Effect of differences in pixel size on image characteristics of digital intraoral radiographic systems: a physical and visual evaluation

被引:5
|
作者
Kuramoto, Taku [1 ]
Takarabe, Shinya [1 ]
Okamura, Kazutoshi [2 ]
Shiotsuki, Kenshi [3 ]
Shibayama, Yusuke [1 ]
Tsuru, Hiroki [1 ]
Akamine, Hiroshi [1 ]
Tatsumi, Masato [1 ]
Kato, Toyoyuki [1 ]
Morishita, Junji [4 ]
Yoshiura, Kazunori [2 ]
机构
[1] Kyushu Univ Hosp, Dept Med Technol, Div Radiol, Higashi Ku, 3-1-1 Maidashi, Fukuoka 8128582, Japan
[2] Kyushu Univ, Fac Dent Sci, Dept Oral & Maxillofacial Radiol, Higashi Ku, 3-1-1 Maidashi, Fukuoka 8128582, Japan
[3] Oita Univ Hosp, Med Technol Dept, 1-1 Idaigaoka,Hasama Cho, Yufu, Oita 8795593, Japan
[4] Kyushu Univ, Fac Med Sci, Dept Hlth Sci, Higashi Ku, 3-1-1 Maidashi, Fukuoka 8128582, Japan
关键词
digital intraoral radiographic system; modulation transfer function; normalized noise power spectrum; photostimulable phosphor imaging plate; quality assurance and quality control; MODULATION TRANSFER-FUNCTION; COMPUTED RADIOGRAPHY; DETECTOR PERFORMANCE; DENTAL RADIOGRAPHY; PHOSPHOR; QUALITY; CARIES; MTF; SPECTRA; PLATES;
D O I
10.1259/dmfr.20190378
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To quantify and validate the effect of pixel size on a digital intraoral radiographic system according to International Electrotechnical Commission standards through physical and visual evaluations. Methods: The digital intraoral radiographic system used was the photostimulable phosphor imaging plate and scanner system. The system had three image capture modes: high-speed (HS), high-resolution (HR), and super high-resolution (SHR) with different pixels. The physical characteristics of the system were evaluated using presampled modulation transfer function (MTF) and the normalized noise power spectrum (NNPS). An aluminum (Al) step phantom with different depths of holes was used to acquire images under various exposure condition. The average number of perceptible holes from all steps was plotted against each exposure dose. The results were compared to analyze the effects of pixel size on image quality of intraoral radiographs. Results: The MTF was slightly higher with SHR than with HR and HS. The NNPS with SHR showed about a 40% decrease in magnitude compared to HS. The total number of perceptible holes in the Al step phantom was higher with SHR than with HS and HR in all exposure conditions. Conclusions: The MTF and NNPS obtained with different pixel size could be quantified by physical evaluation, and the differences were visually validated with Al step phantom. The SHR mode has the potential to decrease the radiation dose without compromising the image quality.
引用
收藏
页数:10
相关论文
共 17 条
  • [1] Effects of differences in pixel size on image characteristics of digital intraoral radiographic systems: a physical and visual evaluation
    Watanabe, Hiroshi
    Kurabayashi, Tohru
    DENTOMAXILLOFACIAL RADIOLOGY, 2022, 51 (03)
  • [2] Evaluation of image quality parameters of representative intraoral digital radiographic systems
    Udupa, Hema
    Mah, Peter
    Dove, Stephen B.
    McDavid, William D.
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2013, 116 (06): : 774 - 783
  • [3] THE EFFECT OF PIXEL SIZE ON THE DETECTION RATE OF EARLY PULMONARY SARCOIDOSIS IN DIGITAL CHEST RADIOGRAPHIC SYSTEMS
    MACMAHON, H
    VYBORNY, C
    POWELL, G
    DOI, K
    METZ, CE
    PROCEEDINGS OF THE SOCIETY OF PHOTO-OPTICAL INSTRUMENTATION ENGINEERS, 1984, 486 : 14 - 20
  • [4] Evaluation of resolution characteristics of digital intraoral radiographic images using a task transfer function
    Taku Kuramoto
    Shinya Takarabe
    Hiroki Tsuru
    Yusuke Shibayama
    Toyoyuki Kato
    Kazunori Yoshiura
    Oral Radiology, 2022, 38 : 638 - 644
  • [5] Evaluation of resolution characteristics of digital intraoral radiographic images using a task transfer function
    Kuramoto, Taku
    Takarabe, Shinya
    Tsuru, Hiroki
    Shibayama, Yusuke
    Kato, Toyoyuki
    Yoshiura, Kazunori
    ORAL RADIOLOGY, 2022, 38 (04) : 638 - 644
  • [6] The effect of differences in customer characteristics on the evaluation of a store image
    Bir ticari mekan imajinin müs¸teri karakteristiklerine bagli olarak farkli yorumlanmasi
    Yildirim, K. (kemaly@gazi.edu.tr), 2005, Gazi Universitesi Muhendislik-Mimarlik (20):
  • [7] Numerical Evaluation of Image Contrast for Thicker and Thinner Objects among Current Intraoral Digital Imaging Systems
    Dashpuntsag, Oyunbat
    Yoshida, Midori
    Kasai, Ryosuke
    Maeda, Naoki
    Hosoki, Hidehiko
    Honda, Eiichi
    BIOMED RESEARCH INTERNATIONAL, 2017, 2017
  • [8] Evaluation of the effect scan pattern has on the trueness and precision of six intraoral digital impression systems
    Mennito, Anthony S.
    Evans, Zachary P.
    Lauer, Abigail W.
    Patel, Ravi B.
    Ludlow, Mark E.
    Renne, Walter G.
    JOURNAL OF ESTHETIC AND RESTORATIVE DENTISTRY, 2018, 30 (02) : 113 - 118
  • [9] Physical evaluation of prototype high-performance anti-scatter grids: potential for improved digital radiographic image quality
    Fetterly, Kenneth A.
    Schueler, Beth A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (02): : N37 - N42