Dosimetric Comparison between Single-energy Computed Tomography and Dual-energy Computed Tomography Relative to Stopping Power Estimation in Proton Therapy

被引:0
|
作者
Chirdchid, Thamonwan [1 ]
Ruangchan, Sirinya [2 ,3 ]
Sanghangthum, Taweap [1 ,2 ,4 ]
机构
[1] Chulalongkorn Univ, Dept Radiol, Fac Med, Bangkok, Thailand
[2] Her Royal Highness Princess Maha Chakri Sirindhor, Bangkok, Thailand
[3] Thai Red Cross Soc, King Chulalongkorn Mem Hosp, Div Radiat Oncol, Dept Radiol, Bangkok, Thailand
[4] Chulalongkorn Univ, Div Radiat Oncol, Dept Radiol, Fac Med, Bangkok, Thailand
关键词
Dual-energy computed tomography; proton therapy; single-energy computed tomography; stoichiometric method; DOSE CALCULATION; CT; CALIBRATION;
D O I
10.4103/jmp.jmp_27_23
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The focus of this work was given on the relative stopping power (RSP) using the water equivalent thickness (WET) validation on tissue substitutes and real pig organs, as well as a dosimetric comparison of proton treatment plans between single-energy computed tomography (SECT) and dual-energy computed tomography (DECT)-based dose calculations. Materials and Methods: The CT calibration curve of SECT and DECT data was generated using the stoichiometric calibration method. WET measurement was performed for RSP validation using a Giraffe dosimeter (IBA dosimetry) in various substitute tissues (Gammex) and real pig tissues. The thorax (008A, CIRS) and head (731-HN, CIRS) phantoms were used to generate proton plans. The dosimetric evaluations of SECT and DECT-based plans were performed using the gamma analysis with 1%/1 mm and the dose-volume histograms (DVHs) comparison. Results: For RSP validation of substitute tissues, the largest percent WET difference between measurement and calculation was observed up to 17.9% (4 mm) in lung tissue, using SECT based. In real pig tissues, the average WET difference was 2.3% +/- 2.1% and 2.5% +/- 2.3% for SECT and DECT, respectively. The average gamma passed of about 92.1% for the lung and 96.8% for the head regions was reported. For the lung region, the DVH of the target dose was observed with a higher predicted dose in SECT than in DECT, while results in the head region were in good agreement for both SECT and DECT. Conclusion: The performed dosimetric comparison indicates the dose differences between SECT and DECT. The impact of the CT calibration curve is more pronounced for the thorax region.
引用
收藏
页码:292 / 297
页数:6
相关论文
共 50 条
  • [1] Dual-Energy Computed Tomography: Is There a Penalty in Image Quality and Radiation Dose Compared With Single-Energy Computed Tomography?
    Thomas, Christoph
    Ketelsen, Dominik
    Tsiflikas, Ilias
    Reimann, Anja
    Brodoefel, Harald
    Claussen, Claus D.
    Heuschmid, Martin
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2010, 34 (02) : 309 - 315
  • [2] Comparison of stopping power estimators from dual-energy computed tomography for protontherapy
    Vilches-Freixas, G.
    Letang, J. M.
    Rit, S.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S869 - S869
  • [3] Dosimetric comparison of stopping power calibration with dual-energy CT and single-energy CT in proton therapy treatment planning
    Zhu, Jiahua
    Penfold, Scott N.
    MEDICAL PHYSICS, 2016, 43 (06) : 2845 - 2854
  • [4] A Convolutional Neural Network Approach with Dual-Energy Computed Tomography for Proton Stopping-Power-Ratio Estimation
    Lee, H.
    Duan, X.
    Park, Y.
    Jia, X.
    Jiang, S.
    Yang, M.
    MEDICAL PHYSICS, 2020, 47 (06) : E445 - E446
  • [5] Accuracy of Dual-Energy Computed Tomography Techniques for Fat Quantification in Comparison With Magnetic Resonance Proton Density Fat Fraction and Single-Energy Computed Tomography in an Anthropomorphic Phantom Environment
    Wu, Zi Jun
    Hippe, Daniel S.
    Zamora, David A.
    Briller, Noah
    Amin, Kathan A.
    Kolokythas, Orpheus
    Mileto, Achille
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2021, 45 (06) : 877 - 887
  • [6] Should Dual-Energy Computed Tomography Pulmonary Angiography Replace Single-Energy Computed Tomography Pulmonary Angiography in Pregnant and Postpartum Patients?
    McDermott, Shaunagh
    Otrakji, Alexi
    Flores, Efren J.
    Kalra, Mannudeep K.
    Shepard, Jo-Anne O.
    Digumarthy, Subba R.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (01) : 25 - 32
  • [7] Utility of single-energy and dual-energy computed tomography in clot characterization: An in-vitro study
    Brinjikji, Waleed
    Michalak, Gregory
    Kadirvel, Ramanathan
    Dai, Daying
    Gilvarry, Michael
    Duffy, Sharon
    Kallmes, David F.
    McCollough, Cynthia
    Leng, Shuai
    INTERVENTIONAL NEURORADIOLOGY, 2017, 23 (03) : 279 - 284
  • [8] Image quality comparison of single-energy and dual-energy computed tomography for head and neck patients: a prospective randomized study
    Bedernik, Andreas
    Wuest, Wolfgang
    May, Matthias Stefan
    Heiss, Rafael
    Uder, Michael
    Wiesmueller, Marco
    EUROPEAN RADIOLOGY, 2022, 32 (11) : 7700 - 7709
  • [9] Image quality comparison of single-energy and dual-energy computed tomography for head and neck patients: a prospective randomized study
    Andreas Bedernik
    Wolfgang Wuest
    Matthias Stefan May
    Rafael Heiss
    Michael Uder
    Marco Wiesmueller
    European Radiology, 2022, 32 : 7700 - 7709
  • [10] Introduction to Dual-Energy Computed Tomography
    Elster, Allen D.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (06) : 823 - 823