A three-dimensional Magnetic Resonance Spin Tomography in Time-domain protocol for high-resolution multiparametric quantitative magnetic resonance imaging

被引:2
|
作者
Liu, Hongyan [1 ]
van der Heide, Oscar [1 ]
Versteeg, Edwin [1 ]
Froeling, Martijn [2 ]
Fuderer, Miha [1 ]
Xu, Fei [1 ]
van den Berg, Cornelis A. T. [1 ]
Sbrizzi, Alessandro [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiotherapy, Computat Imaging Grp MRI Therapy & Diagnost, Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Dept Radiol, Imaging Div, Utrecht, Netherlands
关键词
MR-STAT; multiparametric MRI; parallel imaging; quantitative MRI; relaxometry; 7; T; MRI; RELAXATION; ARTIFACTS; ROBUST; BRAIN;
D O I
10.1002/nbm.5050
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Magnetic Resonance Spin TomogrAphy in Time-domain (MR-STAT) is a multiparametric quantitative MR framework, which allows for simultaneously acquiring quantitative tissue parameters such as T1, T2, and proton density from one single short scan. A typical two-dimensional (2D) MR-STAT acquisition uses a gradient-spoiled, gradient-echo sequence with a slowly varying RF flip-angle train and Cartesian readouts, and the quantitative tissue maps are reconstructed by an iterative, model-based optimization algorithm. In this work, we design a three-dimensional (3D) MR-STAT framework based on previous 2D work, in order to achieve better image signal-to-noise ratio, higher though-plane resolution, and better tissue characterization. Specifically, we design a 7-min, high-resolution 3D MR-STAT sequence, and the corresponding two-step reconstruction algorithm for the large-scale dataset. To reduce the long acquisition time, Cartesian undersampling strategies such as SENSE are adopted in our transient-state quantitative framework. To reduce the computational burden, a data-splitting scheme is designed for decoupling the 3D reconstruction problem into independent 2D reconstructions. The proposed 3D framework is validated by numerical simulations, phantom experiments, and in vivo experiments. High-quality knee quantitative maps with 0.8 x 0.8 x 1.5 mm(3) resolution and bilateral lower leg maps with 1.6 mm isotropic resolution can be acquired using the proposed 7-min acquisition sequence and the 3-min-per-slice decoupled reconstruction algorithm. The proposed 3D MR-STAT framework could have wide clinical applications in the future.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Three-dimensional high-resolution magnetic resonance imaging of ocular and orbital malignancies
    McCaffery, S
    Simon, EM
    Fischbein, NJ
    Rowley, HA
    Shimikawa, A
    Lin, S
    O'Brien, JM
    ARCHIVES OF OPHTHALMOLOGY, 2002, 120 (06) : 747 - 754
  • [2] High-resolution three-dimensional magnetic resonance imaging of mouse lung in situ
    Scadeng, Miriam
    Rossiter, Harry B.
    Dubowitz, David J.
    Breen, Ellen C.
    INVESTIGATIVE RADIOLOGY, 2007, 42 (01) : 50 - 57
  • [3] Three-Dimensional High-Resolution Magnetic Resonance Imaging for the Assessment of Cervical Artery Dissection
    Zhu, Xianjin
    Shan, Yi
    Guo, Runcai
    Zheng, Tao
    Zhang, Xuebin
    Liu, Zunjing
    Liu, Kunpeng
    FRONTIERS IN AGING NEUROSCIENCE, 2022, 14
  • [4] Visualization of inner ear structures by three-dimensional high-resolution magnetic resonance imaging
    Arnold, B
    Jager, L
    Grevers, G
    AMERICAN JOURNAL OF OTOLOGY, 1996, 17 (03): : 480 - 485
  • [5] Quantitative analysis with volume rendering of pathological myopic eyes by high-resolution three-dimensional magnetic resonance imaging
    Zhou, Jinqiong
    Tu, Ying
    Chen, Qinghua
    Wei, Wenbin
    MEDICINE, 2020, 99 (42) : E22685
  • [6] Three-dimensional high-resolution fast spin-echo coronary magnetic resonance angiography
    Stuber, M
    Botnar, RM
    Spuentrup, E
    Kissinger, KV
    Manning, WJ
    MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (02) : 206 - 211
  • [7] Domain approach to three-dimensional structure of rhodopsin using high-resolution nuclear magnetic resonance
    Albert, AD
    Yeagle, PL
    VERTEBRATE PHOTOTRANSDUCTION AND THE VISUAL CYCLE, PART A, 2000, 315 : 107 - +
  • [8] Three-dimensional ultrashort echo time magnetic resonance imaging in assessment of idiopathic pulmonary fibrosis, in comparison with high-resolution computed tomography
    Yang, Xiaoyan
    Liu, Min
    Duan, Jianghui
    Sun, Haishuang
    An, Jing
    Benkert, Thomas
    Dai, Huaping
    Wang, Chen
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2022, 12 (08) : 4176 - +
  • [9] Correlation of Prostate High-Resolution Microultrasound With Multiparametric Magnetic Resonance Imaging
    Pickersgill, Nicholas A.
    Alkazemi, Muhammad Hassan
    Ostergar, Adam
    Joseph, Karan
    Vetter, Joel M.
    Barashi, Nimrod S.
    Kim, Eric H.
    Andriole, Gerald L.
    Sivaraman, Arjun
    UROLOGY, 2025, 197 : 33 - 39
  • [10] Evaluation of Three-Dimensional High-Resolution Magnetic Resonance Imaging in the Diagnosis of Intracranial Atherosclerotic Stenosis
    Cheng, Yu-Fei
    Chen, Chen
    Liu, Hong
    Wu, Jiang
    INTERNATIONAL JOURNAL OF MORPHOLOGY, 2022, 40 (05): : 1344 - 1348