Wear analysis of cross-linked polyethylene inserts articulating with alumina and ion-treated cobalt-chromium femoral heads under third-body conditions

被引:6
|
作者
Hembus, Jessica [1 ]
Lux, Laura [1 ]
Jackszis, Mario [1 ]
Bader, Rainer [1 ]
Zietz, Carmen [1 ]
机构
[1] Univ Med Rostock, Dept Orthopaed, Biomech & Implant Technol Res Lab, Doberaner Str 142, D-18057 Rostock, Germany
关键词
Hip joint simulator; Ceramics; Ion treatment; Cross-linked polyethylene; Third-body wear; Tribology; TOTAL KNEE ARTHROPLASTY; TOTAL HIP REPLACEMENTS; ALTERNATIVE MATERIALS; ACETABULAR LINERS; SIMULATOR MODEL; JOINT SIMULATOR; ABRASIVE WEAR; BONE-CEMENT; PARTICLES; DAMAGE;
D O I
10.1016/j.wear.2018.02.017
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Aseptic implant loosening due to wear debris is the main cause of revision of total hip replacement. Since the presence of third-body particles in the joint gap can lead to highly increased amount of wear, the aim of this study was to evaluate wear of hard-on-soft couplings under third-body wear conditions. For this, alumina ceramic and nitrogen ion-treated cobalt-chromium femoral heads were coupled with sequentially cross-linked polyethylene inserts in a hip joint simulator following ISO 14242-1. Bone cement particles containing zirconium dioxide were added. After 5 million cycles, the amount of wear on the inserts was determined gravimetrically and compared to the amount of wear without third-body particles. The abrasion at the joint components was evaluated by digital microscope. The mean gravimetric wear rates of the polyethylene inserts under third-body wear conditions measured 16.57 +/- 5.98 mg per million cycles when coupled with nitrogen ion-treated CoCr heads and 15.14 +/- 3.30 mg per million cycles with alumina ceramic heads. The results revealed no significant differences between the material pairings. The amount of wear was approximately eight times higher in both cases than that resulting under standard wear test conditions. In the microscopic analysis, the ceramic femoral heads showed higher resistance against abrasive third-body wear.
引用
收藏
页码:216 / 223
页数:8
相关论文
共 15 条
  • [1] Third-body wear of highly cross-linked polyethylene in a hip simulator
    Bragdon, CR
    Jasty, M
    Muratoglu, OK
    O'Connor, DO
    Harris, WH
    JOURNAL OF ARTHROPLASTY, 2003, 18 (05): : 553 - 561
  • [2] Differences in Highly Cross-Linked Polyethylene Wear Between Zirconia and Cobalt-Chromium Femoral Heads in Japanese Patients A prospective, randomized study
    Kawate, Kenji
    Ohmura, Tetsuji
    Kawahara, Ikuo
    Tamai, Katsuya
    Ueha, Tomoyuki
    Takemura, Kazuo
    JOURNAL OF ARTHROPLASTY, 2009, 24 (08): : 1221 - 1224
  • [3] Wear Performance of Sequentially Cross-Linked Polyethylene Inserts against Ion-Treated CoCr, TiNbN-Coated CoCr and Al2O3 Ceramic Femoral Heads for Total Hip Replacement
    Fabry, Christian
    Zietz, Carmen
    Baumann, Axel
    Bader, Rainer
    LUBRICANTS, 2015, 3 (01): : 14 - 26
  • [4] Fifteen-Year Results of Total Hip Arthroplasty With Cobalt-Chromium Femoral Heads on Highly Cross-Linked Polyethylene in Patients 50 Years and Less
    Rames, Richard D.
    Stambough, Jeffrey B.
    Pashos, Gail E.
    Maloney, William J.
    Martell, John M.
    Clohisy, John C.
    JOURNAL OF ARTHROPLASTY, 2019, 34 (06): : 1143 - 1149
  • [5] Excellent 20-Year Results of Total Hip Arthroplasty With Highly Cross-Linked Polyethylene on Cobalt-Chromium Femoral Heads in Patients ≤50 Years
    Parilla, Frank W.
    Youngman, Tyler R.
    Layon, Daniel R.
    Ince, Deniz C.
    Pashos, Gail E.
    Maloney, William J.
    Clohisy, John C.
    JOURNAL OF ARTHROPLASTY, 2024, 39 (02): : 409 - 415
  • [6] Minimum Five-Year Follow-Up Wear Measurement of Longevity Highly Cross-Linked Polyethylene Cup Against Cobalt-Chromium or Zirconia Heads
    Nakahara, Ichiro
    Nakamura, Nobuo
    Nishii, Takashi
    Miki, Hidenobu
    Sakai, Takashi
    Sugano, Nobuhiko
    JOURNAL OF ARTHROPLASTY, 2010, 25 (08): : 1182 - 1187
  • [7] Delamination of a highly cross-linked polyethylene liner associated with titanium deposits on the cobalt-chromium modular femoral head following dislocation
    Patten, E. W.
    Atwood, S. A.
    Van Citters, D. W.
    Jewett, B. A.
    Pruitt, L. A.
    Ries, M. D.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (09): : 1306 - 1311
  • [8] Long-Term Results of Total Hip Arthroplasty with 28-Millimeter Cobalt-Chromium Femoral Heads on Highly Cross-Linked Polyethylene in Patients 50 Years and Less
    Stambough, Jeffrey B.
    Pashos, Gail
    Bohnenkamp, Frank C.
    Maloney, William J.
    Martell, John M.
    Clohisy, John C.
    JOURNAL OF ARTHROPLASTY, 2016, 31 (01): : 162 - 167
  • [9] Prediction of scratch resistance of cobalt chromium alloy bearing surface, articulating against ultra-high molecular weight polyethylene, due to third-body wear particles
    Mirghany, M
    Jin, ZM
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2004, 218 (H1) : 41 - 50
  • [10] Wear and Migration of Highly Cross-Linked and Conventional Cemented Polyethylene Cups with Cobalt Chrome or Oxinium Femoral Heads: A Randomized Radiostereometric Study of 150 Patients
    Kadar, Thomas
    Hallan, Geir
    Aamodt, Arild
    Indrekvam, Kari
    Badawy, Mona
    Skredderstuen, Arne
    Havelin, Leif Ivar
    Stokke, Terje
    Haugan, Kristin
    Espehaug, Birgitte
    Furnes, Ove
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2011, 29 (08) : 1222 - 1229