The effect of two-part dead layer thickness on the efficiency curve of a P-TYPE HPGe detector calculated using the Monte Carlo method

被引:0
|
作者
Aksoy, Gulper [1 ]
Unlu, Hasan [1 ]
Orhan, Nilguen [2 ]
Yuksel, Ayhan [2 ]
Bolukdemir, Mustafa Hicabi [3 ,4 ]
机构
[1] Gazi Univ, Grad Sch Nat & Appl Sci, Adv Technol, Ankara, Turkiye
[2] Turkish Energy Nucl & Min Res Author, Nucl Energy Res Inst, Istanbul, Turkiye
[3] Gazi Univ, Fac Sci, Dept Phys, Ankara, Turkiye
[4] Gazi Univ, Basic & Engn Sci Cent Lab Applicat & Res Ctr GUTMA, Ankara, Turkiye
关键词
HPGe; FEPE; Dead layer; Monte Carlo; PHITS; GESPECOR; SIMULATION; INCREASE;
D O I
10.1016/j.radphyschem.2023.111314
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The full energy peak efficiency (FEPE) of a gamma ray obtained from experimental studies using a high purity germanium detector (HPGe) can be easily reproduced by simulating the experiment in any desired geometry and energy range with Monte Carlo (MC) simulation methods. However, the dead layer (DL) thickness, which is one of the geometric information provided by the manufacturer, changes over time and affects the efficiency calculation, so it must be determined and compared with the initial value for the increases during the life-time of the detector. In this study, the general-purpose MC code PHITS and the special-purpose MC code GESPECOR were used to simulate a p-type HPGe detector and to determine the thickness of the frontal and lateral dead layer. As a result, it was observed that the frontal dead layer thickness increased from 0.7 mm to 1.075 mm, and the lateral dead layer thickness increased from 0.7 mm to 1.7 mm after 7 years of operation of the detector. This increase in dead layer thickness causes both a decrease in the detector's active volume and a decrease in efficiency in the problematic low-energy region.
引用
收藏
页数:6
相关论文
共 32 条
  • [1] Efficiency transfer in the calibration of a coaxial p-type HpGe detector using the Monte Carlo method
    Vargas, MJ
    Díaz, NC
    Sánchez, DP
    APPLIED RADIATION AND ISOTOPES, 2003, 58 (06) : 707 - 712
  • [2] The influence of dead layer thickness increase on efficiency decrease for a coaxial HPGe p-type detector
    Ngo Quang Huy
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3): : 390 - 394
  • [3] MONTE CARLO SIMULATION OF P-TYPE HPGe DETECTORS - THE DEAD LAYER PROBLEM
    Stancu, E.
    Costache, C.
    Sima, O.
    ROMANIAN REPORTS IN PHYSICS, 2015, 67 (02) : 465 - 473
  • [4] Dead-layer thickness effect for gamma spectra measured in an HPGe p-type detector
    Ngo Quang Huy
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 641 (01): : 101 - 104
  • [5] Dead layer thickness characterization of an HPGe detector by measurements and Monte Carlo simulations
    Elanique, A.
    Marzocchi, O.
    Leone, D.
    Hegenbart, L.
    Breustedt, B.
    Oufni, L.
    APPLIED RADIATION AND ISOTOPES, 2012, 70 (03) : 538 - 542
  • [6] Optimization of p-type HPGe detector model using Monte Carlo simulation
    Le Thi Ngoc Trang
    Huynh Dinh Chuong
    Tran Thien Thanh
    Journal of Radioanalytical and Nuclear Chemistry, 2021, 327 : 287 - 297
  • [7] Optimization of p-type HPGe detector model using Monte Carlo simulation
    Trang, Le Thi Ngoc
    Chuong, Huynh Dinh
    Thanh, Tran Thien
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2021, 327 (01) : 287 - 297
  • [8] STUDY ON VARIATION OF HPGE DETECTOR DEAD LAYER THICKNESS AND ITS EFFECT ON THE MEASUREMENTS OF THE DETECTOR RESPONSE AND SAMPLES CHARACTERIZATION USING MONTE CARLO SIMULATION
    Abd El Gawad, K.
    Song, Yushou
    PROCEEDINGS OF THE 26TH INTERNATIONAL CONFERENCE ON NUCLEAR ENGINEERING, 2018, VOL 3, 2018,
  • [9] A novel method to estimate the dead layer of HPGe detector for Monte Carlo FEPE computation
    Satti, Khizar Hayat
    Siddique, Muhammad Tariq
    Dilband, Muhammad
    Rehman, Shakeel Ur
    Mansoor, Shahid
    Malik, Azhar Hussain
    APPLIED RADIATION AND ISOTOPES, 2023, 202
  • [10] Characterisation of two p-type HPGe detectors by measurements and Monte Carlo simulations