A case of neural integrity monitor endotracheal tube malfunction: What to blame? Cancelled surgery due to NIM tracheal tube malfunction e a case report

被引:1
|
作者
Dardeer, Abdulrahman [1 ]
Lafir, Ahamed [1 ]
Krishnan, Chitrambika [1 ]
Albassam, Saba [1 ]
Hammad, Yasser [1 ,2 ]
AlAbdulla, Majid [3 ]
Zaki, Hany [4 ]
Shallik, Nabil [1 ,2 ,5 ,6 ,7 ]
机构
[1] Hamad Med Corp, Anaesthesia ICU & Perioperat Med Dept, Doha, Qatar
[2] Weill Cornell Med Coll Qatar, Clin Anaesthesia Dept, Doha, Qatar
[3] Hamad Med Corp, ENT Dept, Doha, Qatar
[4] Hamad Med Corp, Emergency Dept, Doha, Qatar
[5] Qatar Univ, Coll Med, Clin Anaesthesia Dept, Doha, Qatar
[6] Tanta Univ, Clin Anaesthesia & SICU Dept, Tanta, Egypt
[7] POB 3050, Doha, Qatar
关键词
Neural integrity monitor (NIM); Dental braces; Dental retainer; Equipment malfunction; Thyroid surgery; Recurrent laryngeal nerve; RECURRENT LARYNGEAL NERVE; THYROID-SURGERY;
D O I
10.1016/j.tacc.2023.101259
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Intraoperative neuromonitoring of the recurrent laryngeal nerve (RLN) is becoming the standard of care in surgeries of the brainstem, skull base, thyroid, and parathyroid with the potential risk for RLN damage. The Neural Integrity Monitor (NIM) endotracheal tube is a specialized tube that utilizes electromyog-raphy (EMG) for identifying the RLN intraoperatively. Vocal cords movement (if the touched structure is a nerve) is captured and shown on a monitor visually with a distinctive sound to alert the team. We describe a case of recurrent thyroid papillary carcinoma who needed surgical intervention for the third time, in which a NIM tube was used. After careful troubleshooting and failure to solve the issue, the artefacts were attributed to the patient's dental retainers (braces). The surgery was aborted. Subse-quently, several tests were done on dental models with and without braces using the very NIM tube used on the patient. Our tests showed no difference in artefacts, suggesting a faulty NIM tube. However, no research, evidence, or manufacturer recommendations exist related to the use of the NIM ETT with dental retainers. There is no evidence suggesting signal interference due to metal braces. The surgery was carried out uneventfully on a subsequent day with a new NIM tube.(c) 2023 The Authors. Published by Elsevier Ltd.This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
引用
收藏
页数:3
相关论文
共 31 条
  • [1] Tracheal Obstruction as a Complication of Tracheostomy Tube Malfunction Case Report and Review of the Literature
    Lois, Manuel
    Oltermann, Mark
    [J]. JOURNAL OF BRONCHOLOGY & INTERVENTIONAL PULMONOLOGY, 2010, 17 (03) : 253 - 257
  • [2] Difficult extubation of a damaged neural integrity monitor electromyogram tracheal tube A case report
    Moon, HoSik
    Hong, SungJin
    Sung, ChoonHo
    Chon, JinYoung
    Kwak, JuEun
    Lee, JiYung
    [J]. MEDICINE, 2020, 99 (25) : E20250
  • [3] Difficult tracheal extubation due to endotracheal tube malfunction: A challenge during the COVID-19 pandemic
    Bamgbade, Olumuyiwa A.
    Magboh, Vivian O.
    Otegbeye, Ajibola U.
    King, Mmakgomo B.
    Oluwole, Olusanya J.
    Olatunji, Bolusefe T.
    [J]. JOURNAL OF TAIBAH UNIVERSITY MEDICAL SCIENCES, 2021, 16 (06): : 935 - 937
  • [4] Case report: Endotracheal tube malposition in a patient with a tracheal bronchus
    Wong, David T.
    Kumar, Ashok
    [J]. CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2006, 53 (08): : 810 - 813
  • [5] Ventilation Failure due to Endotracheal Tube Migration Into a Tracheal Pouch in a Toddler With a Repaired Tracheoesophageal Fistula: A Case Report
    Mathew, Preethy J.
    Ashok, Vighnesh
    Malik, Muneer Abas
    [J]. A & A PRACTICE, 2020, 14 (02): : 48 - 50
  • [6] Avoidable tracheal injury using a double-lumen endotracheal tube: a case report
    Lemay, Judith
    Conti, Massimo
    Somma, Jacques
    Vigneault, Louise
    Bussieres, Jean S.
    [J]. CANADIAN JOURNAL OF ANESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2018, 65 (06): : 735 - 736
  • [7] Avoidable tracheal injury using a double-lumen endotracheal tube: a case report
    Judith Lemay
    Massimo Conti
    Jacques Somma
    Louise Vigneault
    Jean S. Bussières
    [J]. Canadian Journal of Anesthesia/Journal canadien d'anesthésie, 2018, 65 : 735 - 736
  • [8] Airway obstruction due to intraoperative endotracheal tube cuff herniation: case report
    Karakaya, Muhammet Ahmet
    Tukac, Ismail Cem
    Kutlu, Esra
    Karaaslan, Pelin
    Tercan, Elvan
    [J]. BIOMEDICAL RESEARCH-INDIA, 2017, 28 (06): : 2391 - 2392
  • [9] Adverse event using Medtronic NIM™ EMG endotracheal tube on a patient receiving anesthesia for hemithyroidectomy: a case report
    Carpenter, Emilee
    Norris, Lorraine
    Beniamin, Myriam
    [J]. BMC ANESTHESIOLOGY, 2022, 22 (01)
  • [10] Adverse event using Medtronic NIM™ EMG endotracheal tube on a patient receiving anesthesia for hemithyroidectomy: a case report
    Emilee Carpenter
    Lorraine Norris
    Myriam Beniamin
    [J]. BMC Anesthesiology, 22