Introducing multi-energy ratios as an alternative to multi-energy calibration for Br determination via high-resolution continuum source graphite furnace molecular absorption spectrometry. A case study

被引:2
|
作者
Garde, Raul [1 ]
Nakadi, Flavio, V [1 ]
Garcia-Ruiz, Esperanza [1 ]
Resano, Martin [1 ]
机构
[1] Univ Zaragoza, Aragon Inst Engn Res I3A, Dept Analyt Chem, Pedro Cerbuna 12, Zaragoza 50009, Spain
关键词
INDUCTIVELY-COUPLED PLASMA; ISOTOPE DETERMINATION; SOLID SAMPLES; BROMINE; INTERFERENCE; NONMETALS; SULFUR; FLAMES; CABR;
D O I
10.1039/d0ja00359j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This manuscript explores the advantages of using multi-signal calibration approaches for the determination of non-metals via high-resolution continuum source graphite furnace molecular absorption spectrometry (HR CS GFMAS), targeting Br as an example. Besides multi-energy calibration (MEC), a novel approach deriving from it, multi-energy ratios (MER), is introduced and compared under different conditions. This approach makes use of the same data but in a different way, such that no linear regression is performed; instead, ratios are calculated. This article investigates the potential errors deriving from the use of amounts of spike dissimilar from the sample content, leading to too high (close to 1) or too low (close to 0) slopes/ratios, setting the best conditions in terms of precision and accuracy for the intended determination in the range of approx. 0.5 to 0.6. Also, situations where the use of MER could be recommended over MEC are identified: namely when only a few transitions of sufficient sensitivity and free from overlaps are available or else, many transitions but of similar sensitivity, which may occur when HR CS GFMAS is deployed. Otherwise, for multiple transitions covering a wider sensitivity range, use of linear regression and thus, of MEC, seems favoured, as a better precision can be achieved. The calculation of limits of detection and quantification for both approaches is also discussed. It is finally further demonstrated that these multi-signal strategies help in solving chemical interferences, which very often hamper the determination of non-metals with HR CS GFMAS, and they do so in a simple way, without the need for laborious work or for the preparation of several standards and sample aliquots, therefore making them a very intriguing option when this technique is deployed.
引用
收藏
页码:2606 / 2619
页数:14
相关论文
共 50 条
  • [1] Multi-energy calibration for the determination of non-metals by high-resolution continuum source molecular absorption spectrometry
    Vieira, Alan L.
    Goncalves, Daniel A.
    Virgilio, Alex
    Ferreira, Edilene C.
    Jones, Bradley T.
    Donati, George L.
    Gomes Neto, Jose A.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2019, 34 (05) : 972 - 978
  • [2] Determination of boron isotope ratios by high-resolution continuum source molecular absorption spectrometry using graphite furnace vaporizers
    Abad, Carlos
    Florek, Stefan
    Becker-Ross, Helmut
    Huang, Mao-Dong
    Heinrich, Hans-Joachim
    Recknagel, Sebastian
    Vogl, Jochen
    Jakubowski, Norbert
    Panne, Ulrich
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2017, 136 : 116 - 122
  • [3] Determination of bromine using high-resolution continuum source molecular absorption spectrometry in a graphite furnace
    Limburg, Tobias
    Einax, Jurgen W.
    MICROCHEMICAL JOURNAL, 2013, 107 : 31 - 36
  • [4] Chlorine isotope determination via the monitoring of the AlCl molecule by high-resolution continuum source graphite furnace molecular absorption spectrometry - a case study
    Nakadi, F. V.
    da Veiga, M. A. M. S.
    Aramendia, M.
    Garcia-Ruiz, E.
    Resano, M.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2015, 30 (07) : 1531 - 1540
  • [5] Determination of residual dimethyl sulfoxide by high-resolution continuum source graphite furnace molecular absorption spectrometry
    Huang, Mao Dong
    Esser, Norbert
    Hinrichs, Karsten
    Tan, Qin
    Rappich, Jorg
    Nickel, Norbert H.
    Dittrich, Thomas
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2021, 177
  • [6] Br isotope determination via the monitoring of CaBr transitions using high-resolution continuum source graphite furnace molecular absorption spectrometry. Potential for direct determination of Br in solid samples using isotope dilution
    Nakadi, F. V.
    da Veiga, M. A. M. S.
    Aramendia, M.
    Garcia-Ruiz, E.
    Resano, M.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2016, 31 (07) : 1381 - 1390
  • [7] Determination of non-metals via molecular absorption using high-resolution continuum source absorption spectrometry and graphite furnace atomization
    Heitmann, Uwe
    Becker-Ross, Helmut
    Florek, Stefan
    Huang, Mao Dong
    Okruss, Michael
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2006, 21 (11) : 1314 - 1320
  • [8] Determination of Chlorine in Milk via Molecular Absorption of SrCI Using High-Resolution Continuum Source Graphite Furnace Atomic Absorption Spectrometry
    Ozbek, Nil
    Akman, Suleyman
    JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2016, 64 (28) : 5767 - 5772
  • [9] High-resolution continuum source graphite furnace molecular absorption spectrometry for the monitoring of Sr isotopes via SrF formation: a case study
    Bazo, Antonio
    Garde, Raul
    Garcia-Ruiz, Esperanza
    Aramendia, Maite
    Nakadi, Flavio, V
    Resano, Martin
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2022, 37 (12) : 2517 - 2528
  • [10] Indirect determination of cysteine in pharmaceutical formulations by high-resolution continuum source graphite furnace molecular absorption spectrometry
    Almeida, Jorge S.
    Meira, Lucilia A.
    Teixeira, Leonardo S. G.
    MICROCHEMICAL JOURNAL, 2018, 143 : 155 - 159