Accuracy of Continuous Glucose Monitoring During Exercise in Type 1 Diabetes Pregnancy

被引:30
|
作者
Kumareswaran, Kavita [1 ,2 ]
Elleri, Daniela [1 ,2 ]
Allen, Janet M. [1 ,2 ]
Caldwell, Karen [1 ,2 ]
Nodale, Marianna [1 ,2 ]
Wilinska, Malgorzata E. [1 ,2 ]
Amiel, Stephanie A. [3 ]
Hovorka, Roman [1 ,2 ]
Murphy, Helen R. [1 ,2 ]
机构
[1] Univ Cambridge, Metab Res Labs, Cambridge CB2 0QQ, England
[2] Univ Cambridge, Natl Inst Hlth Res, Cambridge Biomed Res Ctr, Cambridge CB2 0QQ, England
[3] Kings Coll London, Diabet Res Grp, London WC2R 2LS, England
基金
英国医学研究理事会;
关键词
SUBCUTANEOUS INSULIN INFUSION; HYPOGLYCEMIA; TEMPERATURE; ADOLESCENTS; EFFICACY; SAFETY; SYSTEM;
D O I
10.1089/dia.2012.0292
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Performance of continuous glucose monitors (CGMs) may be lower when glucose levels are changing rapidly, such as occurs during physical activity. Our aim was to evaluate accuracy of a current-generation CGM during moderate-intensity exercise in type 1 diabetes (T1D) pregnancy. Subjects and Methods: As part of a study of 24-h closed-loop insulin delivery in 12 women with T1D (disease duration, 17.6 years; glycosylated hemoglobin, 6.4%) during pregnancy (gestation, 21 weeks), we evaluated the Freestyle Navigator (R) sensor (Abbott Diabetes Care, Alameda, CA) during afternoon (15:00-18:00 h) and morning (09:30-12:30 h) exercise (55 min of brisk walking on a treadmill followed by a 2-h recovery), compared with sedentary conditions (18:00-09:00 h). Plasma (reference) glucose, measured at regular 15-30-min intervals with the YSI Ltd. (Fleet, United Kingdom) model YSI 2300 analyzer, was used to assess CGM performance. Results: Sensor accuracy, as indicated by the larger relative absolute difference (RAD) between paired sensor and reference glucose values, was lower during exercise compared with rest (median RAD, 11.8% vs. 18.4%; P < 0.001). These differences remained significant when correcting for plasma glucose relative rate of change (P < 0.001). Analysis by glucose range showed lower accuracy during hypoglycemia for both sedentary (median RAD, 24.4%) and exercise (median RAD, 32.1%) conditions. Using Clarke error grid analysis, 96% of CGM values were clinically safe under resting conditions compared with only 87% during exercise. Conclusions: Compared with sedentary conditions, accuracy of the Freestyle Navigator CGM was lower during moderate-intensity exercise in pregnant women with T1D. This difference was particularly marked in hypoglycemia and could not be solely explained by the glucose rate of change associated with physical activity.
引用
收藏
页码:223 / 229
页数:7
相关论文
共 50 条
  • [1] Point Accuracy of Interstitial Continuous Glucose Monitoring During Exercise in Type 1 Diabetes
    Yardley, Jane E.
    Sigal, Ronald J.
    Kenny, Glen P.
    Riddell, Michael C.
    Lovblom, Leif E.
    Perkins, Bruce A.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 (01) : 46 - 49
  • [2] Accuracy Of Continuous Glucose Monitoring During Exercise In Type I Diabetes Patients
    Vandoni, Matteo
    Pellino, Vittoria Carnevale
    Correale, Luca
    Liberali, Giulia
    Natalucci, Valentina
    Bonfadini, Silvia
    Girelli, Angela
    Buzzachera, Cosme Franklim
    MEDICINE & SCIENCE IN SPORTS & EXERCISE, 2020, 52 (07) : 133 - 133
  • [3] Accuracy of continuous glucose monitoring system during exercise in type 2 diabetes
    Figueira, Franciele R.
    Umpierre, Daniel
    Ribeiro, Jorge P.
    Tetelbom, Pedro S.
    Henn, Nicoli T.
    Esteves, Jorge F.
    Schaan, Beatriz D.
    DIABETES RESEARCH AND CLINICAL PRACTICE, 2012, 98 (03) : E36 - E39
  • [4] ACCURACY OF CONTINUOUS GLUCOSE MONITORING DURING RESTING AND EXERCISE CONDITIONS IN INDIVIDUALS WITH TYPE 1 DIABETES
    Bally, L.
    Paganini, D.
    Pasi, N.
    Loher, H.
    Feller, K.
    Zueger, T.
    Stettler, C.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 : A93 - A93
  • [5] Accuracy of continuous glucose monitoring during exercise-related hypoglycemia in individuals with type 1 diabetes
    Maytham, Kaisar
    Hagelqvist, Per G.
    Engberg, Susanne
    Forman, Julie L.
    Pedersen-Bjergaard, Ulrik
    Knop, Filip K.
    Vilsboll, Tina
    Andersen, Andreas
    FRONTIERS IN ENDOCRINOLOGY, 2024, 15
  • [6] Factory Calibrated Continuous Glucose Monitoring System Accuracy During Exercise In Adolescents With Type 1 Diabetes Mellitus
    Dyess, Ryan
    Thrasher, Bradly
    Wintergerst, Kupper
    Feygin, Yana
    McKay, Tim
    HORMONE RESEARCH IN PAEDIATRICS, 2022, 95 (SUPPL 1): : 45 - 46
  • [7] Accuracy of continuous glucose monitoring systems during exercise-related hypoglycaemia in patients with type 1 diabetes
    Maytham, K.
    Hagelqvist, P. G.
    Pedersen-Bjergaard, U.
    Knop, F. K.
    Andersen, A.
    Vilsboll, T.
    DIABETOLOGIA, 2022, 65 (SUPPL 1) : S118 - S118
  • [8] Accuracy of Continuous Glucose Monitoring before, during, and after Aerobic and Anaerobic Exercise in Patients with Type 1 Diabetes Mellitus
    Biagi, Lyvia
    Bertachi, Arthur
    Quiros, Carmen
    Gimenez, Marga
    Conget, Ignacio
    Bondia, Jorge
    Vehi, Josep
    BIOSENSORS-BASEL, 2018, 8 (01):
  • [9] ACCURACY AND PRECISION OF CONTINUOUS GLUCOSE MONITORING BEFORE, DURING, AND AFTER AEROBIC AND RESISTANCE EXERCISE IN SUBJECTS WITH TYPE 1 DIABETES
    Biagi, L.
    Bertachi, A.
    Quiros, C.
    Gimenez, M.
    Conget, I.
    Bondia, J.
    Vehi, J.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2018, 20 : A92 - A93
  • [10] Factory-Calibrated Continuous Glucose Monitoring System Accuracy During Exercise in Adolescents With Type 1 Diabetes Mellitus
    Dyess, Ryan J.
    McKay, Timothy
    Feygin, Yana
    Wintergerst, Kupper
    Thrasher, Bradly J.
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2024, 18 (03): : 584 - 591