Comparative osteoconductivity of bone void fillers with antibiotics in a critical size bone defect model

被引:8
|
作者
Oliver, Rema A. [1 ]
Lovric, Vedran [1 ]
Christou, Chris [1 ]
Walsh, William R. [1 ]
机构
[1] UNSW Sydney, Surg & Orthopaed Res Labs, Prince Wales Clin Sch, Prince Wales Hosp, Level 1 Clin Sci Bldg, Randwick, NSW, Australia
关键词
CALCIUM-SULFATE; GRAFT SUBSTITUTE; SURGICAL DEBRIDEMENT; IN-VITRO; OSTEOMYELITIS; TOBRAMYCIN; VANCOMYCIN; IMPLANTS; CEMENT; BEADS;
D O I
10.1007/s10856-020-06418-1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The study aimed to evaluate the comparative osteoconductivity of three commercially available bone void fillers containing gentamicin with respect to new bone, growth, host tissue response and resorption of the implant material. Defects were created in the cancellous bone of the distal femur and proximal tibia of 12-skeletally mature sheep and filled with three commercially available bone void fillers containing gentamicin (Stimulan-G, Cerament-G, Herafill-G). Peripheral blood was taken pre-operatively and at the time of implantation, as well as at intermittent timepoints following surgery to determine systemic gentamicin levels (5-,15- and 30-minutes, 1, 2, 3, 6, 12, 24, 48- and 72-hours, 3-, 6- and 12-weeks). Decalcified, embedded samples were stained with haematoxylin and eosin (H&E) and used to assess the host tissue response and the formation of new bone in the presence of test implant materials. No adverse reactions were noted at harvest at any time points for any cancellous implantation sites with the various implant materials. Comparative microCT analysis of the Stimulan-G, Cerament-G and Herafill-G test materials revealed a similar increase in bone surface area and volume between animals implanted with Stimulan-G or Cerament-G test materials. Animals implanted with Herafill-G test materials demonstrated the lowest increases in bone volume and surface area of the test materials tested, at levels similar to the negative control sites. By 12-weeks, Stimulan-G defects were completely closed with mature bone and bone marrow whilst the Cerament-G material was still present after 12 weeks by histological examination. In conclusion, this study demonstrated differences in the bone regenerative capacity of a range of bone void fillers in an in vivo setting. [GRAPHICS] .
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Comparative osteoconductivity of bone void fillers with antibiotics in a critical size bone defect model
    Rema A. Oliver
    Vedran Lovric
    Chris Christou
    William R. Walsh
    Journal of Materials Science: Materials in Medicine, 2020, 31
  • [2] Author Correction: Comparative osteoconductivity of bone void fillers with antibiotics in a critical size bone defect model
    Rema A. Oliver
    Vedran Lovric
    Chris Christou
    William R. Walsh
    Journal of Materials Science: Materials in Medicine, 2021, 32
  • [3] Comparative osteoconductivity of bone void fillers with antibiotics in a critical size bone defect model (vol 31, 80, 2020)
    Oliver, Rema A.
    Lovric, Vedran
    Christou, Chris
    Walsh, William R.
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2021, 32 (10)
  • [4] Evaluation of comparative soft tissue response to bone void fillers with antibiotics in a rabbit intramuscular model
    Oliver, Rema A.
    Lovric, Vedran
    Christou, Chris
    Walsh, William R.
    JOURNAL OF BIOMATERIALS APPLICATIONS, 2019, 34 (01) : 117 - 129
  • [5] Bone Void Fillers
    Kirkpatrick, John S.
    Cornell, Charles N.
    Hoang, Bang H.
    Hsu, Wellington
    Watson, J. Tracy
    Watters, William C., III
    Turkelson, Charles M.
    Wies, Janet L.
    Anderson, Sara
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2010, 18 (09) : 576 - 579
  • [6] A new standardized critical size bone defect model in the pig forehead for comparative testing of bone regeneration materials
    Moest, Tobias
    Schlegel, Karl Andreas
    Kesting, Marco
    Fenner, Matthias
    Lutz, Rainer
    Beck, Daniele Machado
    Nkenke, Emeka
    von Wilmowsky, Cornelius
    CLINICAL ORAL INVESTIGATIONS, 2020, 24 (05) : 1651 - 1661
  • [7] A new standardized critical size bone defect model in the pig forehead for comparative testing of bone regeneration materials
    Tobias Moest
    Karl Andreas Schlegel
    Marco Kesting
    Matthias Fenner
    Rainer Lutz
    Daniele Machado Beck
    Emeka Nkenke
    Cornelius von Wilmowsky
    Clinical Oral Investigations, 2020, 24 : 1651 - 1661
  • [8] Size Matters: Defining Critical in Bone Defect Size!
    Schemitsch, Emil H.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2017, 31 : S20 - S22
  • [9] The monocortical critical size bone defect as an alternative experimental model in testing bone substitute materials
    Schlegel, Karl Andreas
    Lang, Friedemann Johannes
    Donath, Karl
    Kulow, Jochen Theodor
    Wiltfang, Joerg
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2006, 102 (01): : 7 - 13
  • [10] Natural bovine anorganic apatite and collagen presents osteoconductivity and contribute to bone repair of rat calvaria critical size defect
    Silva, I.I.C.
    Pimentel-Soares, S.
    Bittencourt, R.C.
    Granjeiro, J.M.
    Key Engineering Materials, 2009, 396-398 : 249 - 252