Fluorination of PVC medical devices to prevent plasticizers migration

被引:0
|
作者
Dhifallah, Imen [1 ]
Claves, Daniel [1 ]
Batisse, Nicolas [1 ]
Cueff, Regis [1 ]
Sapin, Anne-Francoise [2 ]
Bouattour, Yassine [2 ]
Petit, Elodie [1 ]
Dubois, Marc [1 ]
Sautou, Valerie [2 ]
机构
[1] Univ Clermont Auvergne, Inst Chim Clermont Ferrand, F-63178 Aubiere, France
[2] Univ Clermont Auvergne, CHU Clermont Ferrand, CNRS, SIGMA Clermont,ICCF, F-63000 Clermont Ferrand, France
关键词
Plasticizer; PVC; Migration; Fluorination; Polymer; Cytotoxicity; SURFACE FREE-ENERGY; POLY(VINYL CHLORIDE); DI-2-ETHYLHEXYL PHTHALATE; BETA-CYCLODEXTRIN; DEHP; ENHANCEMENT; XPS;
D O I
10.1016/j.ijpharm.2023.123254
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Medical devices (MD) are often made of plasticized polyvinylchloride (PVC). However, plasticizers may leach out into infused solutions and expose the patients to a toxic risk. The aim of the present work is to fluorinate plasticized PVC tubular MDs to create a barrier layer on their internal surface, and to study the impact of such a chemical treatment on the migration of the plasticizers. Following fluorination by pure molecular fluorine, the physico-chemical characterization of these modified MDs was carried out using various spectroscopic and microscopic techniques or tensile tests, evidencing the formation of covalent C-F bonds on the surface of the treated samples without modification of their mechanical and optical properties. The migration of plasticizers from fluorinated MDs was assessed using gas chromatography coupled with mass spectrometry and was found considerably decreased in comparison with the pristine MDs. After 24 h, the amount of tri-octyltrimellitate plasticizer (TOTM) detected in migrates from fluorinated MDs was even lower than the limit of quantification. Complementary cytotoxicity assays were performed according to the ISO EN 10993-5 standard, showing that the new fluorinated material does not cause a cytotoxic effect on L929 cells.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Migration of plasticizers from PVC medical devices: Development of an infusion model
    Bernard, L.
    Cueff, R.
    Chagnon, M. C.
    Abdoulouhab, F.
    Decaudin, B.
    Breysse, C.
    Kauffmann, S.
    Cosserant, B.
    Souweine, B.
    Sautoua, V.
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2015, 494 (01) : 136 - 145
  • [2] Migration of Phthalate-based Plasticizers from PVC and non-PVC Containers and Medical Devices
    Veiga, Marlei
    Bohrer, Denise
    Nascimento, Paulo C.
    Ramirez, Adrian G.
    Carvalho, Leandro M.
    Binotto, Regina
    JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY, 2012, 23 (01) : 72 - 77
  • [3] Effects of flow rate on the migration of different plasticizers from PVC infusion medical devices
    Bernard, Lise
    Eljezi, Teuta
    Clauson, Helene
    Lambert, Celine
    Bouattour, Yassine
    Chennell, Philip
    Pereira, Bruno
    Sautou, Valerie
    PLOS ONE, 2018, 13 (02):
  • [4] Leaching of plasticizers from PVC medical devices: A molecular interpretation of experimental migration data
    Sahnoune Millot M.
    Devémy J.
    Chennell P.
    Pinguet J.
    Dequidt A.
    Sautou V.
    Malfreyt P.
    Journal of Molecular Liquids, 396
  • [5] Study of the migration of PVC plasticizers
    Marcilla, A
    Garía, S
    García-Quesada, JC
    JOURNAL OF ANALYTICAL AND APPLIED PYROLYSIS, 2004, 71 (02) : 457 - 463
  • [6] MIGRATION OF PVC PLASTICIZERS INTO ALCOHOLS
    KAMPOURIS, EM
    REGAS, F
    ROKOTAS, S
    POLYCHRONAKIS, S
    PANTAZOGLOU, M
    POLYMER, 1975, 16 (11) : 840 - 844
  • [7] Migrability of PVC plasticizers from medical devices into a simulant of infused solutions
    Bernard, L.
    Cueff, R.
    Breysse, C.
    Decaudin, B.
    Sautou, V.
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2015, 485 (1-2) : 341 - 347
  • [8] Migration of monomeric and polymeric PVC plasticizers
    Hakkarainen, Minna
    CHROMATOGRAPHY FOR SUSTAINABLE POLYMERIC MATERIALS: RENEWABLE, DEGRADABLE AND RECYCLABLE, 2008, 211 : 159 - 185
  • [9] Analysis of plasticizers in PVC medical devices: Performance comparison of eight analytical methods
    Bernard, L.
    Bourdeaux, D.
    Pereira, B.
    Azaroual, N.
    Barthelemy, C.
    Breysse, C.
    Chennell, P.
    Cueff, R.
    Dine, T.
    Eljezi, T.
    Feutry, F.
    Genay, S.
    Kambia, N.
    Lecoeur, M.
    Masse, M.
    Odou, P.
    Radaniel, T.
    Simon, N.
    Vaccher, C.
    Verlhac, C.
    Yessad, M.
    Decaudin, B.
    Sautou, V.
    TALANTA, 2017, 162 : 604 - 611
  • [10] Analysis of PVC plasticizers in medical devices and infused solutions by GC-MS
    Bourdeaux, Daniel
    Yessaad, Mouloud
    Chennell, Philip
    Larbre, Virginie
    Eljezi, Teuta
    Bernard, Lise
    Sautou, Valerie
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2016, 118 : 206 - 213