Dosimetric advantages of using multichannel balloons compared to single-channel cylinders for high-dose-rate vaginal cuff brachytherapy

被引:20
|
作者
Serrano, Nicholas [1 ]
Moghanaki, Drew [1 ,2 ]
Asher, David [1 ]
Karlin, Jeremy [1 ]
Schutzer, Matthew [1 ,2 ]
Chang, Michael [1 ,2 ]
Hagan, Michael P. [1 ,2 ]
机构
[1] Virginia Commonwealth Univ, Dept Radiat Oncol, Massey Canc Ctr, Richmond, VA 23221 USA
[2] Hunter Holmes McGuire VA Med Ctr, Dept Radiat Oncol, Richmond, VA 23249 USA
关键词
Endometrial carcinoma; Vaginal cuff brachytherapy; Multichannel balloon; LOCALIZED PROSTATE-CANCER; EXTERNAL-BEAM RADIOTHERAPY; PERMANENT SOURCE BRACHYTHERAPY; VOLUME HISTOGRAM ANALYSIS; RADIATION-THERAPY; LONG-TERM; GASTROINTESTINAL TOXICITY; AMERICAN BRACHYTHERAPY; MORBIDITY OUTCOMES; INTERMEDIATE-RISK;
D O I
10.1016/j.brachy.2016.04.002
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: To evaluate the dosimetric advantages of using multichannel balloons (MCBs) vs. single-channel cylinders (SCCs) for high-dose-rate vaginal cuff brachytherapy. METHODS AND MATERIALS: A total of 91 consecutive high-dose-rate vaginal cuff brachytherapy including 45 MCB and 46 SCC treatments were reviewed. The clinical target volume (CTV) was defined as a 0.5-cm uniform expansion of the applicator surface from vaginal apex for 3 cm. For dosimetric comparison, we normalized prescription dose per fraction to 700 cGy and optimized each plan to cover at least 90% of CTV. CTV-1 cm, the true vaginal cuff volume, was defined as proximal 1 cm of CTV from vaginal apex. Four quality indices including conformity index (CI), dose homogeneity index, dose nonuniformity index, and overdose index were compared. RESULTS: The CTV and CTV-1 cm were significantly larger for MCB cases compared to SCC cases. Evaluating CTV coverage, the mean dose homogeneity index and dose nonuniformity index were superior for MCB than SCC. No differences were noted regarding CI and overdose index between MCB and SCC cases. However, focusing on CTV-1 cm, the difference of CI became significant in favor of MCB cases. In addition, the mean point dose at 0.5-cm depth from the apex was significantly lower in SCC cases compared to cases by MCB treatment, indicating inadequate vaginal apex coverage by SCC treatment. CONCLUSIONS: Compared to SCC, MCB treats a larger volume and offers a more conformal and homogeneous target coverage. In addition, a lower dose at the vaginal apex due to SCCs source anisotropy can be minimized. (C) 2016 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:435 / 441
页数:7
相关论文
共 50 条
  • [1] Dosimetric comparison of multichannel with one single-channel vaginal cylinder for vaginal cancer treatments with high-dose-rate brachytherapy
    Kim, Hayeon
    Rajagopalan, Malolan S.
    Houser, Chris
    Beriwal, Sushil
    BRACHYTHERAPY, 2014, 13 (03) : 263 - 267
  • [2] A Dosimetric Study by Using a Multichannel Cylinder in High-Dose-Rate Brachytherapy Treatment of the Vaginal Cuff
    Syh, J. H.
    Zhang, M.
    White, M.
    Patel, B.
    Syh, J.
    Song, X.
    Wu, H.
    Rosen, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S501 - S501
  • [3] The use and advantages of a multichannel vaginal cylinder in high-dose-rate brachytherapy
    Demanes, DJ
    Rege, S
    Rodriquez, RR
    Schutz, KL
    Altieri, GA
    Wong, T
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 44 (01): : 211 - 219
  • [4] Dosimetric Advantages in Critical Structure Dose Sparing by Using a Multichannel Cylinder in High Dose Rate Brachytherapy to Treat Vaginal Cuff Cancer
    Syh, J.
    Syh, J.
    Patel, B.
    Zhang, J.
    Wu, H.
    Rosen, L.
    MEDICAL PHYSICS, 2014, 41 (06) : 388 - 388
  • [5] Dosimetric Evaluation of Multi-Channel Cylinders for High Dose Rate Vaginal Cuff Brachytherapy Using a Cylindrical Target Volume
    Dogan, S. Kucuker
    Kang, Z.
    Cates, J.
    Zhang, H.
    MEDICAL PHYSICS, 2019, 46 (06) : E576 - E576
  • [6] (Potential) Mishaps of High-Dose-Rate Vaginal Cuff Brachytherapy
    Wallner, Kent
    Sherertz, Tracy
    Anderson, August
    Blau, Molly
    Panjwani, Neil
    PRACTICAL RADIATION ONCOLOGY, 2023, 13 (04) : 334 - 339
  • [7] Dose Optimization for Single-Channel Vaginal Cylinder High-Dose-Rate Brachytherapy: A Double Prescription Method for Patients With Endometrial Adenocarcinoma
    Wu, Xingen
    McDonald, Andrew
    Shen, Sui
    Stanley, Dennis
    Popple, Richard
    Marcrom, Samuel
    Kim, Robert
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (06)
  • [8] Single or multi-channel vaginal cuff high-dose-rate brachytherapy: Is replanning necessary prior to each fraction?
    Zhou, Jessica
    Prisciandaro, Joann
    Lee, Choonik
    Schipper, Matthew
    Eisbruch, Avraham
    Jolly, Shruti
    PRACTICAL RADIATION ONCOLOGY, 2014, 4 (01) : 20 - 26
  • [9] An Optimization Study and Comparison of a Multichannel Vaginal Cylinder Versus Single Channel Cylinder Applicator in High-dose-rate Brachytherapy
    Syh, J. H.
    Wu, T.
    Syh, J.
    Patel, B.
    Rosen, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S456 - S457
  • [10] Assessment of air pockets in high-dose-rate vaginal cuff brachytherapy using cylindrical applicators
    Hassouna, Ashraf
    Bahadur, Yasir Abdulaziz
    Constantinescu, Camelia
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2014, 6 (03) : 271 - 275