engineering; Issue; 113; Dewar; cryostat; thermocouple; luminescence; luminescence quenching; activation energy; oxygen quenching of luminescence; MULTIPLE-STATE EMISSION; EXCITED-STATES; COMPLEXES; CRYOSTAT;
D O I
10.3791/54267
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The design and operation of a simple liquid nitrogen Dewar/cryostat apparatus based upon a small fused silica optical Dewar, a thermocouple assembly, and a CCD spectrograph are described. The experiments for which this Dewar/cryostat is designed require fast sample loading, fast sample freezing, fast alignment of the sample, accurate and stable sample temperatures, and small size and portability of the Dewar/cryostat cryogenic unit. When coupled with the fast data acquisition rates of the CCD spectrograph, this Dewar/cryostat is capable of supporting cryogenic luminescence spectroscopic measurements on luminescent samples at a series of known, stable temperatures in the 77-300 K range. A temperature-dependent study of the oxygen quenching of luminescence in a rhodium(III) transition metal complex is presented as an example of the type of investigation possible with this Dewar/cryostat. In the context of this apparatus, a stable temperature for cryogenic spectroscopy means a luminescent sample that is thermally equilibrated with either liquid nitrogen or gaseous nitrogen at a known measureable temperature that does not vary (Delta T < 0.1 K) during the short time scale (similar to 1-10 sec) of the spectroscopic measurement by the CCD. The Dewar/cryostat works by taking advantage of the positive thermal gradient dT/dh that develops above liquid nitrogen level in the Dewar where h is the height of the sample above the liquid nitrogen level. The slow evaporation of the liquid nitrogen results in a slow increase in h over several hours and a consequent slow increase in the sample temperature T over this time period. A quickly acquired luminescence spectrum effectively catches the sample at a constant, thermally equilibrated temperature.
引用
收藏
页数:9
相关论文
共 7 条
  • [1] Accurate Solar Cell Measurements at Low Temperatures using a Cryostat
    Siefer, Gerald
    Fehrenbacher, Elvira
    Schachtner, Michael
    Wekkeli, Alexander
    2019 EUROPEAN SPACE POWER CONFERENCE (ESPC), 2019,
  • [2] A SIMPLE DEWAR FOR SPECTROSCOPIC MEASUREMENTS AT CONTROLLED TEMPERATURES DOWN TO 77 DEGREES K
    SIMONS, JP
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1968, 1 (08): : 872 - &
  • [3] DESIGN OF A SIMPLE CRYOSTAT FOR THERMOLUMINESCENCE MEASUREMENTS ON SOLID STATE LOW THERMAL CONDUCTIVITY SAMPLES
    FLEMING, RJ
    LEGGE, CA
    RANICAR, JH
    BORONKAY, S
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1971, 4 (04): : 286 - &
  • [4] The development of a simple TRL microstrip calibration and test fixture suitable for measurements at cryogenic temperatures
    vanNiekerk, C
    Meyer, P
    AFRICON '96 - 1996 IEEE AFRICON : 4TH AFRICON CONFERENCE IN AFRICA, VOLS I & II: ELECTRICAL ENERGY TECHNOLOGY; COMMUNICATION SYSTEMS; HUMAN RESOURCES, 1996, : 919 - 922
  • [5] A method for accurate thermal conductivity measurements of small samples at ultra-low temperatures
    Orendacova, A.
    Tibenska, K.
    Cizmar, E.
    Orendac, M.
    Feher, A.
    CRYOGENICS, 2007, 47 (01) : 61 - 63
  • [6] Strain-free FBG sensors for accurate temperature measurements from cryogenic temperatures to 300 °C
    Gao, Shenghan
    Swinehart, Philip R.
    Zhong, Shuda
    Wuenschell, Jeffrey
    Lalam, Nageswara
    Buric, Michael
    Chen, Kevin. P.
    OPTICAL WAVEGUIDE AND LASER SENSORS II, 2023, 12532
  • [7] Simple guarded hot plate technique for making thermal conductivity measurements of composite materials at cryogenic temperatures
    Oram, RJ
    Wolff, EG
    ADVANCES IN CRYOGENIC ENGINEERING MATERIALS, VOL 44, PTS A AND B, 1998, 44 : 349 - 357