Determination of Inducible Clindamycin Resistance and Correlation with Vitek2 Inducible Clindamycin Resistance Test in Staphylococcus aureus Isolated from Clinical Samples

被引:0
|
作者
Zafar, Adeel [1 ]
Khursheed, Nazia [1 ]
Adnan, Fareeha [1 ]
Zahid, Qurat ul Ain [1 ]
机构
[1] Indus Hosp & Hlth Network, Dept Pathol Microbiol, Karachi, Pakistan
关键词
Methicillin-resistant Staphylococcus aureus; Inducible clindamycin resistance; D zone; Vitek2; PERFORMANCE; MACROLIDE; SYSTEM;
D O I
10.29271/jcpsp.2024.02.183
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To compare the performance of Vitek2 with the gold standard D test in terms of inducible clindamycin resistance (ICR) detection. Study Design: Descriptive study. Place and Duration of the Study: Indus Hospital and Health Network Karachi, Pakistan, from November 2021 to April 2022. Methodology: Standard operating procedures of the laboratory were followed for processing of clinical samples. Methicillin-resistant Staphylococcus aureus (MRSA) isolates were included. The isolates from the same patient within two-week time period were excluded. Clinical laboratory standards institute guidelines were followed for performing and interpreting D test. The results of the D test were compared with Vitek2 results for ICR. Results: A total of 313 isolates were MRSA, of which 93 isolates tested positive for ICR on both the D test and Vitek2. Nine isolates were positive for ICR on Vitek2 and negative on the Kirby-Bauer disk diffusion method. One isolate tested positive on the disk method and negative on Vitek2. Conclusion: Vitek 2 appeared to give false positive results. Reporting false susceptibility of clindamycin can cause therapeutic failure which can markedly affect the patient's outcome. This discordance needs to be investigated further with a large sample size and stringent observation of D-test results to pick laboratory error.
引用
收藏
页码:183 / 186
页数:4
相关论文
共 50 条
  • [1] Inducible clindamycin resistance in Staphylococcus aureus isolated from clinical samples
    Deotale, V.
    Mendiratta, D. K.
    Raut, U.
    Narang, P.
    INDIAN JOURNAL OF MEDICAL MICROBIOLOGY, 2010, 28 (02) : 124 - 126
  • [2] Prevalence of inducible clindamycin resistance in Staphylococcus aureus strains isolated from clinical samples
    Colakoglu, Sule
    Aliskan, Hikmet
    Turunc, Tuba
    Demiroglu, Yusuf Ziya
    Arslan, Hande
    MIKROBIYOLOJI BULTENI, 2008, 42 (03): : 407 - 412
  • [3] Inducible clindamycin resistance in clinical isolates of Staphylococcus aureus
    Gadepalli, Ravisekhar
    Dhawan, Benu
    Mohanty, Srujana
    Kapil, Arti
    Das, Bimal K.
    Chaudhry, Rama
    INDIAN JOURNAL OF MEDICAL RESEARCH, 2006, 123 (04) : 571 - 573
  • [4] Clindamycin treatment of Staphylococcus aureus expressing inducible clindamycin resistance
    Drinkovic, D
    Fuller, ER
    Shore, KP
    Holland, DJ
    Ellis-Pegler, R
    JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 2001, 48 (02) : 315 - 316
  • [5] Methicillin resistance & inducible clindamycin resistance in Staphylococcus aureus
    Ghosh, Soumyadeep
    Banerjee, Mandira
    INDIAN JOURNAL OF MEDICAL RESEARCH, 2016, 143 : 362 - 364
  • [6] Inducible clindamycin resistance in staphylococci isolated from clinical samples
    Delialioglu, N
    Aslan, G
    Ozturk, C
    Baki, V
    Sen, S
    Emekdas, G
    JAPANESE JOURNAL OF INFECTIOUS DISEASES, 2005, 58 (02) : 104 - 106
  • [7] Inducible Clindamycin Resistance, Glycopeptide Resistance and Mupirocin Resistance in Methicillin Resistant Staphylococcus aureus (MRSA) Isolated from Clinical Samples
    Mangala, G. K.
    Rao, P. N. S.
    Vishwanath, G.
    Suresh, K.
    Vijayanath, V.
    JOURNAL OF PURE AND APPLIED MICROBIOLOGY, 2012, 6 (01): : 407 - 410
  • [8] Inducible resistance to clindamycin in Staphylococcus aureus: validation of Vitek-2 against CLSI D-test
    Gardiner, B. J.
    Grayson, M. L.
    Wood, G. M.
    PATHOLOGY, 2013, 45 (02) : 181 - 184
  • [9] INDUCIBLE RESISTANCE TO CLINDAMYCIN IN STAPHYLOCOCCUS AUREUS: VALIDATION OF VITEK-2 AGAINST CLSI D-TEST
    Gardiner, B.
    Wood, G.
    Grayson, L.
    INTERNAL MEDICINE JOURNAL, 2012, 42 : 5 - 5
  • [10] Evaluation of constitutive and inducible resistance to clindamycin in clinical samples of Staphylococcus aureus from a tertiary hospital
    Bottega, Angelita
    Rodrigues, Monica de Abreu
    Carvalho, Fernanda Aguirre
    Wagner, Tatiana Feyh
    Souza Leal, Isabel Agne
    dos Santos, Silvana Oliveira
    Rampelotto, Roberta Filipini
    Hoerner, Rosmari
    REVISTA DA SOCIEDADE BRASILEIRA DE MEDICINA TROPICAL, 2014, 47 (05) : 589 - 592