Automatic image registration on intraoperative CBCT compared to Surface Matching registration on preoperative CT for spinal navigation: accuracy and workflow

被引:1
|
作者
Frisk, Henrik [1 ]
Burstrom, Gustav [1 ]
Persson, Oscar [1 ]
El-Hajj, Victor Gabriel [1 ]
Coronado, Luisa [2 ]
Hager, Susanne [2 ]
Edstrom, Erik [1 ,3 ]
Elmi-Terander, Adrian [1 ,3 ,4 ]
机构
[1] Karolinska Inst, Dept Clin Neurosci, S-17177 Stockholm, Sweden
[2] Clin Affairs, Brainlab AG, Munich, Germany
[3] Lowenstromska Hosp, Capio Spine Ctr Stockholm, Upplands Vasby, Sweden
[4] Uppsala Univ, Dept Surg Sci, Uppsala, Sweden
关键词
Patient tracking; Reference frame; Surface Matching; CBCT; Spine surgery; Surgical navigation; PEDICLE SCREW PLACEMENT; TO-WORLD REGISTRATION; COMPUTED-TOMOGRAPHY; ARM NAVIGATION; SYSTEM; SURGERY; RISK;
D O I
10.1007/s11548-024-03076-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
IntroductionSpinal navigation solutions have been slower to develop compared to cranial ones. To facilitate greater adoption and use of spinal navigation, the relatively cumbersome registration processes need to be improved upon. This study aims to validate a new solution for automatic image registration and compare it to a traditional Surface Matching method.MethodAdult patients undergoing spinal surgery requiring navigation were enrolled after providing consent. A registration matrix-Universal AIR (= Automatic Image Registration)-was placed in the surgical field and used for automatic registration based on intraoperative 3D imaging. A standard Surface Matching method was used for comparison. Accuracy measurements were obtained by comparing planned and acquired coordinates on the vertebrae.ResultsThirty-nine patients with 42 datasets were included. The mean accuracy of Universal AIR registration was 1.20 +/- 0.42 mm, while the mean accuracy of Surface Matching registration was 1.94 +/- 0.64 mm. Universal AIR registration was non-inferior to Surface Matching registration. Post hoc analysis showed a significantly greater accuracy for Universal AIR registration. In Surface Matching, but not automatic registration, user-related errors such as incorrect identification of the vertebral level were seen.ConclusionAutomatic image registration for spinal navigation using Universal AIR and intraoperative 3D imaging provided improved accuracy compared to Surface Matching registration. In addition, it minimizes user errors and offers a standardized workflow, making it a reliable registration method for navigated spinal procedures.
引用
收藏
页码:665 / 675
页数:11
相关论文
共 41 条
  • [1] Automatic image registration on intraoperative CBCT compared to Surface Matching registration on preoperative CT for spinal navigation: accuracy and workflow
    Henrik Frisk
    Gustav Burström
    Oscar Persson
    Victor Gabriel El-Hajj
    Luisa Coronado
    Susanne Hager
    Erik Edström
    Adrian Elmi-Terander
    International Journal of Computer Assisted Radiology and Surgery, 2024, 19 : 665 - 675
  • [2] Automatic Image Registration Provides Superior Accuracy Compared with Surface Matching in Cranial Navigation
    Frisk, Henrik
    Jensdottir, Margret
    Coronado, Luisa
    Conrad, Markus
    Hager, Susanne
    Arvidsson, Lisa
    Bartek, Jiri
    Burström, Gustav
    El-Hajj, Victor Gabriel
    Edström, Erik
    Elmi-Terander, Adrian
    Persson, Oscar
    Sensors, 2024, 24 (22)
  • [3] Preoperative CT and Intraoperative CBCT Image Registration and Evaluation in Robotic Cochlear Implant Surgery
    Lu, Chenxi
    Dong, Bo
    Hu, Xi
    Zhao, Yang
    He, Hongjian
    Wang, Jing
    MEDICAL IMAGE UNDERSTANDING AND ANALYSIS, MIUA 2022, 2022, 13413 : 116 - 129
  • [4] Comparison of radiation exposure and surgery time between an intraoperative CT with automatic surface registration and a preoperative CT with manual surface registration in navigated spinal surgeries
    Martin Mohammed Marzouk
    Yama Afghanyar
    Mark Mahmoud Marzouk
    Sarah Halima Boussouf
    Philipp Hartung
    Marcus Richter
    European Spine Journal, 2022, 31 : 685 - 692
  • [5] Comparison of radiation exposure and surgery time between an intraoperative CT with automatic surface registration and a preoperative CT with manual surface registration in navigated spinal surgeries
    Marzouk, Martin Mohammed
    Afghanyar, Yama
    Marzouk, Mark Mahmoud
    Boussouf, Sarah Halima
    Hartung, Philipp
    Richter, Marcus
    EUROPEAN SPINE JOURNAL, 2022, 31 (03) : 685 - 692
  • [6] Current accuracy of surface matching compared to adhesive markers in patient-to-image registration
    Mireli A. Mongen
    Peter W. A. Willems
    Acta Neurochirurgica, 2019, 161 : 865 - 870
  • [7] Current accuracy of surface matching compared to adhesive markers in patient-to-image registration
    Mongen, Mireli A.
    Willems, Peter W. A.
    ACTA NEUROCHIRURGICA, 2019, 161 (05) : 865 - 870
  • [8] Accuracy evaluation of shape-based registration for image-guided hip surgery - Effect of quality of preoperative CT model and intraoperative data on registration accuracy
    Sasama, T
    Sato, Y
    Sugano, N
    Yamaguchi, K
    Tamura, S
    Ohzono, K
    Ochi, T
    CAR '98 - COMPUTER ASSISTED RADIOLOGY AND SURGERY, 1998, 1165 : 911 - 911
  • [9] The Use of Skin Staples as Fiducial Markers to Confirm Intraoperative Spinal Navigation Registration and Accuracy
    Alvarez-Breckenridge, Christopher
    Muir, Matthew
    Rhines, Laurence D.
    Tatsui, Claudio E.
    OPERATIVE NEUROSURGERY, 2021, 21 (03) : E193 - E198
  • [10] A multiscale adaptive mask method for rigid intraoperative ultrasound and preoperative CT image registration
    Zhang, Zhijun
    Liu, Feng
    Tsui, Hungtat
    Lau, Yunwong
    Song, Xubo
    MEDICAL PHYSICS, 2014, 41 (10)