Aerosol Measurement Degradation in Low-Cost Particle Sensors Using Laboratory Calibration and Field Validation

被引:1
|
作者
Peck, Angela [1 ]
Handy, Rodney G. [1 ]
Sleeth, Darrah K. [1 ]
Schaefer, Camie [2 ]
Zhang, Yue [3 ]
Pahler, Leon F. [1 ]
Ramsay, Joemy [1 ]
Collingwood, Scott C. [4 ]
机构
[1] Univ Utah, Dept Family & Prevent Med, Occupat & Environm Hlth, Salt Lake City, UT 84108 USA
[2] Univ Utah, Dept Family & Prevent Med, Salt Lake City, UT 84108 USA
[3] Univ Utah, Dept Internal Med, Salt Lake City, UT 84108 USA
[4] Univ Utah, Dept Pediat, Salt Lake City, UT 84108 USA
基金
美国国家卫生研究院;
关键词
particulate matter; low-cost sensors; exposure assessment; sensor degradation; citizen science; PARTICULATE AIR-POLLUTION; MASS CONCENTRATION; DAILY MORTALITY; MATTER; FINE; AMBIENT; QUALITY; ASSOCIATIONS; NETWORK; COUNTER;
D O I
10.3390/toxics11010056
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Increasing concern over air pollution has led to the development of low-cost sensors suitable for wide-scale deployment and use by citizen scientists. This project investigated the AirU low-cost particle sensor using two methods: (1) a comparison of pre- and post-deployment calibration equations for 24 devices following use in a field study, and (2) an in-home comparison between 3 AirUs and a reference instrument, the GRIMM 1.109. While differences (and therefore some sensor degradation) were found in the pre- and post-calibration equation comparison, absolute value changes were small and unlikely to affect the quality of results. Comparison tests found that while the AirU did tend to underestimate minimum and overestimate maximum concentrations of particulate matter, similar to 88% of results fell within +/- 1 mu g/m(3) of the GRIMM. While these tests confirm that low-cost sensors such as the AirU do experience some sensor degradation over multiple months of use, they remain a valuable tool for exposure assessment studies. Further work is needed to examine AirU performance in different environments for a comprehensive survey of capability, as well as to determine the source of sensor degradation.
引用
下载
收藏
页数:13
相关论文
共 50 条
  • [1] Suitability of Low-Cost Sensors for Submicron Aerosol Particle Measurement
    Stoll, Daniel
    Kerner, Maximilian
    Paas, Simon
    Antonyuk, Sergiy
    APPLIED SYSTEM INNOVATION, 2023, 6 (04)
  • [2] Laboratory Evaluation and Calibration of Low-Cost Sensors for Air Quality Measurement
    Ionascu, Marian-Emanuel
    Gruicin, Iasmina
    Marcu, Marius
    2018 IEEE 12TH INTERNATIONAL SYMPOSIUM ON APPLIED COMPUTATIONAL INTELLIGENCE AND INFORMATICS (SACI), 2018, : 395 - 400
  • [3] Calibration of Low-cost Sensors for Measurement of Indoor Particulate Matter Concentrations via Laboratory/Field Evaluation
    Kim, Doheon
    Shin, Dongmin
    Hwang, Jungho
    AEROSOL AND AIR QUALITY RESEARCH, 2023, 23 (08)
  • [4] Calibration of Low-Cost Particle Sensors by Using Machine-Learning Method
    Chen, Chen-Chia
    Kuo, Chih-Ting
    Chen, Ssu-Ying
    Lin, Chih-Hsing
    Chue, Jin-Ju
    Hsieh, Yi-Jie
    Cheng, Chun-Wen
    Wu, Chieh-Ming
    Huang, Chun-Ming
    2018 IEEE ASIA PACIFIC CONFERENCE ON CIRCUITS AND SYSTEMS (APCCAS 2018), 2018, : 111 - 114
  • [5] A machine learning field calibration method for improving the performance of low-cost particle sensors
    Patra, Satya S.
    Ramsisaria, Rishabh
    Du, Ruihang
    Wu, Tianren
    Boor, Brandon E.
    BUILDING AND ENVIRONMENT, 2021, 190
  • [6] A machine learning field calibration method for improving the performance of low-cost particle sensors
    Patra, Satya S.
    Ramsisaria, Rishabh
    Du, Ruihang
    Wu, Tianren
    Boor, Brandon E.
    Building and Environment, 2021, 190
  • [7] Perspectives on the Calibration and Validation of Low-Cost Air Quality Sensors
    Ottosen, Thor-Bjorn
    ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2021, 55 (19) : 12773 - 12775
  • [8] Field and laboratory evaluation of PurpleAir low-cost aerosol sensors in monitoring indoor airborne particles
    Park, Seongjun
    Lee, Shinhye
    Yeo, Myoungsouk
    Rim, Donghyun
    BUILDING AND ENVIRONMENT, 2023, 234
  • [9] Laboratory Evaluation and Calibration of Three Low- Cost Particle Sensors for Particulate Matter Measurement
    Wang, Yang
    Li, Jiayu
    Jing, He
    Zhang, Qiang
    Jiang, Jingkun
    Biswas, Pratim
    AEROSOL SCIENCE AND TECHNOLOGY, 2015, 49 (11) : 1063 - 1077
  • [10] CFD Simulations Of An Aerosol Chamber For Calibration Of Low-Cost Particulate Matter Sensors
    Drzymalla, Jan
    Lay, Yannic
    Sauermann, Marc
    Henne, Andreas
    INDOOR ENVIRONMENTAL QUALITY PERFORMANCE APPROACHES (IAQ 2020), PT 1, 2021,